Introduction

Intramedullary nailing is one of the most common surgical operations for fixation of long bone fractures. Nail deformation affects the accuracy of distal interlocking of the intramedullary nails (IMN) which is a common procedure in tibia and femoral fractures1,2,3,4. Moreover, holes locations of the intramedullary nail are unknown which adds to the deformation problem5,6,7,8,9. Numerous techniques for interlocking screw insertion have been proposed to optimize number of drilling and operation time and avoid ionizing radiation exposure10,11. However, these techniques have not gained widespread clinical application. Other methods are based on fluoroscopy that called free-hand technique (FH) or based on a special device designed for this purpose, knows as targeting-arm devices (TAD)12,13,14,15. Fluoroscopy is a high-cost operation in which patients and the surgical team are exposed to X-ray. Targeting device is a time-consuming process that involves trial and error16,17.

To reduce the high risk of X‐ray exposure, increases surgical accuracy, and accelerate the interlocking procedure, significant progress has been developed. The flag and grid technique increases the accuracy of intramedullary nailing by displaying the position of nail holes relative to the flag and grid in fluoroscopic images18. In addition to the risk of X-ray exposure, there is still the possibility of error due to the slipping and flag and grid displacement. Computer‐aided IMN localization techniques were also introduced which representing a significant reduction in fluoroscopic process time for interlocking screw insertion in a high-cost procedure19,20,21,22. Robotic techniques were also proposed in which a robotic system is utilized to drilling and installing fixation screws, requiring a large robot in the operating room representing a high cost, cumbersome, and X-ray-based method23,24,25. Naked‐eye 3D AR enables observation of the surgical path by costly and cumbersome equipment26,27,28,29. Electro-Magnetic (EM) and laser navigation systems can increase surgical accuracy without any X-ray exposure, but with excessive electrical noise causing large magnetic‐field distortion resulting in unstable tracking paths30,31,32,33,34,35,36,37,38. Moreover, the tracking accuracy is degraded with increasing distance of sensor and EM transmitter. This method is time-consuming and requires installation of various equipment39. Augmented fluoroscopy can decrease radiation exposure, surgical time, and simplify the distal interlocking process. But the patient and the surgical team are exposed to X-ray and it is a high-cost procedure40,41,42.

In this work, have been developed a non-invasive optical method for easy determining IMN holes position inside the bone canal utilizing a safe laser system with a safe visible wavelength that possesses proper transmittance, reflectance, and absorbance properties in human tissues. In order to find the optimized wavelength, Ex-Vivo spectroscopy experiments were performed on animal tissues i.e. on two groups of cow and sheep in the visible range of electromagnetic spectrum. Two parameters which are procedure time and drilling quality, are selected as the merit criteria which proves the acceptable performance of our proposed method. Then an In-Vivo performance experiment performed on a middle-aged man with a desired outcomes. To compare the provided method, targeting-arm, and free-hand techniques, two human experiments were performed on a middle-aged and a young man. The results indicate the advantage of the proposed technique. Intramedullary nail holes laser indicator is introduced here as a simple, non-invasive, and safe tool that reduces surgical time and simplifies the process. The presented Ex-Vivo and In-Vivo results pave the way for development of an effective and safe intramedullary nailing process in clinical applications.

Materials and methods

In this study have been proposed a novel radiation-free technique, using laser light to target the cross lock of the IMN. To achieve optimal wavelength, transmittance and reflectance spectroscopy of biological tissues were performed. Then, IMN holes laser indicator was made based on spectroscopy results. Animal and human tests were performed to evaluate the performance of the proposed technique. Figure 1 shows the study flow diagram.

Figure 1
figure 1

Flow diagram depicting the study selection procedure.

