Abstract
This study evaluated artefacts on computed tomography (CT) images using Hounsfield units (HU) in patients with spinal oligometastatic disease who received carbon-fiber (CF; n = 11) or titanium (n = 11) spine implants and underwent stereotactic ablative radiotherapy (SABR). Pre- and postoperative HU were measured at the vertebral body, pedicle, and spinal cord at three different levels: the lower instrumented vertebra, the level of metastatic spinal cord compression, and an uninvolved level. Areas measured at each level were delicately matched pre- and postoperatively. Significant differences in HU were observed at the vertebral body, the pedicle, and the spinal cord at the lowest instrumented vertebra level for both CF and titanium (average increase 1.54-fold and 5.11-fold respectively). At the metastatic spinal cord compression level, a trend towards a higher HU-increase was observed in titanium compared with CF treated patients (average increase 2.51-fold and 1.43-fold respectively). The relatively high postoperative HU-increase after insertion of titanium implants indicated CT artefacts, while the relatively low HU-increase of CF implants was not associated with artefacts. Less CT artefacts could facilitate an easier contouring phase in radiotherapy planning. In addition, we propose a CT artefact grading system based on postoperative HU-increase. This system could serve as a valuable tool in future research to assess if less CT artefacts lead to time savings during radiotherapy treatment planning and, potentially, to better tumoricidal effects and less adverse effects if particle therapy would be administered.
Similar content being viewed by others
Introduction
Spinal metastases are common in oncological care as approximately 70% of all bony metastasized cancers are located in the spine1,2. Spinal oligometastatic disease (OMD) is defined as a subgroup of patients with limited (≤ 5) metastatic lesions in the spine where all metastatic sites are safely treatable3,4. Treatment of spinal OMD is a multidisciplinary team effort, and management must be individualized for each patient. Factors that impact treatment strategy include histology, tumor location, symptoms, radiosensitivity, and prior treatment5. Surgery can be performed in case of mechanical pain, decompression, correction of instability or deformity, and with the purpose of oncological cytoreduction6. In patients with limited spinal OMD, surgery combined with postoperative radiotherapy (RT) to improve local control is an established practice7.
Conventional external beam radiation therapy (EBRT) to the entire spine has been the golden standard for decades due to its excellent palliative effect. However, EBRT doses are too low to ensure long term local control, and raising the dose is not an option because the spinal cord is often at risk8. Fortunately, stereotactic ablative radiotherapy (SABR) is an emerging noninvasive approach for the treatment of spinal OMD9. It has drastically changed the treatment from palliative to curative care for several (early detected) cancers, including lung-, liver-, prostate-, breast-, and spine cancer10,11,12,13. SABR can precisely deliver tumoricidal radiation doses to the tumor(s), while sparing adjacent tissues, thereby achieving durable local tumor control with low complication rates10. This is a delicate procedure that highly depends on accuracy, not only because the dose must be high enough to be toxic to tumor cells, but also because it requires high precision as the spinal cord is often right next to the area being treated14. Therefore, precise SABR planning with computed tomography (CT), or magnetic resonance imaging (MRI), is essential to ensure optimal treatment for spinal OMD15. Several prospective trials have already demonstrated that SABR is an effective tool for treating spinal OMD16,17,18,19,20.
Although very promising, a major challenge in the delivery of SABR to spinal OMD is the proximity of the spinal cord. Despite technical evolutions such as surface-guided monitoring systems, metal artefact reduction, and couch corrections in all six degrees of freedom, SABR treatment can be hampered when spinal tumors are treated with titanium (or other metallic) implants21. Commonly used titanium implant materials produce substantial artefacts on CT images22. Consequently, these implants pose problems with respect to (time-consuming) radiation planning and accurate delivery of the calculated dose22. This could lead to complications such as spinal cord radio necrosis, progressive myelopathy, spinal hemorrhage, and fractures23,24. Tedesco et al. reported that scattering of radiotherapy from titanium spine implants can compromise the therapeutic effect and lead to unwanted radiation to adjacent healthy tissue25. In addition, titanium (or other metallic) artefacts also interfere with postoperative radiologic surveillance used to track bone healing and identify recurrences26.
