Abstract
One approach to increasing nuclear magnetic resonance measurement sample throughput is to implement multiple, independent detection sites. However, the presence of radio frequency interference poses a challenge in multi-detector systems, particularly in unshielded coil arrays lacking sufficient electrical isolation. This issue can lead to unwanted coupling of inductive coils, resulting in excitation pulse interference and signal transfer among multiple detection sites. Here we propose a theoretical framework that combines electromagnetic simulation with spin-dynamic calculations. This framework enables the evaluation of coil coupling effects, the design of parallel pulse sequences to mitigate inter-channel coupling, and the separation of composite free induction decays obtained from multiple detectors. The parallel pulse compensation scheme was validated by a 2-channel parallel spectroscopy experiment. These results provide valuable insights for the design of parallel nuclear magnetic resonance hardware and for exploring the limits of parallelization capacity within a fixed magnet system.
Similar content being viewed by others
Introduction
Nuclear magnetic resonance (NMR) spectroscopy has been widely used to reveal detailed molecular structure and structure-function relationships. Whereas NMR technology is routinely applied to analyze large numbers of samples or to monitor dynamic biochemistry processes, high-throughput capability has remained elusive. A straightforward way to increase throughput is to detect multiple samples simultaneously via multiple channels, broadly termed parallel NMR1. Several previous studies utilize multiple radio frequency (RF) coils for parallel excitation and reception1,2,3. Two broad approaches have been explored with regard to hardware design: The first method combines all coils into a single circuit with a shared tuning and matching circuit. It separates signals using a pulsed gradient field, which reduces electronic components but collects noise from the entire coil array. This configuration is limited to cases1,4 which have a relatively high signal-to-noise ratio (SNR). The second design deploys each coil with its own tuning and matching circuit and receiver chain2,5, where each sample can be excited and detected without the use of gradient fields as a means to achieve effective parallelization. The primary obstacle is achieving exceptional electrical isolation6 among parallel channels, in other words, the inter-coil coupling should be sufficiently weak. The method of suppressing coupling was recently discussed in a sample-centered shimming experiment7, where the coil geometry was optimized to minimize the coupling effect. However, the RF shielding design is practically limited by space and probe complexity, so the coupling cannot be completely eliminated. How to handle the unavoidable coupling remains as an open question. Our work explores software-based approaches to track the inter-channel coupling issue, both in the excitation stage (i.e., parallel pulse sequence design), and the reception stage (composite signal splitting).
In the realm of parallel detection, the incorporation of multiple receive coils also stands as a milestone within the MRI community8,9. This approach, known as ‘parallel imaging,’ has garnered approval for its capacity to enhance SNR10 and reduce acquisition time11. The evolution of parallel imaging led to the utilization of multiple transmit coils in the context of ‘parallel transmission’ (pTx)12. This application aimed at improving RF excitation, including enhancements in B1 homogeneity13, minimization of the specific energy absorption rate14, and even RF shimming15. Addressing the spatial sensitivity of each transmit element, the RF pulse underwent optimization to reduce the duration in spatially selective multidimensional excitation16. Given the dependence of the transmit B1 field on the experimental platform, subject-specific pulse optimization involved exploiting B0 and B1 maps as prior knowledge17,18,19. To circumvent field measurement and data processing, the concept of universal pulses emerged, demonstrating robustness against subject-dependent field inhomogeneity20,21. These universal pulses, designed based on a database of B-field maps from representative subjects, have been further refined using the GRAPE algorithm. For instance, the simultaneous optimization of RF and gradient waveforms can address B1 heterogeneity22 and off-resonance effects23. However, in the context of parallel NMR, the parallel concept aims to increase throughput, with each coil possessing a confined B1 field to excite a specific sample. The overarching objective is independent operation for each channel, making the coupling-induced ‘parallel transmission’ a hardware non-ideality that needs resolution. This is distinct from strongly coupled pTx in MRI, where pTx is intentionally employed to enhance excitation.
The mutual coupling within parallel MRI transmit arrays has been extensively studied, prompting the development of diverse decoupling strategies. These approaches including employing decoupling capacitance24, passive decoupling network25,26, or preamplifier decoupling through impedance mismatch27. A recent review article has summarized these decoupling methods28. While the preamplifier decoupling needs substantial effort on system integration, active decoupling found more interest. One example entails controlling the currents in eight-channel transverse electromagnetic transmit coils to attain independent transmit sensitivities, with control coefficients designed based on the impedance matrix of the coil array29,30. This active decoupling concept might be combined with the tailored RF pulses which have become the principal approach for manipulating nuclei spin. In this context, we explore parallel excitation by designing cooperative pulses aimed at mitigating inter-channel coupling, while avoiding the B field mappings. The overall effect of the cooperative pulse would be to achieve excitation, equivalent to that produced by each channel in the absence of coupling, and hence a coupling model for parallel excitation is crucial for this purpose.
As such, the approach is an extension to the design of non-identical, but cooperatively acting pulses as previously introduced for related heteronuclear Hartmann-Hahn sequences31, cooperativity for added scans32, and consecutive scans33.
In signal reception, understanding the impact of coupling on signal transfer is crucial. We need to determine how multiple free induction decays (FIDs) combine to generate detected signals, necessitating the development of a specific coupling model for the reception stage. In addition, we noticed that processing multi-channel NMR signals falls within the realm of handling composite signals from multiple detectors. In this domain, blind source separation (BSS) stands out as an exceptional model with diverse applications, including speech recognition34, image processing35, and bio-medical signal processing36. BSS methods, under appropriate assumptions such as the independent sources condition37, not only identifies signals from multiple sources, but also estimates the mixing matrix, which reveals the inter-detector coupling information. Therefore, the BSS method will serve as a complementary approach for the developed model in order to collaboratively uncover the inter-channel coupling.
In this work, an NMR simulation model was built to investigate the effects of inter-channel coupling in the case of a parallel detector configuration. This model accounts for the coil coupling-induced excitation distortion and FID composition. For these purposes, an electromagnetic simulation module was first built to calculate the B(t) field (B0, B1) within the multiple, parallel samples, the spin-dynamic calculations were performed by importing the B(t) field data and solving the Liouville-von Neumann equation38. Based on this model, a parallel pulse compensation scheme was put forward to address the RF coupling. Furthermore, we utilized the calculated coupling matrix to virtually separate the simulated composite signals and employed the BSS method to split both simulated and experimental signals. This enables the recovery of detector-specific signals and the identification of coupling. Finally, the pulse compensation scheme was validated through a parallel spectroscopy experiment using two channels. These results will help explore the limit of designing parallel NMR probes and experiments in a fixed volume, hence increasing sample throughput.