Spectroscopy of biological tissues

We performed Ex-Vivo spectroscopy of biological tissues including bone, fat, tendon, muscle, and skin. Samples were prepared in dimensions of 30 × 30 ± 2 mm and 2–4 mm thickness from cow and sheep legs which were supplied from the Zarrinshahr slaughterhouse, Iran. Two homemade spectroscopy setups were utilized: one for obtaining the reflectance spectrum (Fig. 2A) and another was designed for transmittance measurements (Fig. 2B). The reflectance measurements performed by a reflective probe mounted on a holder base which is placed at 5 mm hight from the sample. This reflective probe consists of seven optical fibres (750 µm, NA = 0.47 PMMA Optical Fibre-Mitsubishi Electric Co-Tokyo-Japan) which the middle one is coupled to the xenon source (ASB-XE-175-Spectral Products Co- Putnam-US) by a lens array. The induced light to the sample surface is collected by six side fibres and then the reflected spectrum is monitored by an optical spectrometer (Spectronix Ar 2015v- Teifsanje Co-Tehran-Iran). To eliminate the source spectral characteristics, a BaSO4 pill was used as a reference. We molded 4.49gr of BaSO4 powder (CAS Number: 7727-43-7-Titrachem Co-Iran) in 250mpa pressure. The reflectance spectrum of the homemade BaSO4 pill was collected and divided by the source spectrum yielding a flat spectral response through the visible spectrum.

Figure 2
figure 2

(A) Reflectance and (B) transmittance spectroscopy setups.

Two lens arrays are utilized to obtain the transmittance spectrum, one mounted on the holder base for focusing the xenon light on the sample, and the second mounted in front of the first array to collect transmitted rays. The sample mounted on a sample holder in the proper distance from the lens arrays.

IMN holes laser indicator

As shown in Fig. 3A, the IMN holes laser indicator consists of a 680 nm–350 mW solid-state laser and a flexible biocompatible probe. The laser light is guided through the designed probe and then passes through the IMN hole, making the position of the IMN hole appears on the skin for the naked eyes. This portable device capable of washing, disinfecting, and autoclaving. Moreover, it uses a rechargeable lithium battery and can provide constant optical power during operation. Figure 3B shows a schematic image of positioning IMN hole using the IMN holes laser indicator.

Figure 3
figure 3

(A) IMN holes laser indicator and (B) a schematic image of positioning IMN hole using the IMN holes laser indicator.

IMN holes laser indicator in animal tests

Measurements were performed on 10 animal samples, 5 sheep's legs, and 5 cows' legs which were supplied from the Zarrinshahr slaughterhouse, Iran. Sheep's samples had a small diameter between 16 and 21 mm and a large diameter between 29 and 35 mm. Also, the cows' samples had a small diameter between 40 and 48 mm and a large diameter between 65 and 73 mm. For each sample, intramedullary nailing (Pooyandegan Pezeshki Pardis-Golestan-Iran) was performed inside the tibia bone, and distal locking was done three times using the laser intramedullary nail holes indicator. Figure 4, shows intramedullary nailing procedures for one of the samples.

Figure 4 
figure 4

Intramedullary nailing for one of the samples, (A) determining the position of the IMN hole and skin incision, (B) bone drilling, (C) fastening the interlocking screw, (D) checking the screw insertion inside the IMN hole.

Output parameters were procedure time and drilling quality. The procedure time as one of the performance parameters is defined as the total time needed between determining the position of the IMN hole and checking the screw insertion. Another performance parameter is the drilling quality which is evaluated as follows: 3 points for successful operation and if the drill does not hit the nail, 2 points for successful operation but with a slight collision with the nail,1 point for severe interference of the drill with nail (in this case, the drilling site must be corrected), 0 point in case of failure. Mean value, standard deviation, and p-value were calculated for both the procedure time, and drilling quality. Paired t-tests were used to compare these performance parameters for cows' and sheep`s samples considering α = 0.05.

IMN holes laser indicator in a human test

The evaluation was performed on a middle-aged man with an acute tibia and fibula fracture of the distal area of the right leg. The study was approved by the ethics committee of the Shohada Lenjan Hospital, Iran, in January 2020. Also, informed consent was obtained from participants and the study was performed under relevant guidelines and regulations. Figure 5A illustrates an X-ray image of the patient's right leg before the surgery.

Figure 5 
figure 5

Interlocking of the intramedullary nail by the laser system, (A) X-ray image of the patient's right foot before the surgery, (B) Finding drilling position by the IMN holes laser indicator in the dark surgery room, (C) Interlocking the IMN, (D) The X-ray image of the patient's right foot one day after intramedullary nailing.

