Abstract
Anatomic variations, calcified, tortuous, angulated lesions, and lack of support to increase the complexity of transradial intervention (TRI). Guidezilla is a mother-and-child catheter enabling increased support during complex interventions. As there are few published reports of its use, we describe our experience using this device to assist TRI in Chinese patients. The aim of this study was to investigate the efficacy and safety of the Guidezilla guide extension catheter in complex coronary interventions. Thirty-two patients’ clinical characteristics, angiographic details, and in-hospital outcome retrospectively collected between June 2015 and August 2017. Patients were 59.44 ± 10.48 years of age and 26 (81%) were men. The most frequent target vessels were the RCA (34%) and LAD (31%), patients had complex type C (53%) or B (47%) lesions, severely tortuous (41%) and angulated (22%).With the use of Guidezilla, technique success was 100%, and procedural success was 94%. The mean diameter of the deployed stents was 2.97 ± 0.37 mm, and the length was 27.19 ± 8.14 mm. The estimated mean distance of Guidezilla intubation into the target vessel was 7.66 ± 2.29 cm.The Guidezilla catheter extension safely facilitated successful completion of TRI in complex coronary artery lesions. This device can help interventionalists successfully perform difficult procedures.
Similar content being viewed by others
Introduction
Transradial intervention (TRI) is often chosen for percutaneous coronary interventional (PCI) procedures because of fewer complications, earlier patient mobilization, and improved clinical outcomes compared with the transfemoral approach1,2,3,4,5. However, in many patients, the size of the radial artery limits the guide catheter (GC) to a size no larger than 7F. Also procedures involving complex coronary artery anatomies, extreme vessel tortuosity, calcification, angulation, and chronic total occlusions (CTO) are often time consuming and challenging6,7. TRI procedures can fail because of arterial spasms, failure to puncture the access site, failure to cannulate the lesion vessels, and lack of adequate guide support. Most of these obstacles can be dealt with by certain tips and tricks. One solution is use of a guide catheter extension system (GCES).
Commercially available GCES include the Proxis device (St Jude Medical, St Paul, MN, USA), Heartrail II catheter (Terumo, Tokyo Japan), GuideLiner (Vascular Solutions, Minneapolis, MN, USA), and Guidezilla (Boston Scientific, Natick, MA, USA). The Proxis device and the Heartrial II catheter are 120 cm catheters that are introduced into the mother guide and require removing the Y-connector8,9,10,11. The GuideLiner catheter is a rapid GCES with a long flexible tubular end that can be deeply advanced into lesions, providing support without the need to disconnect it from the mother guide12,13,14. The Guidezilla GCES is a unique, rapid exchange mother-and-child catheter. For difficult or complex procedures such as anchor balloons, deep seating, buddy wire technique, super-stiff guide wires, or GC change to a more supportive configuration15,16,17, the Guidezilla is simpler, quicker, more effective especially for transradial PCI, where the GC lacks support. Guidezilla, has a 25 cm catheter with a polymer proximal collar, and a 1.45 mm inner-diameter that has more room to deliver intervention device, and the 1.68 mm outer diameter reduces GC interference, which is smaller than the GuideLiner (Fig. 1). The braid design facilitates additional back-up support for complex or tortuous vessels, calcification, and angulated coronary anatomy lesions. The Food and Drug Administration approved Guidezilla in July 201318, and the a limited number of published reports of its use for complex PCI found it to be noninferior to the other three guide extension catheters.
Thus, the purpose of this study was to describe our preliminary clinical experience with Guidezilla for back-up support and balloon and stent delivery in the treatment of complex coronary lesions via the transradial approach report at a single center in China.
Methods
Study population
Thirty-two consecutive patients with coronary artery disease (CAD) and TRI procedures utilize Guidezilla at between June 2015 and August 2017. All had undergone angiography and all had failed conventional techniques. The study was approved by the ethics committee of West China Hospital and complied with ethical principles of the Declaration of Helsinki. All patients gave written informed consent before enrollment.
Baseline characteristics
Patient demographics, target vessel, ACC/AHA lesion type19, lesion complexity, access site, procedure success and complications, and in-hospital outcome were recorded. Data collected and reviewed by two cardiologists experienced in analysis of quantitative coronary arteriography. Indications for use of the Guidezilla GCES included: (1) anomalous or angulated take-off of native coronary arteries, (2) proximal tortuous vessels, (3) extreme calcification or long lesions, (4) for back-up support of delivery of balloons, stents, rotational atherectomy or aspiration devices, (5) or failure with other techniques. Success was defined as stent placement in the target lesion with residual stenosis of less than 20% and TIMI 3 flow. Procedural success was defined as technique success without major Guidezilla-related complications. Deep intubation was defined as depth more than 2 cm.