Results
The ultimate objective of this simulation study is to explore methods to handle RF coupling in parallel NMR, i.e., parallel pulse sequence optimization, and parallel signal decomposition. The initial step concerns developing a model to reveal how coupling influences parallel NMR, including an electromagnetic simulation to predict the resulting magnetic field, and a means to quantify coupling, as well as the spin-dynamic calculation using the simulated magnetic fields in order to reveal coupling effects on the spin.
Firstly, we constructed a simulation workflow for single-channel NMR experiments, inspired by a simulation methodology for predicting MR noise39 (Fig. 1). This workflow includes three parts: magnetic field acquisition, spin-dynamic simulation, and pulse sequence optimization. In the first part, we use COMSOL multi-physics to calculate the static magnetic field (B0) and RF field (B1) inside a sample, with well-defined coil geometry and electrical parameters (e.g. permittivity and permeability). The calculated fields were imported into the simulation package Spinach40 for spin dynamics calculations in part 2. Finally, part 3 focuses on pulse optimization considering the magnetic fields predicted from part 1.
This workflow was then adapted for a parallel NMR scenario, where the electromagnetic (EM) simulation considers a detector array, and the spin-dynamic calculation accommodates the spin evolution of multiple samples. To illustrate the framework without losing generality, we selected water samples and solenoid RF coils for the EM simulation, and an 11.74 T magnet for the primary magnetic field.
Modeling of parallel NMR
EM simulation
In parallel NMR experiments, multiple RF coils and samples are arranged as an array inside the magnet bore, where the coils couple with each other via an inductive-coupling effect (inter-coil coupling). In addition, the magnetic field from a local coil can spill into the external space and affect neighboring samples (coil-sample coupling), especially in unshielded coil arrays. To quantify the coupling and its impact on the parallel system performance, the RF coupling chain was modeled with the S-parameters, which relates the voltage waves incident on the RF ports to those reflected from the ports41. The B1 field was calculated by full-wave simulation with the calculated currents on the coils (Supplementary Note 1).
In Fig. 2a, a schematic of a single-channel RF probe is depicted, and a parallel NMR detector consists of multiple repetitions of this probe. Each channel is equipped with an individual tuning and matching (T&M) circuit, the T&Mcircuit is connected to the spectrometer, where the RF switch alternates between the Tx and Rx modes for excitation and reception, respectively. In Fig. 2c, we depict an RF chain built for modeling the coil array and T&Mnetwork. The amplifiers are substituted by matched loads, assuming that all the amplifiers are tuned and matched to the cable’s characteristic impedance. We use S-parameters referenced to 50 Ω to combine each part, and the forward and reverse traveling waves are indicated as a and b, respectively. These two waves can incorporate externally induced signals, including the FID and RF pulse.
To illustrate the RF chain, we first built a 4-solenoid array model together with cylindrical water samples in COMSOL, as shown in Fig. 2b, the feed ports connected to the T&Mnetwork were substituted by lumped ports with 50 Ω impedance. The RF simulation was carried out using the COMSOL ‘emw’ module, employing a frequency domain solver that solved the electric field wave equation through the finite element method. The frequency was set at 500 MHz, corresponding to the 1H Larmor frequency under an 11.74 T magnetic field. The simulation provides the coil array S matrix, denoted as SC, which correlates the voltage waves incident on the coils with those reflected from the coils. Additionally, the simulation generates the prototype B1 field under unit excitation (1 V). The T&Mnetwork was virtually designed, with each coil individually tuned and matched based on simulated Sc. In cases of very weak coupling (SC,21 < − 60dB), capacitance sets were calculated using diagonal elements of SC, achieving reflections below -20 dB. However, when coupling increases and leads to undesirable mismatches (SCM,11 > − 20dB), we employed the Nelder-Mead simplex algorithm42 for capacitance optimization (Supplementary Note 2). In this process, all channels shared the same capacitance sets, considering the symmetric geometry of the coil array. The calculated capacitances were then integrated into the coil array, resulting in an RF chain characterized by S-parameters.
The B1 field at a specific position under parallel excitation can be expressed as the linear superposition of the field values obtained from separate excitation of individual coils43, represented by B1 = ∑mB1m. As RF coils are linear components, B1m can be expressed as cmIm, where Im represents the actual current flowing through the m-th coil, calculated using the RF chain model. By performing a complete port sweep in COMSOL, we obtain N values of B1 and N combinations of port currents, allowing for the determination of the linear coefficients cm. Consequently, we could simulate the B1 field under a given excitation p using the equation
Here, F transfers the excitation p to the coil current, and T converts the current to the B1 field (Supplementary Note 3). As we iterate through each position, the full-wave simulation results are reserved, including the phase information. Using this approach, we simulated the B1 field for a 4-solenoid array, assuming only the first channel was excited in Fig. 2b. It’s important to note that we apply an incident voltage p at the T&Mnetwork terminal, assuming that the amplifier precisely delivers the required power to the RF probe. Therefore, the amplifier should operate within its linear range, and caution should be exercised when considering very high pulse powers44. In addition, the RF simulation conducted in the frequency domain yields steady-state results, so the transient response of a finite-Q circuit was neglected. This matter was addressed concerning simulating the transient response45 and measuring the impulse response function46.
To estimate the homogeneity of the static B0 field, we replicated the geometry used in the RF simulation within the COMSOL ‘mfnc’ module. The sample susceptibility was set to − 9.035 × 10−6, and the static field was determined based on Gauss’ law ∇ B = 0 within an 11.74 T background field.
Modeling the excitation stage
Figure 3a illustrates the signal flow during parallel excitation. The first stage involves calculating the B1 field using the provided pulse, supposing the synchronous pulse in each channel. The parallel RF pulses are amplified to a predetermined power level and delivered to the terminals of the T&Mnetwork. At the T&Mnetwork port, the excitation pulse p contributes to the forward waves a2, as depicted in Fig. 2c. The B1 field is computed using Eq. (1), enabling the determination of the field distribution across the entire space.
In the second stage, we address the spin evolution for multiple samples by incorporating them into a composite system that encompasses all spin isotopes and considers spin interactions. Mathematically, the system Hamiltonian is constructed by organizing the sub-Hamiltonians into blocks of a block-diagonal matrix, where each block corresponds to one sample. The corresponding spin state vectors are also assembled to align with the Hamiltonian. In Spinach, when defining the spin system, interactions between different samples are neglected under the assumption that sample-sample coupling, e.g. via radiation damping, can be disregarded (Supplementary Note 4). Spin operators are calculated and applied separately to spins from different samples. Therefore, the spins from multiple samples are individually excited, evolved, and detected, despite being within a single composite spin system.