A possible improvement for spinal OMD treatment with SABR is to change traditional titanium (or other metallic) implants to innovative carbon-fiber (CF) implants. CF materials have good biocompatibility, chemical stability, good mechanical properties, and a modulus of elasticity which is similar to human bone and theoretically leads to better bone quality27. Besides, clinical studies have not shown an increase in complications with implementation of CF implants25,28,29,30. Therefore, CF implants could improve SABR planning and lead to more accurate delivery of the calculated dose compared to traditional implants31. Several CF spine implants have shown promising results with regards to reducing artefacts, better radiation planning, and potentially greater safety and quality of radiotherapy25,32,33. However, it has been difficult to quantify the difference in CT artefacts after implementation of CF and titanium implants. In the current study, we utilized a quantitative technique to perform a pre- and postoperative comparison of CT artifacts produced by CF and titanium implants in patients with spinal OMD undergoing postoperative SABR. In addition, we propose an artefact grading system to classify CT artefacts.
Materials and methods
This retrospective single center study included patients ≥ 18 years with spinal OMD who received CF or titanium spinal implants (including pedicles, screws, and rods) between 2018 and 2020. A closely matched gender and age group of patients that received CF and titanium implants was selected because gender- and age-related osteoporotic changes in bone density could potentially influence CT artefact measurements. Patients with traumatic or inflammatory conditions, or previous fusion surgery were not eligible. Additionally, patients that received bone cement (i.e., polymethyl methacrylate or PMMA) at the level of metastasis or instrumented level were excluded because this could potentially interfere with CT artefact measurements. The study protocol was approved by the local ethics committee (St. George’s Research Ethics Committee, clinical audit registration number AUDI003026), and informed consent was obtained from all subjects and/or their legal guardians. All methods were performed in accordance with relevant guidelines and regulations.
Surgery was recommended as a curative treatment strategy in patients with spinal OMD. Most patients presented with pain, some with spinal cord compression, and response to non-surgical treatment was insufficient. Decompression and fixation surgery was predominantly performed for those cases, and there was no standardized protocol to choose for CF implants instead of titanium implants. Therefore, the choice between CF or titanium was made by shared decision making and the preference of the operating surgeon. During this study, patients were treated with various FDA approved and CE marked CF (CarboFix Orthopedics; Herzliya, Israel) and titanium implants (Stryker Corporation, Michigan, United States of America).
Outcomes assessment
Artefacts on CT images were measured preoperatively and within the first postoperative week using Hounsfield units (HU), which were determined by a picture archiving and communication system (PACS) integrated software (Phillips Medical Systems, Eindhoven, The Netherlands). An unmodified standard care CT spine protocol (median tube potential 140kVp, median tube current 60mA, 2.5mm slice thickness) was used to measure the artefacts; the software measured the density of a region of interest (ROI), which electronically overlaid the image, and calculated the corresponding HU. ROI were selected by a neurosurgeon (K.A.K.) and checked by an orthopedic surgeon specialized in spine surgery (D.L.). Pre- and postoperative HU measurements were performed at the vertebral body, pedicle, and spinal cord at three different levels: the lower instrumented vertebra, the level of metastatic spinal cord compression, and at an uninvolved level. These locations were chosen because the lower instrumented vertebra received the implant. The level of metastatic spinal cord compression was chosen because this is the level where radiotherapy is directed, and usually no screws are inserted at this level which means that HU changes come from scattering from different levels. Additionally, an uninvolved level served as internal control. Areas were measured as a circle in the vertebral body and spinal canal, while an ellipse was used for the pedicle (Fig. 1). CT images were analyzed for the CF as well as the titanium implant group, and HU areas measured at each level were delicately matched between pre- and postoperative images (< 10% difference in the examined area measured in cm2). In addition, a CT artefact grading system was developed based on the postoperative HU change.
Statistical analysis
Statistical analyses were performed using SPSS version 25 (IBM Corp, Somer, NY, USA). For continuous data, the Kolmogorov–Smirnov test was used to assess the normal distribution assumption. Mean HU were calculated for each group pre- and postoperatively. These mean HU values were used because the distribution of our data was symmetric without clear outliers, and a paired t-test was used to compare the mean HU pre- versus postoperatively. Significance was set as a p-value ≤ 0.05.