Interventional procedures
Transradial PCI techniques with Guidezilla were performed by experienced interventionalists following standard clinical protocols with right radial artery access and a 6F GC and radial sheath (Terumo, Japan). Dual antiplatelet drugs were administered orally before performing PCI. Intra-arterial nitroglycerin (200 µg) and unfractionated heparin (70–100U/kg) were administered during the PCI procedure after achieving arterial access. The choice of other drugs, intervention approach, equipment, and technique was also at the operator’s discretion. Selective coronary artery angiography (CAG) was performed using diagnostic 5F multifunctional catheters (Terumo, Japan). The 6F radial sheath was removed immediately after the PCI procedure, and a radial compression bandage was used to prevent bleeding of the access puncture site. The bandage was removed after 6 h, provided no bleeding occurred.
The Guidezilla catheter and its use
We used the 6F version Guidezilla GCES that has two radiopaque marker bands to increase visibility and facilitate accurate placement and positioning in the aorto-ostium of the coronary arteries. The radiopaque safety tip is soft and flexible, which reduces the risk of vessel damage and complications. A hydrophilic surface coating provides a smooth finish that reduces friction. The procedures began by positioning the GC and advancing the guide wire (GW) across the target lesion. The Guidezilla was then advanced over the GW through the hemostatic valve of the Y-connector to intubate the target lesion. When the Guidezilla was in position, balloons, and stents were delivered over the same initial GW. This technique permitted deep intubation with back-up support and great coaxial alignment. In patients where a kissing balloon was needed, a wire exchange was performed followed by balloon dilation of the side branch through the stent struts. Before final inflation of the kissing balloon could be performed, the Guidezilla had to be removed.
Statistical analysis
This was not a comparative study. Continuous variables were expressed as means and standard deviation (SD) and categorical variables were expressed as number and percentage (%). Data analysis was performed with SPSS software version 19.0 (IBM Corp., Armonk, NY, USA). All authors take responsibility for all aspects of the reliability, interpretation, and freedom from bias of the data.
Results
Study population and baseline characteristics
A total of 32 consecutive procedures were performed using Guidezilla. Baseline clinical characteristics of the patients are presented in Table 1. They were 59.44 ± 10.48 years of age, 26 (81%) were men, and the majority had risk factors, including hypertension (75%), hypercholesterolemia (34%), current smoking (59%), and diabetes mellitus (47%). About 31%, 47%, 28% had previous coronary revascularization, CAD and myocardial infarction (MI), respectively. Fifty-nine percent of the patients presented with unstable angina pectoris.
Interventional procedures and indication for Guidezilla
Guidezilla GCES only through TRI access was used in 30 of the 32 procedures (94%). Both through the TRI and femoral access was used in 2 procedures (6%), because the 2 cases with chronic total occlusion (CTO), which need both transradial and transfemoral approaches to achieve contralateral injection or retrograde manipulation.
The target vessel was the RCA (34%) and LAD (31%). Patients had complex type C (53%) or B (47%) lesions, severely tortuous (41%) and angulated (22%). Details of the procedures are shown in Tables 2 and 3. The major finding was that severe tortuosity (41%) and angulation (22%) predicted the need for Guidezilla support, suggesting that vessel characteristics could be the primary consideration for use of active guide support in TRI procedures. Technique success was achieved with this GCES technique in all 32 procedures (100%); procedural success was achieved in 30 of 32 (94%). In one procedure, proximal dissection of the RCA occurred with high-pressure balloon dilation, which was not related to the Guidezilla. Stent stripping from the target lesions occurred in two patients. The mean diameter of the stents deployed was 2.97 ± 0.37 mm, the mean length was 27.19 ± 8.14 mm. The estimated mean distance of Guidezilla intubation into the target vessel was 7.66 ± 2.29 cm, and the intubation distance was ≥10 cm in eleven patients (34%). Representative images are taken during a procedure in a patient where initially it was not possible to deliver any equipment, but did succeed with the use of a Guidezilla catheter. In one elder patient CAG showed a left and right coronary common opening, left and right coronary arteries with severely tortuous, long stenotic lesions, calcification, and angulation successful treatment of coronary lesions with Guidezilla (Fig. 2).