This workflow also enables heteronuclear simulation, which applies to polarization transfer and high-dimensional NMR experiments. The relevant B1 fields would only need to be calculated for each desired nucleus.
Modeling the reception stage
Figure 3b shows the signal flow for parallel reception, each channel acquires the FIDs synchronously, i.e., all channels are fixed with the same sweeping frequency and acquisition period. The spin evolution was conducted in this period, giving the trajectories of the spin state, which is projected to the vector representing the coil to calculate the normalized FID, i.e, \({{{{{{{\bf{m}}}}}}}}\cdot {{{{{{{\bf{{b}}}}}}}^{-}}}= \langle{{{{{{{\rm{coil}}}}}}}}| {{{{{{{\boldsymbol{\rho }}}}}}}} \rangle\)47. The real FID was calculated according to the reciprocity principle48,
The relaxation effect was neglected, and the magnetization amplitude ∣M∣ was fixed at 1. The receiving field \(| {B}_{1}^{-}| {e}^{i\phi }={B}_{x}-i{B}_{y}\) was calculated using the same method as the transmit \({B}_{1}^{+}\), assuming a unit current through a specific coil. Signal calculations for the received signal utilized the RF chain depicted in Fig. 2c. Within the RF chain, the FID contributed to the open-circuit voltage at the coil terminals, a total gain was determined to transfer the FID to the T&Mnetwork terminal (Eq. S10). Ultimately, the signal at the T&Mnetwork output formed the input for the signal decomposition module, further details on signal decomposition will be provided in the subsequent section.
The RF coupling between heteronuclear channels is ignored because they are tuned to isolated frequencies. The acquisition for each nucleus shall adapt to an independent RF chain and a given coupling strength.
Parallel pulse compensation scheme
In the parallel NMR experiments, multiple channels deliver simultaneous RF pulses. Poor electrical isolation can cause coil coupling, i.e., the B1 field generated at one channel is sensed by neighboring samples, resulting in undesired manipulations of their spin states. In this sense, normal pulse calibration for a single sample fails in the parallel case, since the B1 field distortion comes from both the individual coil imperfection as well as from the pulse leakage from adjacent channels. To overcome imperfections in single coils, optimal control methods have been extensively employed to generate composite pulses, which exhibit high-performance transfer efficiency against field inhomogeneity and resonance offset49,50,51,52, and have been used to explore physical limits of excitation and inversion pulses53,54,55. In the parallel case, the advantages of optimal parallel transmission for compensation of patient-induced B1 inhomogeneity and reducing RF power in MRI have been demonstrated8, and the cooperative pulses for multiple-scan spectroscopy32 contribute similar inspiration. Utilizing the obtained B-map database, the concurrent optimization of pulses across multiple channels enhances robustness against load-induced B-field inhomogeneity, thereby improving single-sample excitation in MRI. However, in parallel NMR involving multiple and potentially distinct samples, the pulse applied to each channel must prioritize specific sample excitation while guarding against coupled radiation from neighboring channels. If the inter-channel coupling details are known, the simultaneous optimization can be transformed into channel-specific compensation. This approach may grant the signal-channel pulse greater flexibility in addressing field inhomogeneity and spin-spin interactions56,57,58. Here, we combine single-channel pulse optimization with the coupling information provided by RF modeling, to design the parallel pulse sequences.
From parallel detector modeling, we correlate the B1 field with the pulse sequence through two matrices, B1 = T ⋅ F ⋅ p, where p is the excitation at the T&Mnetwork terminal (a2 in Fig. 2c). The F and T matrix indicate a two-step linear supercombination. Specifically, the F transforms the excitation pulse to coil currents, so it represents a circuit-level combination. This combination is position-independent and convenient to be canceled out by pulse compensation. The T matrix transforms the coil currents to the B1 field, so it represents a field-level combination, which contributes to the direct field spill-over effect from one channel to another. In this way, we separate the mutual coupling into circuit-level combination and field-level combination. Generally, the RF coils including striplines and solenoids confine the B1 field at their center, so the outgoing radiation becomes very weak and makes a minimal contribution. Consequently, in weak coupling cases, the simultaneous optimization of parallel pulses incorporating the position-dependent T matrix can be avoided. The parallel pulse compensation is aimed to cancel out circuit-level combination.
Prior to the pulse compensation, we examine the impact of coupling on pulse efficiency, specifically the decrease in excitation fidelity under a range of coupling strengths. In Fig. 4a, by subjecting one 1H spin to a single 90° pulse in both channels, we observed a remarkable decrease in fidelity as the coupling strength increased, indicated by the close proximity of the two solenoid coils. Notably, we assigned the proton chemical shifts as -2 ppm for channel-1 and 5 ppm for channel-2, respectively. Consequently, a slight discrepancy between the two lines arose as a result of the distinct impact of the coupled pulse on each nucleus. Similarly, Fig. 4b presents the excitation fidelity for the optimized control pulse, which was specifically designed to exhibit stability in the presence of resonance offset and B1 field inhomogeneity. In the absence of coupling, the optimized control pulse achieved near-perfect fidelity. However, as the coupling strength increased, a remarkable decline in fidelity was observed, mirroring the behavior of the hard pulse case. Notably, Fig. 4b demonstrates that the fidelity degradation resulting from RF coupling exhibits variations between the two channels. This discrepancy is attributed to the disorganized pulse shape which is distinct for the two channels. In addition, the adverse effects of coupling are heightened with increasing pulse power, the functionality of the 10 kHz pulse is entirely compromised when operating in parallel with a coupling strength of 0.24, as detailed in Supplementary Fig. S8.
Due to its ability to precisely manipulate nuclear spins, the optimal control pulse offers exceptional performance. However, the RF coupling disrupts its waveform by inducing additional currents on the coil. Consequently, the pulse becomes less robust to resonance offset and field inhomogeneities, resulting in a decrease in transfer efficiency. To mitigate RF coupling through pulse compensation, the key concept is to create a cooperative pulse that integrates the optimal control pulse with a compensation term (refer to Fig. 5a). The cooperative pulse aims to exert an equivalent effect on the spin as the optimal control pulse does during individual excitation. This approach can be outlined in the following steps:
-
1.
Run optimal control for each channel individually, generating pulse sequence p0.
-
2.
Assume a compensation term δi for each channel, to satisfy the elimination condition, for which the following equation is established
-
3.
Determine the cooperative pulse p = p0 + δ by solving Eq. (3).
