One of the biggest challenges of glaucoma drainage device surgery is the avoidance of hypotony in the early post-operative phase. The bleb resistance surrounding glaucoma tube plate tends to have minimal resistance to outflow to aqueous drainage until around 6 weeks post-surgery. Should the flow through the glaucoma tubes be unrestricted before that time, there is a real risk of severe hypotony causing vision-threatening complications such as choroidal haemorrhage1 or flat anterior chamber. In UK, for a long period in the 1980s, glaucoma drainage devices were rarely used by glaucoma specialists due to the poor understanding of controlling flow in the early post-operative phase and the common occurrence of supra-choroidal haemorrhage. Over the intervening decades, the emergence of new glaucoma devices with flow control mechanisms as well as better understanding of ways to control the drainage flow meant that tube surgery became popular again in UK in the 1990s.

The pioneering works by Tony Molteno2 in the 1960s which led to the introduction of Molteno tubes has revolutionised glaucoma tube surgery, particularly in those patients with refractory glaucoma. Recent tube vs trabeculectomy study (TVT) data have shown its superiority compared to traditional trabeculectomy with Mitomycin C in those patients which are considered a higher risk group after both 13 and 5 years follow-up.4 However, the subsequent primary tube vs trabeculectomy study (PTVT) with 1 year follow-up data did not re-affirm the superiority of Baerveldt tube vs trabeculectomy with Mitomycin C in low risk patients.5 Nonetheless, glaucoma drainage devices are now an indispensable tool in our management of patients in the glaucoma service.

Although significant advances in tube design have been made since Molteno unveiled his design in the 1960’s,1 all the current effective tubes are essentially made of the similar design first pioneered by Molteno, in that, they all consist of a plate element which created a bleb or aqueous reservoir and a tube element which is inserted into the anterior chamber. The Baerveldt implant was introduced in 19906 and has a larger plate than a Molteno device but it does not have a flow restrictor. While the Ahmed glaucoma valve, which was introduced in 1993, does have a Venturi-valve system to regulate flow in the early post-operative period to prevent hypotony.7 The main design features and the material used are summarised in Table 1. Most of the tubes and plates are made out of silicone as the primary material, and differ only by the presence, or otherwise, of a flow resistor.

Table 1 Contemporary glaucoma drainage devices

The Ahmed glaucoma valve was the first glaucoma drainage device with a flow resistor made out of folded over silicone membrane and pre-tensioned by the plate casting to open and close at a certain level of intraocular pressure (Figure 1). It also utilises the Venturi effect to enable the valve to close at a certain pressure level. An early study on flow control in glaucoma valves by Prata8 appeared to contradict the claim made by the manufacturer regarding the presence and effectiveness of the valve mechanism in the Ahmed glaucoma valve design. However, subsequent studies by Porter, Francis, and Eisenberg appeared to have confirmed the presence of a valve system in the Ahmed glaucoma valve.9, 10, 11

Figure 1
figure 1

A schematic drawing of the resistance mechanism of the Ahmed valve. Aqueous flows (black arrow) through the tube into a chamber within the plate element. This chamber is formed by a folded over silicone membrane (black line) with its free edges forming a one-way valve. Manufacturers claim that the two halves of the polypropylene body of the plate element pre-tension the valve to open at a specific level of IOP. They also claim that the venturi effect produced by the tapering trapezoidal shape of the space enclosed by the folded silicone membrane acts to improve flow regulation (increasing fluid velocity as the chamber tapers act to reduce internal pressure proximal to the slit opening in accordance with the inverse relationship between fluid velocity and pressure expressed in the Bernoulli’s theorem).

The main difference in the Prata study compared to the latter study in terms of their flow rig set-up is the absence of a reservoir to allow the pressure to develop slowly in the flow rig system using an open stopcock system. When the Ahmed glaucoma valve was tested in the Prata’s closed flow rig system, the pressure built up instantaneously to a high level, thereby forcing the valves to open almost instantaneously. This made the Ahmed glaucoma valve appear not to have any flow resistor.