Discussion
To the best of our knowledge, this is the first report of the successful use of the Guidezilla GCES for highly complex procedures in a series of 32 patients in China. The results show that the Guidezilla was a useful and safe adjunct tool during challenging TRI procedures. Previous reports described similar techniques in smaller patient series20,21. Using a Guidezilla GCES within a 6F GC, we performed interventions easily without the possibility of losing access to the coronary arteries while changing out the GC. TRI with complex coronary anatomy is often challenging and time consuming. Thus, proper PCI strategy with proper coronary intervention devices is crucial to ensure the success of the procedure. The PCI procedural success rate with the use of GCES ranges from 93% to 98%22,23. Farooq et al. demonstrated that GCES may allow deeper intubation of the GC, referred to as “rail-roading”14. Takahashi et al. noted that use of GCES provides substantial improvement in back-up support for complex coronary interventions24. The GuideLiner GCES has been used for selective contrast injection, providing better visualization of target lesions with smaller amounts of contrast, which is described in detail in the Twente GuideLiner registry22. Currently, four GCES devices are commercially available, the Proxis, Heartrail II, GuideLiner, and Guidezilla. The Guidezilla became available in China only recently. In our series, the main indications for the Guidezilla GCES were: (1) anomalous or angulated take-off of native coronaries; (2) proximal tortuous vessels; (3) extreme calcification or long lesions; (4) as back-up support for delivery of balloons, stents, and other equipment such as rotational atherectomy or aspiration devices; or (5) failure using other techniques. In our example case, we could not initially to deliver any equipment, and the stent could not pass through the RCA stenosis because of tortuosity, calcification, and angulation. A 6F Guidezilla was successfully introduced into the RCA, and a 2.5 mm anchor balloon was introduced into the distal LCX without difficulty using the mother-and-child maneuver with the Guidezilla catheter to achieve a successful result. We recommend limiting the extension of the Guidezilla catheter to <15 cm beyond the tip of the GC, in order to avoid complication, operator should pay attention not to extend the whole segment in the target vessel. Guidezilla catheter segment length is 25 mm, the manufacturer’s instructions suggested extension of the catheter out of the guide catheter less than 15 cm.Otherwise, it is possible that the mother catheter and child catheter lose the coaxiality and impede withdrawal of the device. In present study, the longest intubation depth of Guidezilla into the target vessel was 12 cm as further distal advancement of the Guidezilla device could cause the entire guide segment to track outside of the guide catheter and impede withdrawal of the device, and it is important to maintain coaxial alignment. For patients with renal insufficiency and the expectation of selective angiography (e.g., in a dominant LCA system), a Guidezilla GCES may be considered as a first choice. The patient described in the present study had a baseline creatinine of 204 µmol/L with compromised renal function. PCI was completed using less than 100cc of contrast, which minimizing contrast load and decreased the possibility of contrast-associated nephropathy. The patient was discharged 3 days later and remained asymptomatic at 1-year follow up. This is in line with a report by Tunuguntla et al.25 on the use of a GuideLiner catheter to minimize contrast during PCI in a patient with chronic kidney disease (CKD).Reports of complications associated with the Guidezilla GCES are limited. We experienced coronary stent stripping in two procedures (6%) and proximal dissection of the RCA in one (3%). Stent stripping from its delivery balloon occurred on the introduction into the proximal Guidezilla catheter edge. The same event has been reported by Waggoner et al.21, who described the proximal junction transition as the site of stent stripping in their case. The stainless steel collar of the Guidezilla is not as flexible as the polymer collar of the GuideLiner, so passing through the acute angle of the aortic arch caused the stent to encounter the proximal edge of the stainless steel collar during transition to the GCES. This appears to work against introducing stents into the device when it is within an acute angulation of a vessel. In our experience the solution was to remove the Guidezilla before pulling back an unimplanted stent. Farooq et al.14, reported that there was a small risk of damage to large or bulky stents as they enter the collar. They advised using low-profile stents with the GCES, avoiding ones than 4 mm in diameter.