To give an example, in the first step, the single-channel optimal control considered transferring one 1H spin from Iz to Ix, accounting for resonance offset ( ± 2.5kHz bandwidth) and RF inhomogeneity determined by the imported B1 field. The B1 field, calculated at a reasonable excitation power, generated an average RF amplitude of 2kHz for 1H. In the second step, the compensation term δ was added to p0 to cancel the coupling effect, enabling the generation of coil currents consistent with individual excitation of the optimal control pulse. Fig. 5b illustrates the normalized excitation waveform for two-channel pulse compensation, displaying pulse sequences in terms of amplitudes and phase with 100 time slices and a duration of 2ms. Each element of p0 had a fixed amplitude of 1, whereas the amplitude and phase of p changed over time. Fig. 5
(c) presents the simulated transfer fidelity under three excitation conditions: applying the optimal control pulse individually without coupling (p0 - no couple), the optimal control pulse in parallel with coupling (p0 - coupled), and the cooperative pulse with coupling (p - coupled). When applying p0 in parallel, the coupling introduced B1 field distortion and led to a drop in transfer fidelity. However, the compensation of p accounted for coupling and achieved the same efficiency as individual excitation of p0. It’s important to note that this drop varied depending on the resonance offset and local B1 field. Additionally, due to the step-by-step optimization process of the initial guess for the optimal control pulse, which contained randomness, the efficiency drop could differ between the two channels, as observed in Fig. 4b.
Under a wide range of coupling strengths, as long as Eq. (3) remains solvable, a cooperative pulse can achieve comparable transfer efficiency relative to the case in which no coupling exists and optimal control pulses apply on isolated channels. The cooperative pulses maintain over 96% mean fidelity with coupling strengths equivalent to 0.09, 0.24, and 0.36 of 2 solenoid coils, provided in Supplementary Fig. S9. As a portion of the cooperative pulse energy is allocated to counteract coupling, its applicability may be constrained by available power, particularly in cases of strong coupling.
Parallel signal decomposition scheme
Coil coupling not only contributes to pulsed current leakage during the excitation stage but also causes FID cross-talk3 in the reception stage. Here we develop a post-processing method to split the composite signal at the output of the T&Mnetwork (Fig. 3b).
Consider synchronous parallel acquisition for n homonuclear channels. In this scenario, each coil records the FID from all samples, i.e., \({s}_{i}={{f}_{\!\!ii}}+{\sum}_{j\ne i} \, {{f}_{\!\!ij}}\), where fij is the FID on channel i from sample j. Eq. (2) indicates that each FID component is determined by the spin magnetization as well as the \({B}_{1}^{-}\) field. When the j-th sample obtains uniform magnetization, for example, Mj = Mx, the coupled signal fij could be given as Rij fjj, where Rij is the ratio of sensitivity. In this way, we reduce the unknowns and summarize that the received signal x can be expressed as (Supplementary Note 6):
where f is the signal vector consisting of the primary FIDs, i.e., \({{{{{{{\bf{f}}}}}}}}={( \, {{f}_{\!\!11}},{{f}_{\!\!22}},...,{{f}_{\!\!nn})}}^{T}\), and \({{{{{{{{\bf{n}}}}}}}}}^{{\prime} }\) is the coupled noise, \({{{{{{{\bf{{G}}}}}}}^{{\prime} }}}\) is a modified coupling matrix, in which \({G}_{ii}^{{\prime} }={G}_{ii},{G}_{ij}^{{\prime} }={G}_{ij}+{G}_{ii}{R}_{ij}\). The G is the reception gain and the Rij is the sensitivity ratio between the coil i and the coil j regarding the sample j. Hence, the primary FIDs can be recovered provided the coupling matrix is known. In a digital twin, the coupling matrix is calculated through EM simulation, in an experimental scenario, Eq. (4) could be solved by BSS59,60, these two aspects are discussed below.
Figure 6 shows the signal decomposition results for a 4-detector, 8-channel array, where the pairs of 1H and 13C channels were simultaneously excited and received, assuming perfect isolation between heteronuclear channels. For simplicity, only one resonance was simulated for each channel and was assigned a different chemical shift per channel. The B1 fields were simulated for 1H and 13C using the electrical parameters of water, i.e., assuming an aqueous sample. Figure 6a, c displays the coupled signal with interference. The interfering signals are relatively lower in the 13C channel due to a smaller coupling strength at the lower frequency. Figure 6b, d gives the decomposed signals, demonstrating complete splitting of the desired signals from the cross-talk component. As the modified coupling matrix originates from reception gain G and sensitivity ratio, the decomposition error was simulated regarding the random shifts of G and B1, see Supplementary Note 7.
Note that Eq. (4) expresses the received signals as a linear combination of the FIDs providing the reception gain is time-invariant, indicating that signal decomposition is a typical BSS problem under instantaneous linear mixtures60. First, the source signals from multiple samples are statistically independent, and second, we suppose they have different spectral content. The second-order blind identification (SOBI) algorithm61 is an exceptional and robust method and suitable in this case. This method is based on jointly diagonalizing a set of covariance matrices, which could be adapted to difficult contexts such as adverse SNR and sources with little spectral difference. In Fig. 7, we use the SOBI algorithm to decompose 1H spectra from a two-channel array, the original spectra and split spectra for each channel are given for comparison. The simulated data in Fig. 7a was generated using the RF chain model. In particular, both the FIDs and noise components of the original signals underwent reception coupling G. The coupled signals are fully filtered, indistinguishable from the noise. Figure 7b gives the decomposition results for experimental data obtained with 2 water samples, the primary peaks, i.e., 4 ppm and –1.8 ppm were distanced by applying a shimming gradient. Compared to the original spectrum (gray lines), the split one (colored lines) removed over 90% of the coupled components, while maintaining the individual signals. Note that the BSS for instantaneous mixtures returns an estimated demixing matrix, the recovered signal is given by multiplying this matrix with the coupled signal, so the lineshape of the spectrum is preserved. When handling the highly parallel probes, the spectral differences between signals from two channels may appear minor, especially amidst the complexities of spin systems, both SNR and linewidth can suffer due to decreased shimming quality. To investigate these factors, we present the figure of merit results for the SOBI method in the context of a large number of mutually uncorrelated sources. Note that when dealing with partially correlated source signals, the ability to separate them diminishes rapidly as the number of sources increases, as discussed in literature62. Each channel has a single harmonic wave described by \({s}_{m}=\exp (({{{{{{{\rm{i2\pi }}}}}}}} \, {{f}_{\!\!m}}-\alpha )t)\), where fm = m and α is the decay rate, the corresponding linewidth in Hz is α/π. Each source signal involves 5000 sampling points over 10 seconds, with a fixed SNR of 30 dB. The coupling matrix is randomly generated as an N by N complex matrix for N channels. The estimation error is calculated based on the mixing matrix criterion62.