Results
In total, 22 patients were included. Six males and five females (n = 11) were included in the CF implant group, with a mean age of 54 years (range 20–70 years). The titanium group consisted of five males and six females (n = 11), with a mean age of 56 years (range 36–66 years). Although two patients in the CF group were primary spine tumors, most of the lesions were spinal metastasis, and the involved location was mostly at the thoracic level of the spine (Table 1).
Pre-versus postoperative HU comparison for carbon fiber implants
The average HU of the vertebral body, pedicle, and spinal cord at the level of the lower instrumented vertebra, at the metastatic spinal cord compression level, and at the uninvolved level was compared before and after insertion of the CF implants. CF instrument scatter artefacts were observed with higher postoperative HU. Although HU only increased with a maximum of 1.67-fold compared to its preoperative value, there was a significant increase at the vertebral body-, pedicle-, and spinal cord at the lowest instrumented vertebra level (p = 0.012, 0.015, and 0.014 respectively; Table 2). No statistically significant HU-increase was observed at the metastatic spinal cord compression level (generally no instrumentation at the tumor level) and at the uninvolved level (internal control) (Table 2).
Pre-versus postoperative HU comparison for titanium implants
The average HU of the vertebral body, pedicle, and spinal cord at the level of the lower instrumented vertebra, at the metastatic spinal cord compression level, and at the uninvolved level was compared before and after insertion of the titanium implants. Titanium instrument scatter artefacts were observed with higher postoperative HU. The same trend was observed as with CF implants, with significantly increased HU postoperatively (maximum 5.65-fold increase) at the vertebral body-, pedicle-, and spinal cord at the lowest instrumented vertebra level (p-values of 0.00, 0.00, and 0.24 respectively; Table 3). Although not statistically significant, a trend towards higher HU was observed in titanium implants (average 2.51-fold HU-increase) at the metastatic spinal cord compression level. As expected, no significant HU-increase was observed at the uninvolved level (internal control).
Artefact grading system to classify CT artefacts
Based on our observations the postoperative HU-increase was associated with an increase in CT artefacts. Therefore, we propose a CT artefact grading system where grade 0 = no increase (i.e., bone allografts), grade 1 = 1–1.3-fold (differences in study planes), grade 2 = 1.3–2-fold increase (CF), grade 3 = 2–4-fold increase (i.e., CF with cement), and 5 = > 4-fold increase (titanium) (Table 4). This grading system provides information on the ability to assess anatomically relevant structures and could be used in future studies to assess if less CT artefacts indeed facilitate an easier contouring phase in radiotherapy planning and possibly lead to better tumoricidal effects and decreased adverse outcomes in cases where newer forms of radiotherapy, such as particle therapy, would be considered.
Discussion
In this study, we utilized a quantitative technique to perform a pre- versus postoperative comparison of HU produced by CF and titanium implants in patients with spinal OMD undergoing postoperative SABR. Significant increases in HU were observed at the vertebral body, the pedicle, and the spinal cord at the lowest instrumented vertebra level for both CF and titanium implants (average 1.54-fold and 5.11-fold HU-increase, respectively). At the metastatic spinal cord compression level, a trend towards a higher HU-increase was observed in titanium implants compared with CF implants (average 2.51-fold and 1.43-fold HU-increase respectively). In general, no screws are inserted at this level, which means the HU-increase comes from the scatter of a cage or rods posteriorly. As expected, no postoperative HU-increase was observed at the uninvolved level for both CF and titanium. Based on our observations, the HU-increase indicates an increase in CT artefacts. Therefore, we propose a CT artefact grading system based on postoperative HU-increase, which provides information on the ability to assess anatomically relevant structures and could be used in future long term follow up studies. These studies could assess if less CT artefacts (low grade artefacts) indeed lead to time savings during radiotherapy planning and, potentially, to enhanced tumoricidal effects with less adverse outcomes in cases where particle therapy would be administered.