Intramedullary nailing was performed using a tibia nail and three locking screws (Osveh Asia Medical Instrument Co—Iran). By utilizing the IMN holes laser indicator, finding drilling position, incisions skin, and drilling toward the IMN hole performed in a dark surgery room. Then, fastening the locking screw was performed in the natural light of the surgery room. Distal and proximal interlocking performed successfully for one hole in the distal area and two holes in the proximal area with no errors. Figure 5B finding drilling position by the IMN holes laser indicator in the dark surgery room, and Fig. 5C shows interlocking the IMN. Figure 5D shows the X-ray image of the right patient's foot one day after surgery.

TAD in a human test

Interlocking was performed for the tibia bone of a middle-aged man with the left leg fracture using a TAD (Osveh Asia Medical Instrument Co—Iran). The study was approved by the ethics committee of the Shohada Lenjan Hospital, Iran, in September 2021. Also, informed consent was obtained from participants and the study was performed under relevant guidelines and regulations.

The IMN was fixed by three fixation screws. The distal interlocking process was performed by two locking screws which encountered errors while the proximal interlocking performed successfully. Figure 6A, illustrates an X-ray image of the patient's left leg before the surgery, Fig. 6B, inserting the IMN inside the tibia bone using a hammer, Fig. 6C, distal interlocking of the IMN by the TAD, and Fig. 6D, shows the X-ray image of the patient's foot one day after the surgery.

Figure 6 
figure 6

Interlocking of the intramedullary nail by the TAD, (A) an X-ray image of the patient's left foot before the surgery, (B) Inserting the IMN inside the tibia bone using the hammer, (C) Distal interlocking of the IMN by the TAD, (D) The X-ray image of the patient's left foot one day after surgery.

FH technique in a human test

Interlocking of the IMN was done by a c-arm device (siemens healthineer-USA). The patient was a young man with an acute tibia and fibula fracture of the distal area of the left leg. The surgery team was used X-ray protective lead clothes with a thyroid shield collar. The study was approved by the ethics committee of the Shohada Lenjan Hospital, Iran, in October 2021. Also, informed consent was obtained from participants and the study was performed under relevant guidelines and regulations.

Three fixation screws were used for the interlocking of the IMN. By controlling the movement direction of the drill under X-ray emission, no severe interference occurred between the drill and IMN. Figure 7A, illustrates an X-ray image of the patient's left foot before the surgery, Fig. 7B, inserted guide-wire inside the tibia bone under fluoroscopy control, and Fig. 7C, shows the X-ray image of the patient's left foot one day after the surgery.

Figure 7 
figure 7

Interlocking of the intramedullary nail using FH technique, (A) The X-ray image of the patient's left foot before the surgery, (B) Inserted guide-wire inside the tibia bone under fluoroscopy control, (C) The X-ray image of the patient's left foot one day after the surgery.

Results

Spectroscopy of biological tissues

Ex-Vivo spectroscopy was performed on biological tissues in the visible range of electromagnetic spectrum. The optical properties of animal tissues of bone, tendon, muscle, skin, and fat were investigated. The transmitted and reflected spectra are shown in Fig. 8 for sheep and cow samples.

Figure 8 
figure 8

Spectral behavior of biological tissue in the range of 400 to 700 nm, transmittance spectrum of (A) sheep and (B) cow samples, reflectance spectrum of (C) sheep and (D) cow samples.

IMN holes laser indicator in animal tests

Animal experiments were performed on 10 samples in two group, including 5 sheep legs and 5 cow legs. For each sample, intramedullary nailing performed inside the tibia bone, and distal locking was done three times using laser intramedullary nail holes detector. The output parameters were procedure time and drilling quality shows in Tables 1 and 2.

Table 1 The procedure times for each interlocking of IMN hole in sheep and cow samples.
Table 2 The drilling quality points for each interlocking of IMN hole in sheep and cow samples.

IMN holes laser indicator in a human test

Intramedullary nailing was performed for the tibia bone on a middle-aged man. The IMN holes laser indicator used in the dark conditions of the surgery room to interlocking IMN. First, one hole was interlocked in the distal area, and then two holes were interlocked in the proximal area. The interlocking process was successful and there were no errors. The output parameters, the procedure time and the drilling quality are shown in Tables 3 and 4.