where E is the identity matrix, \(\hat{{{{{{{{\bf{A}}}}}}}}}\) is the estimated coupling matrix in which the permutation and scale indeterminacies are removed, and A is the real coupling matrix. The calculated CA is shown in Fig. 8. Performance improves with more sources, indicating smaller estimation errors and robust separation for uncorrelated sources. Larger linewidth α results in larger CA, for example, with the linewidth decreasing from 0.4/π to 0.2/π, the estimation efficiency gets an improvement factor of 2 (3 dB). It implies the potential application in dense arrays with enhanced shimming quality. Furthermore, Supplementary Note 8 offers insight into spectral resolution at low SNR signals, indicating that two signals with a 3 Hz spectral difference can be discerned at SNR=30 dB, with a finer resolution of 1.5 Hz achievable at SNR=40 dB.
Pulse compensation experiment
The pulse compensation scheme was verified with two channels. Figure 9a illustrates the workflow for the pulse compensation scheme. Initially, the simulated magnetic field was imported into the optimal control module implemented in Spinach to generate the optimal control pulse. Subsequently, a calibration experiment was performed to generate the coupled time domain FID signals, which were then inputted into the BSS module to estimate the coupling matrix \(\hat{{{{{{{{\bf{F}}}}}}}}}\), an n by n matrix for an n-channel detector. Finally, the cooperative pulse was computed using the equation p = F−1Fdiagp0, where the vector p0 represents the multiple optimal control pulses, each of which is individually optimized.
Utilizing this workflow, we conducted the 1H NMR calibration experiment using two water samples, which were excited with hard pulses. To increase the separation between the two water peaks for easier separation, the global y-shim value was set to a large value of 150,000 on TopSpin, resulting in a spectral difference of 5.3 ppm, for example, water peaks are at –0.7 ppm and 4.6 ppm for detectors 1 and 2, respectively. Parallel FIDs were then acquired for signal decomposition and coupling matrix identification. In Fig. 9b, the coupled component in channel 1 was successfully removed, while the one in channel 2 at -0.7 ppm persisted, indicating a coupling estimation error. This error arises from two factors: Firstly, the shimming quality in channel 1, the peak exhibits a linewidth of 28 Hz along with a prolonged shoulder on its upfield. Secondly, it stems from suppressed coupling due to an impedance mismatch in channel 2, the coupling from channel 1 to channel 2 is approximately 1%. We subsequently calculated the cooperative pulse following the workflow in Fig. 9a, and tested their efficiency under parallel excitation to verify the pulse compensation scheme. Figure 9c presents 1H spectra of the two water samples, with the top and bottom plots representing channels 1 and 2, respectively. Gray lines depict spectra obtained with optimal control pulse in single-channel excitation, i.e., p0 - no couple, for reference. To compare excitation distortions, the three lines in each figure share the same phase correction parameters, optimized based on the lineshape of the reference line. Notably, the shimming was performed using global and local shimming sets, causing a 0.9 ppm shift between the water peaks. In channel 1, spectra for p0 - coupled show varying amplitude and phase changes due to coupling-induced distortion, while p—coupled yields similar signal characteristics to the reference, indicating successful restoration by the cooperative pulse. However, in channel 2, the lines nearly coincide, suggesting weaker coupling, the strength of 1% was observed in Fig. 9b. Here, the error in coupling estimation led to ineffective compensation by the cooperative pulse. Figure 9d displays spectra of L-alanine and L-valine samples excited with the same optimal control pulses, giving a similar distortion as in Fig. 9c. The closely spaced peaks from L-alanine and L-valine are invisible considering the shimming quality, the peaks in the downfield arise from residual water in D2O solvent and the peaks at 0 ppm represent TSP.
Performance of the digital twin
Wall clock time
The wall clock time for the parallel NMR simulation is provided in Table 1. These simulations were carried out on a PC equipped with an AMD Ryzen 5950X 16-core processor, running at a base frequency of 3.40 GHz, and equipped with 64 GB of RAM. Each channel’s sample was divided into 72 voxels. In the spin dynamics simulation within each channel, the calculation considered a single 1H spin, subjected to a hard pulse and acquisition experiment at a 500 MHz magnet. As the number of channels increases, the RF simulation needs to sweep each lumped port to calculate the full n by n scattering matrix, which contributes to the most time-consuming part. The wall clock time of optimal control was given in Supplementary Note 9. For the wall clock time of large-scale spin system simulation, one may refer to the Spinach literature63.
Robustness to poor SNR
Using the test flow in Fig. 10a, the robustness of signal decomposition and pulse compensation with poor SNR signals were examined through a 2-channel simulation. The EM simulation incorporates a 2-solenoid array with a distance of 1.25 mm. The B0 and B1 fields were processed and imported into Spinach. The spin dynamics simulation involved a 1H excitation and acquisition experiment, with the spin system for each channel randomly generated, including varying numbers of spins and their chemical shifts.
In Fig. 10b, where the average SNR is 10 dB and minimal Δf = 0.02 ppm, the cooperative pulse’s fidelity exhibits mean values of 0.988 for channel-1 and 0.987 for channel-2, achieving a recovery of 99.3% fidelity from the optimal control pulse. Figure 10c showcases the robustness of CA concerning SNR, with CA converging to –17 dB (2%) as SNR increases to 50 dB. Notably, the source signals are weakly correlated, given the presence of cross-coupling between coils and samples during the reception, which causes an additional error. The 2% decomposition error means removing over 95% of coupled components and achieving a similar fidelity level for the cooperative pulse compared to the optimal control pulse, as shown in Fig. 10d.
Discussion
We built a digital twin to model parallel NMR experiments, creating a digital environment from which pulse sequence optimization under parallel measurement can be performed. Starting from electromagnetic simulation, the calculated B0 and B1 fields were imported to the Spinach package for spin dynamics simulation, and the FIDs were extracted for signal chain calculation. Based on the model, we adapted cooperative pulses to cancel the inter-coil coupling effects in the excitation stage. In this scheme, pulse optimization only executes on single channels individually, followed by a forward pulse compensation, which can be a potential approach for parallel pulse sequence design. The optimization procedure avoids including inter-channel coupling and allocates its degree of freedom to address specific samples, making this scheme efficient for deployment in highly parallel detectors, in which the B1 field is confined to the coils. In the future, one may also design NMR coil arrays in which the field-level combination becomes notable. In such scenarios, the pulse compensation based on the coupling matrix may not suffice. Exploring parallel transmit design, as extensively addressed in MRI20,22,64, could be beneficial.