A recent in vitro study of Krätzig et al. evaluated the susceptibility of artefacts in CT and MRI of titanium and CF screw-rod constructs for posterior spinal stabilization using a standardized in vitro model34. Here, similar manually placed 2D ROI were defined for each image, and CT imaging with typical implant configuration for thoracic stabilization demonstrated a significant artifact reduction in CF compared with titanium implants for the evaluation of index structures, such as the spinal cord and the vertebra. Coherently, Depauw et al. used a water phantom as a human tissue equivalent and reported no imaging artefacts and minimal dose perturbation of CF compared with titanium35. In addition, Fleege et al. reported reduced artifacts of CF pedicle screws in MRI scans of patients with lumbar spondylodesis36. The authors calculated the surface of the artifact free vertebral body area as percentage of the total vertebral body, and CF displayed significantly less artefacts than titanium (67.1 ± 5.6% vs. 48.3 ± 5.0%; p ≤ 0.01, respectively). Furthermore, Ringel et al. reported reduced artefacts of CF spine implants compared with titanium implants and conclude that CF spine implants are a valuable and feasible option in spine tumors where postoperative imaging and radiation planning are necessary26. Our findings, together with the previously mentioned studies, highlight that CF spine implants show reduced artefacts compared to titanium.
The clinical relevance of the reduced artefacts after CF implementation instead of titanium remains to be further elucidated. New treatment planning systems, which outline metallic materials and associate an atomic number which is used for dose calculation purposes (the density override method), correct for metallic artefacts. Therefore, metallic implants are unlikely to impact the tumoricidal effects of SABR. However, the density override method no longer needs to be applied with CF implants, which results in a simpler method and therefore time savings, as well as an accurate dose distribution37. Besides, radiation oncologists are increasingly interested in CF spinal instrumentation because it enables the use of particle therapy, such as proton beam therapy, in a group of patients where it was previously impossible due to the imaging artefacts and perturbation effect of metallic instrumentation38. Several studies have shown that the use of CF is favorable to titanium instrumentation for the use in particle therapy. Nevelsky et al. investigated the perturbation effect of CF screws compared to titanium screws and found a perturbation effect of less than 5% for CF screws, compared to greater than 30% for titanium screws39. Mastella et al. evaluated the dosimeric perturbation caused by CF screws compared to titanium screws and found less dose degradation caused by CF screws, making CF more suitable for particle therapy40. Ultimately, this might help achieving the goal of durable tumor control with low complication rates.
Drawbacks of CF implants include its potential challenging surgery due to its radiolucency. However, recent research has shown CT-guided navigation of pedicle screws is possible for instrumentation and precision assessment across the thoraco-lumbar spine41. Besides, long-term postoperative results of CF implants, including the effect of radiation on the properties of CF implants, remain to be investigated. Some also question whether availability and costs of CF implants could be a disadvantage. Although we are not aware of its availability, production costs have decreased as CF composites are widely used across other industries, and current costs of CF nails are competitive with conventional metal nails42.
This study has several limitations. First, our objective quantitative assessment using HU is not a perfect measurement of artefacts. Although several studies have used HU to predict osteoporosis and artefacts, we acknowledge that CT scanner configurations, depth of the measurement, location of the measurement, tissue type, implant material and artefacts due to motion during the scan can all influence HU34,43. Artefacts can be bright (high HU) or dark (low HU). Therefore, postoperative HU differences, such as our CT artefact grading system based on postoperative HU-increase, might only be a surrogate marker for CT artefacts. Nevertheless, this real life setting with standard clinical protocols provides relevant clinical insights; the relatively high HU-increase after insertion of titanium implants indicated CT artefacts, while the relatively low HU-increase after insertion of CF implants was not associated with artefacts. A qualitative assessment by a musculoskeletal radiologist could contribute to the validity and reliability of our study because this is generally seen as the ground-truth. Although not performed in this study, a qualitative analysis of post-operative artifact-free vertebrae surface area and its ratio to pre-operative vertebrae surface area, as done by Fleege et al., would presumably show better results in patients treated with CF implants36. Another limitation is that our proposed CT artefacts grading system is solely based on 11 patients receiving CF implants and 11 patients receiving titanium implants. Validation is needed to assess if this HU based classification system holds promise for assessing CT artefacts in future studies. Furthermore, our retrospective study design is inherently susceptible to several forms of bias, including selection- and assessor bias, and causal differences should be interpreted with caution. However, we objectively measured carefully matched ROI and compared the hardware against itself (in different levels) and against its own control (pre- and postoperatively). Postoperative HU differences which are clinically relevant for the ability to assess anatomically relevant structures or metastasis/residual tumor, improve planning (time savings) and accurate administration of newer forms of radiotherapy, such as particle therapy, should be further examined.