One of our patients experienced coronary dissection because of high pressure balloon dilation, which was not related to the Guidezilla use. Similar complications reported during GCES procedures include stent deformation or dislodgement on withdrawal, coronary artery ischemia or dissection, pressure damping, and balloon kinking26,27,28,29,30,31. In difficult lesions, compared with other techniques, such as anchoring balloons, deep seating, buddy wires, super-stiffer wires, or GC change to a more supportive configuration15,16,17, the Guidezilla is simple, quick, and effective. It is especially helpful in transradial PCI where GC lack passive support, or in complex procedures.
Limitations
Our data was retrospective and obtained at a study single center, and the study sample was small, and accordingly native bias might have affected the accuracy and efficacy of our statistical analysis. However, this trial implemented rigorous inclusion criteria to reduce the risk of confounding bias. So, larger studies will be needed to further investigate our findings, meanwhile, we are expanding the sample size and performing a follow-up of these patients. Design modification of future GCES devices, especially the steel collar, may minimize the risk of stent stripping that we observed. Prospective randomized trials of the efficacy and safety of GCES techniques are warranted.
Conclusion
The Guidezilla GCES safely facilitated successful completion of TRI in complex coronary artery lesions. Guidezilla use was associated with a highly success rate with no major complications. We recommend this device to assist in difficult lesions, particular in TRI procedures. It was also suitable for a patient with renal insufficiency. We believe this technique is a first-line choice of radial interventionalists treating patients with complex coronary lesions or CKD, reducing contrast use and total fluoroscopy time.
References
Kiemeneij, F., Laarman, G. J., Odekerken, D., Slagboom, T. & van der Wieken, R. A randomized comparison of percutaneous transluminal coronary angioplasty by the radial, brachial and femoral approaches: the access study. J Am Coll Cardiol 29, 1269–1275 (1997).
Mann, T. et al. Stenting in acute coronary syndromes: a comparison of radial versus femoral access sites. J Am Coll Cardiol. 32, 572–576 (1998).
Louvard, Y. et al. Comparison of transradial and transfemoral approaches for coronary angiography and angioplasty in octogenarians (the OCTOPLUS study). Am J Cardiol. 94, 1177–1180 (2004).
Chase, A. J. et al. Association of the arterial access site at angioplasty with transfusion and mortality: the M.O.R.T.A.L study (Mortality benefit Of Reduced Transfusion after percutaneous coronary intervention via the Arm or Leg). Heart. 94,1019–1025(2008).
Jolly, S. S., Amlani, S., Hamon, M., Yusuf, S. & Mehta, S. R. Radial versus femoral access for coronary angiography or intervention and the impact on major bleeding and ischemic events: A systematic review and meta-analysis of randomized trials. Am Heart J. 157, 132–140 (2009).
Abbate, A. et al. Survival and cardiac remodeling benefits in patients undergoing late percutaneous coronary intervention of the infarct-related artery: evidence from a meta-analysis of randomized controlled trials. J Am Coll Cardiol. 51, 956–964 (2008).
Dehghani, P. et al. Mechanism and predictors of failed transradial approach for percutaneous coronary interventions. JACC Cardiovasc Interv. 2, 1057–1064 (2009).
Mamas, M. A., Fath-Ordoubadi, F. & Fraser, D. Successful use of the Heartrail II Catheter as a stent delivery catheter following failure of conventional techniques. Catheter Cardiovasc Interv. 71, 358–363 (2008).
Banerjee, S. & Brilakis, E. S. Use of the Proxis embolic protection device for guide anchoring and stent delivery during complex saphenous vein graft interventions. Cardiovasc Revasc Med. 10, 183–187 (2009).
Brilakis, E. S. & Banerjee, S. Novel uses of the proxis embolic protection catheter. Catheter Cardiovasc Interv. 74, 438–445 (2009).
Mamas, M. A. et al. Use of the Heartrail II catheter as a distal stent delivery device; an extended case series. Eurointervention. 5, 265–271 (2009).
Mamas, M. A., Fath-Ordoubadi, F. & Fraser, D. G. Distal stent delivery with guideliner catheter: First in man experience. Catheter Cardiovasc Interv. 76, 102–111 (2010).
Cola, C., Miranda, F., Vaquerizo, B., Fantuzzi, A. & Bruguera, J. The Guideliner™ Catheter for Stent Delivery in Difficult Cases: Tips and Tricks. J Interv Cardiol. 24, 450–461 (2011).
Farooq, V., Mamas, M. A., Fath-Ordoubadi, F. & Fraser, D. G. The use of a guide catheter extension system as anaid during transradial percutaneous coronary intervention of coronary artery bypass grafts. Catheter Cardiovasc Interv. 78, 847–863 (2011).