The cable connections between the probe and spectrometer exert a substantial impact on coupling strength, a factor not incorporated into the simulation model. Fortunately, parallel excitation and reception share the same RF chain and accordingly, the same coupling matrix, i.e., Gnorm ≈ Fnorm. We applied the BSS method to split the parallel signals and identify the coupling matrix. It is worth conducting a calibration experiment using samples with strong and distinct spectra to characterize the parallel system, and using this information for the general samples.
Since the SOBI algorithm relies on distinct spectral features from each source, the splitting efficiency drastically decreases when the coupled components are located on the shoulders of the primary peaks, as indicated previously, coupling estimation errors escalate with increasing linewidth. Therefore, there is a demand for more robust post-processing methods to facilitate general signal decomposition. These methods should be capable of managing source signals with dense peaks and broad shoulders, as well as naturally correlated source signals. The concepts developed for DOSY may offer avenues in this regard65,66,67. In heteronuclear experiments, such as HSQC, separate pulse compensation for each nucleus is anticipated, due to the frequency-dependent coupling.
This work focused on addressing RF coupling in parallel NMR. In many routine experiments, pulsed field gradients are crucial for the selection of coherence transfer pathways. To generate the required gradient field at each detection site, local gradient coils are utilized in parallel NMR probes. However, this approach can introduce gradient spillover or coupling. Therefore, further research is needed to investigate and address gradient coupling.
Methods
Electromagnetic simulations
The simulation scenario includes a solenoid array with a T&Mnetwork. The EM simulation was conducted with COMSOL MultiPhysics 6.1 and cross-checked with CST Studio Suite 2022, considering the coil array with water samples inserted. S-parameters were extracted from COMSOL to Matlab R2023a to design a virtual T&Mnetwork and build the RF chain model. Spin dynamics calculations and single-channel optimal control were implemented with the Spinach package v2.840, the details were provided in Supplementary Notes 10 and 11. For easy reading, the physical variables concerning the EM simulation and spin dynamics calculations were concluded in Supplementary Note 12.
Parallel signal acquisition experiments
The parallel signal acquisition experiments (Fig. 7b) was carried out using a 15.2 T Bruker NMR system and a custom-built two-stripline probe68. Each RF coil was deployed with local shim coils powered by an external DC source to improve B0 homogeneity, an x-gradient was applied to separate the two channels via the global shim (see Supplementary Fig. S17). For synchronous single pulse 1H acquisition, the spectra consisted of 64 scans, and 13157 data points were collected for each scan with 20 ppm spectral width.
Parallel pulse compensation experiments
The 2-channel pulse compensation was conducted using an 11.74 T Bruker NMR system equipped with a 4-stripline probe provided by Voxalytic GmbH, with only channels 1 and 2 utilized. The 2 channels were individually tuned to 500 MHz with a tune and match (T&M) box equipped for the probe, the T&M results were given in Supplementary Fig. S18. The local shimming was powered by a 28-channel shim current source68. Pulse calibration was completed within the single channel nutation experiment. Specifically, the nutation experiments were conducted on channel-1 and channel-2 separately to determine the RF amplitude at the suggested maximum power level, 20 W. Then the power level was scaled down to deliver an RF amplitude of 30 kHz, i.e., 8.3 μs of 90 degree pulse, supposing that B1 amplitude is proportional to \(\sqrt{P}\) when the power level located in the linear range of the amplifier. The optimal control pulse has a nominal RF amplitude of 30 kHz, lasts 8 ms, and comprises 4000 time bins. Two channels share the same acquisition parameters, including 3072 data points, a relaxation delay of 5 seconds for water and 10 seconds for L-alanine and L-valine, and a sweep width of 20 ppm for 1H acquisition. A total of 64 scans were recorded, with a receiver gain set to 10 for water and 18 for L-alanine and L-valine.
Sample preparation
The distilled water was prepared for the calibration experiment (Fig. 9b). A solution of 250 mM L-valine and L-alanine (Sigma-Aldrich) samples were prepared in D2O (99.9%, Sigma-Aldrich) with 10 mM 3-(Trimethylsilyl)propionic-2,2,3,3-d4 acid sodium salt (TSP) as reference compound for verifying the pulse compensation. The chemical shifts of L-alanine and L-valine are given in Supplementary Note 14. The samples were filled into the syringes and manually pumped into the individual fluidic chamber of the dedicated detector.
Data availability
The data supporting the findings of this work are available from the corresponding authors upon reasonable request.
Code availability
The Multiphysics simulation code is accessible in the repository https://github.com/kikioh/digital-parallel-NMR.
References
MacNamara, E., Hou, T., Fisher, G., Williams, S. & Raftery, D. Multiplex sample NMR: An approach to high-throughput NMR using a parallel coil probe. Anal. Chim. Acta 397, 9–16 (1999).
Wang, H., Ciobanu, L., Edison, A. S. & Webb, A. G. An eight-coil high-frequency probehead design for high-throughput nuclear magnetic resonance spectroscopy. J. Magn. Reson. 170, 206–12 (2004).
Li, Y., Wolters, A. M., Malawey, P. V., Sweedler, J. V. & Webb, A. G. Multiple solenoidal microcoil probes for high-sensitivity, high-throughput nuclear magnetic resonance spectroscopy. Anal. Chem. 71, 4815–4820 (1999).
Lei, K.-M. et al. Portable NMR with parallelism. Anal. Chem. 92, 2112–2120 (2020).
Ciobanu, L., Jayawickrama, D. A., Zhang, X., Webb, A. G. & Sweedler, J. V. Measuring reaction kinetics by using multiple microcoil NMR spectroscopy. Angew. Chem. Int. Ed. 42, 4669–4672 (2003).
Kupče, Ē., Frydman, L., Webb, A. G., Yong, J. R. J. & Claridge, T. D. W. Parallel nuclear magnetic resonance spectroscopy. Nat. Rev. Methods Primers 1, 1–23 (2021).
Cheng, Y.-T., Jouda, M. & Korvink, J. Sample-centred shimming enables independent parallel NMR detection. Sci. Rep. 12, 14149 (2022).
Katscher, U. & Börnert, P. Parallel RF transmission in MRI. NMR Biomed. 19, 393–400 (2006).
Pruessmann, K. P. Encoding and reconstruction in parallel MRI. NMR Biomed. 19, 288–299 (2006).
Roemer, P. B., Edelstein, W. A., Hayes, C. E., Souza, S. P. & Mueller, O. M. The NMR phased array. Magn. Reson. Med. 16, 192–225 (1990).