Conclusion
It has been difficult to quantify the difference in computed tomography (CT) artefacts after implementation of carbon-fiber (CF) and titanium implants. This study utilized a quantitative technique to compare pre- and postoperative CT artifacts produced by CF and titanium implants in patients with spinal oligometastatic disease (OMD) undergoing stereotactic ablative radiotherapy (SABR). A greater increase in Hounsfield units (HU) was observed in the group treated with titanium spine implants than in the group treated with CF spine implants. This relatively high postoperative HU-increase after insertion of titanium implants indicated CT artefacts, while the relatively low HU-increase of CF implants was not associated with artefacts. Therefore, we propose a CT artefact grading system based on postoperative HU-increase. This could be used in future studies to assess if less CT artefacts due to treatment with CF implants lead to time savings during radiotherapy treatment planning and, potentially, better tumoricidal effects and decreased adverse effects if particle therapy would be administered.
Data availability
Raw data are available with the corresponding author and will be provided upon a written request.
References
Ecker, R. D., Endo, T., Wetjen, N. M. & Krauss, W. E. Diagnosis and treatment of vertebral column metastases. Mayo Clin. Proc. 80, 1177–1186. https://doi.org/10.4065/80.9.1177 (2005).
Wong, D. A., Fornasier, V. L. & MacNab, I. Spinal metastases: The obvious, the occult, and the impostors. Spine (Phila Pa 1976) 15, 1–4 (1990).
Hellman, S. & Weichselbaum, R. R. Oligometastases. J. Clin. Oncol. 13, 8–10. https://doi.org/10.1200/jco.1995.13.1.8 (1995).
Lievens, Y. et al. Defining oligometastatic disease from a radiation oncology perspective: An ESTRO-ASTRO consensus document. Radiother. Oncol. 148, 157–166. https://doi.org/10.1016/j.radonc.2020.04.003 (2020).
Chang, S. Y., Mok, S., Park, S. C., Kim, H. & Chang, B. S. Treatment strategy for metastatic spinal tumors: A narrative review. Asian Spine J. 14, 513–525. https://doi.org/10.31616/asj.2020.0379 (2020).
Barzilai, O. et al. Essential concepts for the management of metastatic spine disease: What the surgeon should know and practice. Global Spine J. 9, 98s–107s. https://doi.org/10.1177/2192568219830323 (2019).
Gerszten, P. C., Mendel, E. & Yamada, Y. Radiotherapy and radiosurgery for metastatic spine disease: What are the options, indications, and outcomes?. Spine (Phila Pa 1976) 34, S78-92. https://doi.org/10.1097/BRS.0b013e3181b8b6f5 (2009).
Yazici, G. et al. The dosimetric impact of implants on the spinal cord dose during stereotactic body radiotherapy. Radiat. Oncol. 11, 71. https://doi.org/10.1186/s13014-016-0649-z (2016).
Glicksman, R. M. et al. Stereotactic ablative radiotherapy for the management of spinal metastases: A review. JAMA Oncol. 6, 567–577. https://doi.org/10.1001/jamaoncol.2019.5351 (2020).
Barzilai, O., Fisher, C. G. & Bilsky, M. H. State of the art treatment of spinal metastatic disease. Neurosurgery 82, 757–769. https://doi.org/10.1093/neuros/nyx567 (2018).
Barzilai, O. et al. Survival, local control, and health-related quality of life in patients with oligometastatic and polymetastatic spinal tumors: A multicenter, international study. Cancer 125, 770–778. https://doi.org/10.1002/cncr.31870 (2019).
David, S. et al. Stereotactic ablative body radiotherapy (SABR) for bone only oligometastatic breast cancer: A prospective clinical trial. Breast 49, 55–62. https://doi.org/10.1016/j.breast.2019.10.016 (2020).
Zeng, K. L. et al. Stereotactic body radiotherapy (SBRT) for oligometastatic spine metastases: An overview. Front. Oncol. 9, 337. https://doi.org/10.3389/fonc.2019.00337 (2019).
Cox, B. W. et al. International Spine Radiosurgery Consortium consensus guidelines for target volume definition in spinal stereotactic radiosurgery. Int. J. Radiat. Oncol. Biol. Phys. 83, e597-605. https://doi.org/10.1016/j.ijrobp.2012.03.009 (2012).