Burzotta, F. et al. Use of a second buddy wire during percutaneous coronary interventions: a simple solution for some challenging situations. J Invasive Cardiol. 17, 171–174 (2005).
Hirokami, M., Saito, S. & Muto, H. Anchoring technique to improve guiding catheter support in coronary angioplasty of chronic total occlusions. Catheter Cardiovasc Interv. 67, 366–371 (2006).
Mario, C. D. & Ramasami, N. Techniques to enhance guide catheter support. Catheter Cardiovasc Interv. 72, 505–512 (2008).
Huang, Z. H. et al. Usefulness and safety of a novel modification of the retrograde approach for the long tortuous chronic total occlusion of coronary arteries. Int Heart J. 31, 351–356 (2017).
Ryan, T. J. et al. Guidelines for percutaneous transluminal coronary angioplasty. A report of the American College of Cardiology/American Heart Association Task Force on Assessment of Diagnostic and Therapeutic Cardiovascular Procedures (Subcommittee on Percutaneous Transluminal Coronary Angioplasty). Circulation. 78, 486–502 (1988).
Yew, K. L. & Kang, Z. Guidezilla guide extension catheter facilitated the delivery of long and bulky In. Pact Falcon drug-coated balloon for the treatment of chronic total occlusion lesion. Int J Cardiol. 201, 220–221 (2015).
Waggoner, T., Desai, H. & Sanghvi, K. A unique complication of the GuideZilla guide extension support catheter and the risk of stent stripping in interventional & endovascular interventions. Indian Heart J. 67, 381–384 (2015).
De Man, F. H. et al. Usefulness and safety of the GuideLiner catheter to enhance intubation and support of guide catheters: insights from the Twente GuideLiner registry. Eurointervention. 8, 336–344 (2012).
Chan, P. H. et al. Extended use of the GuideLiner in complex coronary interventions. EuroIntervention. 11, 325–335 (2015).
Takahashi, S. et al. New method to increase a backup Support of a 6 French guiding coronary catheter. Catheter Cardiovasc Interv. 63, 452–456 (2004).
Tunuguntla, A., Daneault, B. & Kirtane, A. J. Novel use of the GuideLiner catheter to minimize contrast use during PCI in a patient with chronic kidney disease. Catheter Cardiovasc Interv. 80, 453–455 (2012).
Papayannis, A. C., Michael, T. T. & Brilakis, E. S. Challenges associated with use of the GuideLiner catheter in percutaneous coronary interventions. J Invasive Cardiol. 24, 370–371 (2012).
Kumar, S., Gorog, D. A., Secco, G. G., Di Mario, C. & Kukreja, N. The GuideLiner “child” catheter for percutaneous coronary intervention - early clinical experience. J Invasive Cardiol. 22, 495–498 (2010).
Chang, Y. C., Fang, H. Y., Chen, T. H. & Wu, C. J. Left main coronary artery bidirectional dissection caused by ejection of guideliner catheter from the guiding catheter. Catheter Cardiovasc Interv. 82, E215–220 (2013).
Tariq, B., Hassan, B. & Frank, T. A unique complication with use of the GuideLiner catheter in percutaneous coronary interventions and its successful management. J Invasive Cardiol. 26, E42–44 (2014).
Murphy, J. C. & Spence, M. S. Guideliner catheter-friend or foe? Catheter Cardiovasc Interv. 80, 447–450 (2012).
Seto, A. & Kern, M. The Guideliner: Keeping your procedure on track or derailing it? Catheter Cardiovasc Interv. 80, 451–452 (2012).
Acknowledgements
We would like to thank the patients who participated in the study and the staff from departments of Cardiology and Radiology. This work was partly supported by the Applied and Fundamental Study of Sichuan Province (No. 2017JY0026).
Author information
Authors and Affiliations
Contributions
M.M. conceived the idea for the study, acquisition of data and drafting the article. D.K.Y. help acquisition and analysis of data. L.X.J. contributed to analysis, critically revised the manuscript, and H.Y. are finally approval of the version to be submitted. All authors read and approved the final version of the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ma, M., Diao, Ky., Liu, Xj. et al. Early clinical experience with Guidezilla for transradial interventions in China. Sci Rep 8, 5444 (2018). https://doi.org/10.1038/s41598-018-23633-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-23633-7