Table 3 The procedure time for each interlocking of IMN hole in the human test using the IMN holes laser indicator, Targeting-Arm Device (TAD), and Free-Hand (FH) techniques.
Table 4 The drilling quality points for each interlocking of IMN hole in the human test using the IMN holes laser indicator, Targeting-Arm Device (TAD), and Free-Hand (FH) techniques.

TAD in a human test

Four fixation screws were inserted in the IMN using the TAD. The distal interlocking process encountered errors while the proximal interlocking performed successfully. For the first and second holes at the distal area of the nail, the drilling position was changed 4 and 3 times, respectively. While for two holes in the proximal area of the nail, the drilling process was performed while the drill did not hit the nail. The output parameters are shown in Tables 3 and 4.

FH technique in a human test

Interlocking of the IMN was done by four fixation screws using a c-arm device. The movement direction of the drill was controlled under X-ray emission. So, no severe interference occurred between the drill and IMN. But the procedure time was increased comparing using the IMN holes laser indicator device and TAD. Although the surgery team was used X-ray protective lead clothes with a thyroid shield collar, the X-ray emission isn`t ineffective on their health especially the patient. The output parameters, the procedure time and the drilling quality are shown in Tables 3 and 4, respectively.

Discussion

According to the results of spectroscopy of biological tissues, for higher wavelength, higher transmittance is obtained. Due to the white colour of bone, a uniform reflectance is observed on the visible spectral range. Regards to the denser bone of cow, it illustrates lower transmittance. Two dips in wavelengths of 540 nm and 576 nm with a distance of 36 nm in the reflectance spectrum of sheep tendon and two dips in wavelengths of 490 nm and 536 nm, with a distance of 46 nm in the reflectance spectrum of cow samples are observed. These dips can be related to the effect of blood absorption on the endogenous fluorescence signal intensity of biological tissues41,43. Changes amount of blood in the tendon vessels can affect the fluorescence spectra44,45. It may result from α and β attenuation signals due to the absorption capacity of the hemoglobin concentration46,47. Moreover, different amounts of fluorescent coenzyme concentration in cow and sheep samples have a great effect on the transmitted intensity. The muscle transmitted and reflected spectra are clearly related to its dark red color. The sample of sheep`s skin has a lighter brown color than cows according to its reflectance diagrams. Fat also illustrates an almost flat reflectance, like the bone spectrum regards to its white color.

Therefore, the optimized wavelength of 680 nm is selected for the IMN holes laser indicator. In addition to low transmittance in lower wavelengths and thus a mitigated performance of the device, the highest absorbance makes these wavelengths unsafe for tissues and organs.

There is a significant difference in the overall procedure time (sheep legs = 99.866 ± 14.532 s vs. cow legs = 120.333 ± 33.047 s, P < 0.05) between the two groups. But there isn`t a significant difference in drilling quality (sheep legs = 2.733 ± 0.457 points vs. cow legs = 2.666 ± 0.487 points, P > 0.05) between the two groups. The overall procedure time obtained for sheep samples compared to the cow samples is less. It`s due to the larger dimensions of the cow's leg than the sheep's. Cow's leg has larger dimensions and thicker, denser tissues than sheep's leg. So cutting tissues and drilling the bone procedures takes more time. If we compare the sheep's leg to an immature human's leg being, and the cow's leg to an adult human's leg being. It can be said that the nailing procedure for an adult compared to an immature human takes more time.

Weidert et al., during an intramedullary nailing study performed on cow legs, reported 378.76 ± 101.10 s for conventional fluoroscopy (free hand technique), and 380.38 ± 165.11 s for video-augmented fluoroscopy41. The overall procedure time on cow samples obtained 120.333 ± 33.047 s using the IMN holes laser indicator. The time obtained is about 260.047 s less than the video-augmented fluoroscopy, and about 258.427 s less than the conventional fluoroscopy. Using the IMN holes laser indicator, the procedure time can be reduced compared to fluoroscopy techniques. It`s an X-ray-free technique. While in free hand technique, the patient and the surgical team are exposed to X-ray radiation. However, the fluoroscopy technique can evaluate the reliability and validation of the fracture healing methods48,49,50.