Billiet, C. et al. Precision of image-guided spinal stereotactic ablative radiotherapy and impact of positioning variables. Phys. Imaging Radiat. Oncol. 22, 73–76. https://doi.org/10.1016/j.phro.2022.04.006 (2022).
Billiet, C. et al. Outcome and toxicity of hypofractionated image-guided SABR for spinal oligometastases. Clin. Transl. Radiat. Oncol. 24, 65–70. https://doi.org/10.1016/j.ctro.2020.06.011 (2020).
Ryu, S. et al. Partial volume tolerance of the spinal cord and complications of single-dose radiosurgery. Cancer 109, 628–636. https://doi.org/10.1002/cncr.22442 (2007).
Sahgal, A. et al. Stereotactic body radiotherapy versus conventional external beam radiotherapy in patients with painful spinal metastases: An open-label, multicentre, randomised, controlled, phase 2/3 trial. Lancet Oncol. 22, 1023–1033. https://doi.org/10.1016/s1470-2045(21)00196-0 (2021).
Sahgal, A. et al. Probabilities of radiation myelopathy specific to stereotactic body radiation therapy to guide safe practice. Int. J. Radiat. Oncol. Biol. Phys. 85, 341–347. https://doi.org/10.1016/j.ijrobp.2012.05.007 (2013).
Yamada, Y. et al. High-dose, single-fraction image-guided intensity-modulated radiotherapy for metastatic spinal lesions. Int. J. Radiat. Oncol. Biol. Phys. 71, 484–490. https://doi.org/10.1016/j.ijrobp.2007.11.046 (2008).
Boriani, S. et al. Carbon-fiber-reinforced PEEK fixation system in the treatment of spine tumors: A preliminary report. Eur. Spine J. 27, 874–881. https://doi.org/10.1007/s00586-017-5258-5 (2018).
Jia, Y., Zhao, L., Cheng, C. W., McDonald, M. W. & Das, I. J. Dose perturbation effect of metallic spinal implants in proton beam therapy. J. Appl. Clin. Med. Phys. 16, 333–343. https://doi.org/10.1120/jacmp.v16i5.5566 (2015).
Béhin, A. & Delattre, J. Y. Complications of radiation therapy on the brain and spinal cord. Semin. Neurol. 24, 405–417. https://doi.org/10.1055/s-2004-861535 (2004).
Furuya, T. et al. Evaluation of multi-institutional end-to-end testing for post-operative spine stereotactic body radiation therapy. Phys. Imaging Radiat. Oncol. 16, 61–68. https://doi.org/10.1016/j.phro.2020.09.005 (2020).
Tedesco, G., Gasbarrini, A., Bandiera, S., Ghermandi, R. & Boriani, S. Composite PEEK/Carbon fiber implants can increase the effectiveness of radiotherapy in the management of spine tumors. J. Spine Surg. 3, 323–329. https://doi.org/10.21037/jss.2017.06.20 (2017).
Ringel, F. et al. Radiolucent carbon fiber-reinforced pedicle screws for treatment of spinal tumors: Advantages for radiation planning and follow-up imaging. World Neurosurg. 105, 294–301. https://doi.org/10.1016/j.wneu.2017.04.091 (2017).
Xin-ye, N., Xiao-bin, T., Chang-ran, G. & Da, C. The prospect of carbon fiber implants in radiotherapy. J. Appl. Clin. Med. Phys. 13, 3821. https://doi.org/10.1120/jacmp.v13i4.3821 (2012).
Complications of patients with bone tumors treated with carbon-fiber plates: An international multicenter study. Sci. Rep. 12, 18969. https://doi.org/10.1038/s41598-022-23519-9 (2022).
Joerger, A. K. et al. CFR-PEEK pedicle screw instrumentation for spinal neoplasms: A single center experience on safety and efficacy. Cancers (Basel) https://doi.org/10.3390/cancers14215275 (2022).
Rijs, Z. et al. Complications of patients with bone tumors treated with carbon-fiber plates: An international multicenter study. Sci. Rep. 12, 18969. https://doi.org/10.1038/s41598-022-23519-9 (2022).
Sakaura, H. et al. Outcome of total en bloc spondylectomy for solitary metastasis of the thoracolumbar spine. J. Spinal Disord. Tech. 17, 297–300. https://doi.org/10.1097/01.bsd.0000096269.75373.9b (2004).