Hutchinson, M. & Raff, U. Fast MRI data acquisition using multiple detectors. Magn. Reson. Med. 6, 87–91 (1988).
Sotgiu, A. & Hyde, J. S. High-order coils as transmitters for NMR imaging. Magn. Reson. Med. 3, 55–62 (1986).
Ibrahim, T. S., Lee, R., Baertlein, B. A., Kangarlu, A. & Robitaille, P.-M. L. Application of finite difference time domain method for the design of birdcage RF head coils using multi-port excitations. Magn. Reson. Imag. 18, 733–742 (2000).
Zhu, Y. Parallel excitation with an array of transmit coils. Magn. Reson. Med. 51, 775–784 (2004).
Adriany, G. et al. Transmit and receive transmission line arrays for 7 Tesla parallel imaging. Magn. Reson. Med. 53, 434–445 (2005).
Katscher, U., Börnert, P., Leussler, C. & van den Brink, J. S. Transmit SENSE. Magn. Reson. Med. 49, 144–150 (2003).
Cloos, M. A. et al. Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7T. NeuroImage 62, 2140–2150 (2012).
Deniz, C. M., Alon, L., Brown, R. & Zhu, Y. Subject- and resource-specific monitoring and proactive management of parallel radiofrequency transmission. Magn. Reson. Med. 76, 20–31 (2016).
Massire, A. et al. Parallel-transmission-enabled three-dimensional T2-weighted imaging of the human brain at 7 Tesla. Magn. Reson. Med. 73, 2195–2203 (2015).
Gras, V., Vignaud, A., Amadon, A., Le Bihan, D. & Boulant, N. Universal pulses: A new concept for calibration-free parallel transmission. Magn. Reson. Med. 77, 635–643 (2017).
Gras, V. et al. Design of universal parallel-transmit refocusing kT-point pulses and application to 3D T2-weighted imaging at 7T. Magn. Reson. Med. 80, 53–65 (2018).
Le Ster, C., Mauconduit, F., Massire, A., Boulant, N. & Gras, V. Standardized universal pulse: A fast RF calibration approach to improve flip angle accuracy in parallel transmission. Magn. Reson. Med. 87, 2839–2850 (2022).
Van Damme, L., Mauconduit, F., Chambrion, T., Boulant, N. & Gras, V. Universal nonselective excitation and refocusing pulses with improved robustness to off-resonance for magnetic resonance imaging at 7 Tesla with parallel transmission. Magn. Reson. Med. 85, 678–693 (2021).
Adriany, G. et al. A 32-channel lattice transmission line array for parallel transmit and receive MRI at 7 tesla. Magn. Reson. Med. 63, 1478–1485 (2010).
Vernickel, P. et al. Eight-channel transmit/receive body MRI coil at 3T. Magn. Reson. Med. 58, 381–389 (2007).
Avdievich, N. I., Pan, J. W. & Hetherington, H. P. Resonant inductive decoupling (RID) for transceiver arrays to compensate for both reactive and resistive components of the mutual impedance. NMR Biomed. 26, 1547–1554 (2013).
Moody, K. L. et al. An eight-channel T/R head coil for parallel transmit MRI at 3T using ultra-low output impedance amplifiers. J. Magn. Reson. 246, 62–68 (2014).
Choi, C.-H. et al. A review of parallel transmit arrays for ultra-high field mr imaging. IEEE Rev. Biomed. Eng. 17, 351–368 (2024).
Galante, A., Fantasia, M., Sebastiani, P., Sotgiu, A. & Alecci, M. Analytical RF coils decoupling: A theoretical approach for parallel transmission. Proc. Intl. Soc. Mag. Reson. Med. 25 https://cds.ismrm.org/protected/17MProceedings/PDFfiles/4302.html (2017).
Vernickel, P., Findeklee, C., Eichmann, J. & Grässlin, I. Active digital decoupling for multi channel transmit MRI systems. Proc. Intl. Soc. Mag. Reson. Med. 15, 170 (2007).
Carlomagno, T., Luy, B. & Glaser, S. J. “Kin” HEHAHA sequences, heteronuclear Hartmann–Hahn transfer with different bandwidths for spins I and S. J. Magn. Reson. 126, 110–119 (1997).
Braun, M. & Glaser, S. J. Cooperative pulses. J. Magn. Reson. 207, 114–123 (2010).
Braun, M. & Glaser, S. J. Concurrently optimized cooperative pulses in robust quantum control: Application to broadband ramsey-type pulse sequence elements. New J. Phys. 16, 115002 (2014).
Wang, T., Yang, F. & Yang, J. Convolutive transfer function-based multichannel nonnegative matrix factorization for overdetermined blind source separation. IEEE ACM Trans. Audio Speech Lang. Process. 30, 802–815 (2022).
Chang, S. et al. An advanced scheme for range ambiguity suppression of spaceborne sar based on blind source separation. IEEE Trans. Geosci. Remote Sens. 60, 1–12 (2022).
Ye, C., Toyoda, K. & Ohtsuki, T. Blind source separation on non-contact heartbeat detection by non-negative matrix factorization algorithms. IEEE Trans. Biomed. Eng. 67, 482–494 (2020).
Zarzoso, V. & Nandi, A. Blind separation of independent sources for virtually any source probability density function. IEEE Trans. Signal Process. 47, 2419–2432 (1999).
Dumont, R. S., Jain, S. & Bain, A. Simulation of many-spin system dynamics via sparse matrix methodology. J. Chem. Phys. 106, 5928–5936 (1997).
Horneff, A. et al. An EM simulation-based design flow for custom-built MR coils incorporating signal and noise. IEEE Trans. Med. Imag. 37, 527–535 (2018).
Hogben, H. J., Krzystyniak, M., Charnock, G. T., Hore, P. J. & Kuprov, I. Spinach–a software library for simulation of spin dynamics in large spin systems. J. Magn. Reson. 208, 179–194 (2011).
Pozar, D. M. Microwave Engineering, 4th Edition, 178–180 (Wiley, 2011).
Lagarias, J. C., Reeds, J. A., Wright, M. H. & Wright, P. E. Convergence properties of the nelder–mead simplex method in low dimensions. SIAM J. Optim. 9, 112–147 (1998).
Majewski, K. Simultaneous optimization of radio frequency and gradient waveforms with exact hessians and slew rate constraints applied to kT-point excitation. J. Magn. Reson. 326, 106941 (2021).
Keifer, P. A. 90° pulse width calibrations: How to read a pulse width array. Concepts Magn. Reson. 11, 165–180 (1999).