There aren`t any significant differences in drilling quality between sheep and cow samples (p > 0.05). For sheep samples, 11 holes interlocked successfully without drill impact with the nail, and 4 holes interlocked successfully with a slight collision with the nail. Moreover, in cow samples, 10 holes interlocked successfully without drill impact with the nail, and 5 holes interlocked successfully with a slight collision with the nail. This small difference may be due to human error due to experimenter fatigue of multiple tests. Slight collision with the nail is normal and does not interfere with the process.

By comparing the overall procedure times of the IMN holes laser indicator, and the TAD techniques in the human tests, it is clear the IMN holes laser indicator significantly takes less time to interlock the IMN (IMN holes laser indicator technique = 115.333 ± 8.993 s vs. TAD technique = 326.666 ± 70.343 s, P < 0.05). Also, there is a significant difference between the IMN holes laser indicator, and the FH techniques (IMN holes laser indicator technique vs. FH technique = 916.666 ± 14.613 s, P < 0.05). There is significant difference in drilling quality between the IMN holes laser indicator, and the TAD techniques (the IMN holes laser indicator technique = 2.666 ± 0.471 points vs. the TAD technique = 1.666 ± 0.942 points, P < 0.05). Moreover, the drilling quality between the IMN holes laser indicator, and FH techniques is equal (the IMN holes laser indicator technique vs. the FH technique = 2.666 ± 0.471 points, P = 0.05).

As shown in Table 5, there are different reports of the IMN interlocking in human tests which use the TAD, and the FH techniques. Among these, Ramireddy et al. have reported less procedure time for the IMN interlocking by the TAD technique (180–300 s)62. Also, Windolf et al., have reported less procedure time for the FH technique (246 ± 126)39. Comparing the reported times with the obtained time for the IMN holes laser indicator, it is obvious that the IMN holes laser indicator technique takes less procedure time (115.333 ± 8.993).

Table 5 Different reported procedure times for the free hand technique (FH) and the targeting-arm devices (TAD).

Transmitted light intensity through soft tissues such as fat is more than hard tissues like bone (as shown in Fig. 8). This reduces the intensity of the transmitted light through the patient's foot in areas with thicker soft tissue. Regarding the Gaussian intensity profile of the laser beam (Fig. 9A), in a plane perpendicular to the beam axis, the emitted beam from the laser is more intense in the center and decreases in a Gaussian shape63. Therefore, the transmitted beam through the patient's foot in the center is more intense, which makes it easier to detect the center of the circle created by the laser beam on the patient's foot. This prevents errors in determining the center position of the IMN hole placed in the patient's foot. Also, the shape of the soft tissue layers on the bone is almost identical, which has no effect on the accuracy of the system.

Figure 9 
figure 9

(A) The Gaussian intensity profile of the laser beam, (B) the schematic image of the light passing through the nail holes in proximal and distal regions and (C) the cortex of the patient tibia bone in three regions before the surgery.

During evaluating the IMN holes laser indicator in a human test, by inserting the probe inside the IMN, the light passing through the nail holes in the proximal region was more intense than in the distal region as shown in schematic image Fig. 9B. Moreover, the diameter of the circle created on the patient's foot due to the passing of light in the distal was smaller than the proximal. However, the diameter of the foot is larger in the proximal, and we should have seen less light passing through this region. As shown in Fig. 9C, the cortex in the proximal of the tibia bone has a small thickness and a large diameter. Approaching the distal, the thickness of the cortex increases, and the diameter of the tibia bone decreases. Therefore, the tibia bone is strong enough in a small diameter area. The cortex is denser than other parts of the bone. The passage of the beam through the dense area is more difficult than the less dense area. As a result, as approaching the distal of the tibia, the passage of the beam decreases so is observed a circle with a smaller diameter in the distal. By reducing the diameter of the circle, the position of the nail hole can be determined more accurately.

Conclusion

IMN holes laser indicator is a non-invasive accurate method that reduced surgical time and simplifies the process. This is an X-ray-free technique and prevents patient and surgical team from being exposed to X-ray. This new technology makes it easier to determine the drilling position and interlocking IMN especially distal interlocking which is a big problem for orthopedic surgeons. This new technology uses 680 nm visible light which has optimal transmission, reflectance, and absorption properties that can be seen with the naked eye. This instrument paves the way for development of effective and safe intramedullary nailing in clinical applications.