Takayanagi, A. et al. Radiolucent carbon fiber-reinforced implants for treatment of spinal tumors-clinical, radiographic, and dosimetric considerations. World Neurosurg. 152, 61–70. https://doi.org/10.1016/j.wneu.2021.05.100 (2021).
Neal, M. T. et al. Carbon fiber-reinforced PEEK instrumentation in the spinal oncology population: A retrospective series demonstrating technique, feasibility, and clinical outcomes. Neurosurg. Focus. 50, E13. https://doi.org/10.3171/2021.2.Focus20995 (2021).
Krätzig, T. et al. Carbon fiber–reinforced PEEK versus titanium implants: An in vitro comparison of susceptibility artifacts in CT and MR imaging. Neurosurg. Rev. 44, 2163–2170. https://doi.org/10.1007/s10143-020-01384-2 (2021).
Depauw, N., Pursley, J., Lozano-Calderon, S. A. & Patel, C. G. Evaluation of carbon fiber and titanium surgical implants for proton and photon therapy. Pract. Radiat. Oncol. https://doi.org/10.1016/j.prro.2023.01.009 (2023).
Fleege, C. et al. Carbon fiber-reinforced pedicle screws reduce artifacts in magnetic resonance imaging of patients with lumbar spondylodesis. Sci. Rep. 10, 16094. https://doi.org/10.1038/s41598-020-73386-5 (2020).
Akyol, O., Dirican, B., Toklu, T., Eren, H. & Olgar, T. Investigating the effect of dental implant materials with different densities on radiotherapy dose distribution using Monte-Carlo simulation and pencil beam convolution algorithm. Dentomaxillofac Radiol. 48, 20180267. https://doi.org/10.1259/dmfr.20180267 (2019).
Gaito, S. et al. Proton beam therapy in the Oligometastatic/Oligorecurrent setting: Is there a role? A literature review. Cancers (Basel) 15, 2489. https://doi.org/10.3390/cancers15092489 (2023).
Nevelsky, A., Borzov, E., Daniel, S. & Bar-Deroma, R. Perturbation effects of the carbon fiber-PEEK screws on radiotherapy dose distribution. J. Appl. Clin. Med. Phys. 18, 62–68. https://doi.org/10.1002/acm2.12046 (2017).
Mastella, E. et al. Dosimetric characterization of carbon fiber stabilization devices for post-operative particle therapy. Phys. Med. 44, 18–25. https://doi.org/10.1016/j.ejmp.2017.11.008 (2017).
Hubertus, V. et al. Navigation accuracy and assessability of carbon fiber-reinforced PEEK instrumentation with multimodal intraoperative imaging in spinal oncology. Sci. Rep. 12, 15816. https://doi.org/10.1038/s41598-022-20222-7 (2022).
Herzog, L. N., Traven, S. A., Walton, Z. J. & Leddy, L. R. The use of carbon fiber implants for impending or existing pathologic fractures. J. Orthop. Trauma. 36, e260–e264. https://doi.org/10.1097/bot.0000000000002320 (2022).
Zaidi, Q., Danisa, O. A. & Cheng, W. Measurement techniques and utility of Hounsfield unit values for assessment of bone quality prior to spinal instrumentation: A review of current literature. Spine (Phila Pa 1976) 44, E239-e244. https://doi.org/10.1097/brs.0000000000002813 (2019).
Author information
Authors and Affiliations
Contributions
Z.R., K.A.K., M.v.d.S., and D.L.: Conceptualization and Methodology. Z.R., P.S.: Data curation, Writing- Original draft preparation (including figures and tables). Z.R., K.A.K., P.S. and D.L.: Investigation. K.A.K., M.v.d.S., and D.L.: Supervision. Z.R., K.A.K., M.v.d.S., and D.L.: Writing- Reviewing and Editing.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Rijs, Z., Kawsar, K.A., Saha, P. et al. Evaluation of computed tomography artefacts of carbon-fiber and titanium implants in patients with spinal oligometastatic disease undergoing stereotactic ablative radiotherapy. Sci Rep 14, 6700 (2024). https://doi.org/10.1038/s41598-024-52498-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-024-52498-2
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.