Rasulov, U., Acharya, A., Carravetta, M., Mathies, G. & Kuprov, I. Simulation and design of shaped pulses beyond the piecewise-constant approximation. J. Magn. Reson. 353, 107478 (2023).
Tošner, Z. et al. Overcoming volume selectivity of dipolar recoupling in biological solid-state NMR spectroscopy. Angew. Chem. Int. Ed. 57, 14514–14518 (2018).
Kuprov, I. Large-scale NMR simulations in liquid state: A tutorial. Magn. Reson. Chem. 56, 415–437 (2018).
Hoult, D. I. The principle of reciprocity in signal strength calculations—a mathematical guide. Concepts Magn. Reson. 12, 173–187 (2000).
Khaneja, N., Reiss, T., Kehlet, C., Schulte-Herbruggen, T. & Glaser, S. J. Optimal control of coupled spin dynamics: Design of NMR pulse sequences by gradient ascent algorithms. J. Magn. Reson. 172, 296–305 (2005).
Kobzar, K., Luy, B., Khaneja, N. & Glaser, S. J. Pattern pulses: Design of arbitrary excitation profiles as a function of pulse amplitude and offset. J. Magn. Reson. 173, 229–235 (2005).
Skinner, T. E., Reiss, T. O., Luy, B., Khaneja, N. & Glaser, S. J. Application of optimal control theory to the design of broadband excitation pulses for high-resolution NMR. J. Magn. Reson. 163, 8–15 (2003).
Skinner, T. E., Reiss, T. O., Luy, B., Khaneja, N. & Glaser, S. J. Reducing the duration of broadband excitation pulses using optimal control with limited RF amplitude. J. Magn. Reson. 167, 68–74 (2004).
Kobzar, K., Skinner, T. E., Khaneja, N., Glaser, S. J. & Luy, B. Exploring the limits of broadband excitation and inversion pulses. J. Magn. Reson. 170, 236–243 (2004).
Kobzar, K., Skinner, T. E., Khaneja, N., Glaser, S. J. & Luy, B. Exploring the limits of broadband excitation and inversion: II. RF-power optimized pulses. J. Magn. Reson. 194, 58–66 (2008).
Kobzar, K., Ehni, S., Skinner, T. E., Glaser, S. J. & Luy, B. Exploring the limits of broadband 90° and 180° universal rotation pulses. J. Magn. Reson. 225, 142–160 (2012).
Görling, B., Bräse, S. & Luy, B. HR-HSBC: Measuring heteronuclear one-bond couplings with enhanced resolution. Magn. Reson. Chem. 50, S58–S62 (2012).
Ehni, S. & Luy, B. A systematic approach for optimizing the robustness of pulse sequence elements with respect to couplings, offsets, and B1-field inhomogeneities (COB). Magn. Reson. Chem. 50, S63–S72 (2012).
Ehni, S. & Luy, B. BEBEtr and BUBI: J-compensated concurrent shaped pulses for 1H–13C experiments. J. Magn. Reson. 232, 7–17 (2013).
Tong, L., Liu, R.-w, Soon, V. & Huang, Y.-F. Indeterminacy and identifiability of blind identification. IEEE Trans. Circuits Syst. 38, 499–509 (1991).
Jutten, C. & Herault, J. Blind separation of sources, part I: An adaptive algorithm based on neuromimetic architecture. Signal Processing 24, 1–10 (1991).
Belouchrani, A., Abed-Meraim, K., Cardoso, J.-F. & Moulines, E. A blind source separation technique using second-order statistics. IEEE Trans. Signal Process. 45, 434–444 (1997).
Bobin, J., Rapin, J., Larue, A. & Starck, J.-L. Sparsity and adaptivity for the blind separation of partially correlated sources. IEEE Trans. Signal Process. 63, 1199–1213 (2015).
Edwards, L. J. & Kuprov, I. Parallel density matrix propagation in spin dynamics simulations. J. Chem. Phys. 136, 044108 (2012).
Tomi-Tricot, R. et al. Smartpulse, a machine learning approach for calibration-free dynamic RF shimming: Preliminary study in a clinical environment. Magn. Reson. Med. 82, 2016–2031 (2019).
Nilsson, M. & Morris, G. A. Improved DECRA processing of DOSY data: correcting for non-uniform field gradients. Magn. Reson. Chem. 45, 656–660 (2007).
Nilsson, M. & Morris, G. A. Speedy component resolution: An improved tool for processing diffusion-ordered spectroscopy data. Anal. Chem. 80, 3777–3782 (2008).
Lin, E., Yang, Y., Huang, Y. & Chen, Z. High-resolution reconstruction for diffusion-ordered NMR spectroscopy. Anal. Chem. 92, 634–639 (2020).
Becker, M. et al. Artificial intelligence-driven shimming for parallel high field nuclear magnetic resonance. Sci. Rep. 13, 17983 (2023).
Acknowledgements
We are supported by the Joint Lab Virtual Materials Design (JLVMD) of the Helmholtz Association, Germany. J.G.K. acknowledges support from the ERC-SyG (HiSCORE, 951459). Y.C., J.G.K., and M.J. were supported by the DFG under contract AN 984/10-1. N.M., M.J., B.L., and J.G.K. acknowledge partial support from CRC 1527 HyPERiON. The authors acknowledge the support of the Helmholtz Society, both through the program Materials Systems Engineering, as well as for financing the 15.2 T magnet system. We appreciate the suggestions on Spinach simulations provided by Prof. Ilya Kuprov.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
N.M. and J.G.K. conceived of the initial digital twin. M.H. performed the simulations with input from all coauthors. Y.C. and M.J. designed and fabricated the 2-stripline probe. M.H., D.F., N.M., Y.C., and D.B. conducted the experiments. M.H., D.F., N.M., and Y.C. drafted the manuscript. All authors reviewed and refined the manuscript. J.G.K., B.L, and N.M. provided supervision and secured the funding.
Corresponding authors
Ethics declarations
Competing interests
J.G.K. is a shareholder of Voxalytic GmbH, a spinoff company that produces and markets microscale NMR devices. The other authors declare no competing interests.
Peer review
Peer review information
Communications Engineering thanks Soumyajit Mandal, Angelo Galante and the other, anonymous, reviewer(s) for their contribution to the peer review of this work. Primary Handling Editors: Anastasiia Vasylchenkova and Rosamund Daw. A peer review file is available.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
He, M., Faderl, D., MacKinnon, N. et al. A digital twin for parallel liquid-state nuclear magnetic resonance spectroscopy. Commun Eng 3, 90 (2024). https://doi.org/10.1038/s44172-024-00233-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s44172-024-00233-0