However, in the late 1990s the design of the Ahmed glaucoma valve was changed from polypropylene material to construct the plate to a softer silicone plate material. Since then, uncertainty regarding the consistency of the opening and closing pressure of Ahmed valves again re-surfaced. This theory was tested by Moss et al in the study published in 2008.12 They use a gravity driven flow system to test the opening and closing pressure of Ahmed valve and found that there was significant variability in the closing pressure. Three out of the six did not appear to close, even if the pressure is below 6 mm Hg. This would appear to confirm the clinical picture of the frequent post-op hypotony encounter in some Ahmed valve patients after surgery. Surgeons in that period tend to inject viscoelastic material into the anterior chamber at the end of Ahmed valve surgery to prevent hypotony. The reason for the apparent change in the consistency of the flow mechanism in Ahmed glaucoma valve may be related to the change in the plate material used. Polypropylene is a much stiffer plastic material compared to the newer version of FP7 which uses silicone plate. It is probable that the polypropylene cast was able to exert a more precise opening and closing pressure for the Ahmed valve than a softer silicone casting would have.

Due to this discrepancy, our unit started testing every single Ahmed glaucoma valve implantation from July 2008 using a gravity driven flow rig similar to those described by Moss.13 We use bottles of balanced salt solution and a giving set with a 26 gauge cannula. All the Ahmed valves were primed prior to testing on the rig, the cannula was inserted into the tube opening and the water bottle was elevated or lowered, according to the testing requirement. We tested 24 Ahmed glaucoma valves between 2008 and 2010 out of which we rejected four devices; three were due to hypotony, and one had very high opening pressure. However, in recent years our rejection rate has been much lower than the initial reported rate. The reason for this is unknown, but we speculate that this may be due to more stringent quality control during the manufacturing process.

Flow control in non-valve device

If constriction of tube diameter is to be used as the flow resistance element for non-valved glaucoma drainage devices, the internal dimensions required to provide a given pressure gradient can be calculated using Poiseuille’s equation,14 assuming a constant rate of production of human aqueous, and assuming that flow through the tube is laminar. Aqueous flow in normal subjects during waking hours is thought to be 2.75±0.63 μl/min,15, 16 but can be as low as 1.4±0.19 μl/min while sleeping.17

Turbulent flow occurs in tubes when the Reynolds number reaches a certain level. At this point, the pressure gradient across the tube will increase markedly as more mechanical energy is lost. It was therefore essential that flow remains laminar. It is generally agreed that the lower critical Reynolds number is 2000, above which the transition from laminar to turbulent flow occurs. Reynolds numbers can be calculated by using the following formula:

where Re=Reynolds number, ρ=density of aqueous=1000 kg/m3, d=tube diameter in metre (10–30 μm), u=average flow velocity, and μ=viscosity of the fluid=0.001 N s/m2.

In Figure 2, Reynolds numbers are plotted against a range of possible diameters for short cylinders designed to provide IOP control in the desired range (5–15 mm Hg) at physiological flow rates. It is clear that the Reynolds numbers for all the various tube sizes used are well below 2000.

Figure 2
figure 2

The relationship between Reynolds number and tube diameter at physiological human aqueous flow rate.

The Poiseuille’s equation (Figure 3) is a function of the aqueous viscosity, aqueous flow rate, diameter of tube, and the length of the tube. Aqueous viscosity is usually assumed to be similar to water and as you can see from the formula, the diameter change has significant impact on the pressure gradient due to the relationship shown in the Poiseuille’s equation.

Figure 3
figure 3

Simplified Poiseuille’s formula, as applied by McEwen14 for aqueous outflow. The pressure drop across the tube is given by: , where μ=aqueous viscosity=0.001 N s/m2, l=length (thickness of polymer disc used in this study)=0.075 mm, Q=aqueous flow rate=1.4 μl/min, and d=diameter in metre.

The relationship between tube diameter and length, calculated from the Poiseuille formula, for a pressure drop across the tube of 5 mm Hg at a flow rate of 0.7 μl/min (plain line), 1.4 μl/min (dotted line), and 2.8 μl/min (dashed line) are shown in Figure 4. Therefore, in order to prevent hypotony, the diameter of the tube must be smaller or the length of the tube must be longer that those shown in the graph.18

Figure 4
figure 4

The relationship between tube diameter and length, calculated from the Poiseuille formula, for a pressure drop (gradient) across the tube of 5 mm Hg at a flow rate of 0.7 μl/min (plain line), 1.4 μl/min (dotted line), and 2.8 μl/min (dashed line). where n=aqueous viscosity =0.001 N s/m2, l=length=0.075 mm in this study, Q=aqueous flow rate, and d=diameter in metres.

As the standard tube diameter is too big as in the case of a Baerveldt tube or Molteno tube, the early post-operative flow has to be limited by either a vicryl suture tie19 and/or internal suture stenting.20 The vicryl tie technique usually uses a 7.0 vicryl suture and completely ties off the tube at the end of the surgery. In general, the vicryl sutures tend to loosen and dissolve around week 6 and 8, around the same time when the resistance to flow surrounding the plate has become significant, thereby bringing the pressure down to within a reasonable range without hypotony. However, these effects are not always predictable, as shown by the regular occurrence of early post-operative hypotony in the tube vs trabeculectomy study,21 despite the fact that all the tubes were tied with 7.0 vicryl suture at the end of the procedure.

Internal suture stenting technique was first described by Egbert and Liebert.20 We have described a technique similar to those described by Egbert and Liebert in a video presentation in 2005 at an American Academy of Ophthalmology meeting. Using this technique, we inserted the Supramid suture for the whole length of the tube including the portion protruding into the anterior chamber. This technique utilises a tight scleral tunnel which provides extra constriction around the Supramid tube and thereby eliminates all flow through the tube in the early operative period. The Supramid can then be removed in stages postoperatively, depending on the pressure level of the eye. Obviously the length of the Supramid in the tube will determine the pressure gradient in the eye, again following the Poiseuille’s formula to a certain extent.

New generations of glaucoma tube implant

More recently, various advancements have been made in glaucoma tube development in particular three implants utilising the Poiseuille’s formula for flow restriction. These include the XEN implant (Allergan Inc, CA, USA), MicroShunt (Senten, CA, USA), and eyeWatch implant (Rheon Medical, Switzerland).

The XEN implant is 6 mm in length, initially launched with two diameters of either 140 or 45 μm. Using the estimation as illustrated in Figure 4, one would assume that a 6 mm tube will require around 50 μm diameter to prevent hypotony. This prediction was confirmed by the early hypotony rate in the preliminary study which showed that the XEN with 145 μm diameter encountered 54% of hypotony while the XEN 45 micron only had 13% hypotony rate on day 1.22 MicroShunt also utilise similar formula and its tube has a lumen of 70 μm with a length of 8.5 mm. A laboratory study performed by the designer have shown that this combination will provide significant amount of outflow resistance to prevent hypotony in the early post-operative phase as predicted by using the Poiseuille’s equation as stated earlier in Figure 4.23 However, none of these two implants address the problem of resistor in series. In theory, once the bleb resistance has occurred the flow restrictor within the tube actually becomes redundant. In fact, the residual resistance provided by the tube might be detrimental in achieving lowest possible IOP once the bleb resistance have formed surrounding the plate. This issue of residual resistance can potentially be tackled by the eyeWatch implant,24 the eyeWatch implant utilises a magnetic compression mechanism onto the external portion of the tube. The amount of squeeze on the tube can be adjusted postoperatively in the clinic using a magnetic device. It is therefore easier to control the amount of resistance provided by the tube at any time point and one can potentially remove all resistance through the tube once the bleb resistance has formed in the post-operative period using the magnetic device in the out-patient, thereby achieving the lowest possible intraocular pressure level.

Conclusion

In conclusion, the Ahmed glaucoma valve does have a flow control mechanism that utilises the Venturi principle to regulate flow but can be inconsistent in some implants, although this issue appears to have been addressed by the manufacturer in recent years. Non-valve devices all utilise the Poiseuille’s equation as a principle of flow restriction. These include new generations of glaucoma implants such as XEN, eyeWatch, and MicroShunt. However, issues with residual resistance remains in most of these implants once the bleb has developed a significant amount of outflow resistance. Currently, only the eyeWatch implant provides an easily reversible flow restrictor mechanism in their tube and this could potentially transform the management of glaucoma in the future.