Abstract
A feasible, inexpensive, rapid, and easy-to-use method to measure vestibular vertical movement perception is needed to assess the sacculus-mediated low-frequency otolith function of dizzy patients. To evaluate the feasibility of reaction time assessment in response to vertical motion induced by an elevator in healthy young individuals. We recorded linear acceleration/deceleration reaction times (LA-RT/LD-RT) of 20 healthy (13 female) subjects (mean age: 22 years ± 1 SD) as a measure of vertical vestibular motion perception. LA-RT/LD-RT were defined as the time elapsed from the start of elevator acceleration or deceleration to the time at which subjects in a sitting position indicated perceiving a change in velocity by pushing a button with their thumb. The light reaction time was measured as a reference. All 20 subjects tolerated the assessment with repeated elevator rides and reported no adverse events. Over all experiments, one upward and four downward rides had to be excluded for technical reasons (2.5%). The fraction of premature button presses varied among the four conditions, possibly related to elevator vibration (upward rides: LA-RT-up 66%, LD-RT-up 0%; downward rides: LA-RT-down 12%, LD-RT-down 4%). Thus LD-RT-up yielded the most robust results. The reaction time to earth-vertical deceleration elicited by an elevator provides a consistent indicator of linear vestibular motion perception in healthy humans. The testing procedure is inexpensive and easy to use. Deceleration on upward rides yielded the most robust measurements.
Similar content being viewed by others
Introduction
The perception of one’s own head and/or body movement in space is critical for gait, posture control, and spatial orientation1. This self-motion perception results from the integration of vestibular, visual, and somatosensory information2. Among the sensors in the inner ear, the otolith organs—the sacculus and the utriculus—provide the necessary signal for the perception of linear acceleration or tilt of the head relative to gravity3. Due to its orientation within the skull, the sacculus is more sensitive to earth-vertical linear translation (e.g., elevator acceleration) when the head is upright, whereas the utriculus is more sensitive to horizontal linear translation (e.g., car acceleration)4.
The most widely used auxiliary test of otolith function used by neuro-otologists are vestibular-evoked myogenic potentials (VEMPs)5,6. VEMPs, however, have several limitations7, including being technically complex, time-consuming, and relatively expensive. Moreover, VEMPs demonstrate significant response variabilities that depend on non-vestibular factors, such as hearing acuity, skull thickness, and the degree of tonic muscle activation7. While ocular VEMPs (oVEMPs) reflect utricular high-frequency function, low-frequency utricular function is typically assessed with subjective visual vertical (SVV) test8. In routine saccular assessment, however, only high-frequency function is measured using oVEMPs. In other words, a low-frequency sacculus-mediated equivalent to the mainly utriculus-mediated SVV is missing. Thus, a simple, easy to employ, and relatively low-cost method to assess low-frequency vertical otolith function could be a valuable element of the vestibular testing battery for evaluating dizzy patients.
As gravity is equivalent to a constant earth-vertical acceleration, an earth-vertical body movement only increases (upward acceleration or downward deceleration) or decreases (downward acceleration or upward deceleration) the length of the gravito-inertial (GI) vector, but not its direction. With subjects in the upright position, sensing GI-vector length changes by the otolith organs relies predominantly on the function of the saccular maculae due to their orientation relative to the skull9. As is the case for any vestibular stimulus, other sensory inputs, such as vision, proprioception, and hearing, contribute to the perception of linear motion. Thus, disorders of saccular pathways and pathways of other involved sensory inputs may increase the threshold of vertical linear motion perception10.
Motion platforms has been successfully employed utilized to provide earth-vertical motion stimuli; they are, however, technologically complex and expensive and therefore their clinical application is rare11,12,13. A suitable elevator, such as one used in hospitals for the transport of beds, could provide a simple, easy to employ, and relatively low-cost alternative to clinically assessing vertical linear motion perception in patients, including subjects with suspected deficits in saccular pathways.
Reaction times are frequently used in auxiliary neurological assessments, e.g., in patients with suspected neurocognitive impairment or concussion14,15. Reaction times could also permit the evaluation of linear self-motion perception and thus provide a valuable vestibular screening tool for the detection of otolith dysfunction, such as in patients with chronic peripheral vestibular hypofunction16 or with otolith hypersensitivity caused by vestibular migraine17.
To our knowledge, no studies have attempted to assess vestibular perception via reaction time to elevator-induced vertical acceleration and deceleration. This study evaluated the feasibility of a linear acceleration/deceleration reaction time (LA-RT or LD-RT) assessment in response to low-frequency vertical motion induced by a hospital elevator in healthy young individuals.
Results
Feasibility of elevator rides
There were no adverse events such as dizziness or disorientation. All participants confirmed that the assessment procedure was tolerable and acceptable for being repeated several times. In every subject, the testing, including preparation, could be performed in less than 20 min. One upward ride and four downward rides were excluded for technical reasons (2.5%). The acceleration curves of 25 upward and 25 downward elevator rides showed excellent reliability: curves of upward had an ICC score of 0.986 (CI 0.986–0.987) and a Cronbach’s alpha of 0.986, while curves of downward had an ICC score of 0.990 (CI 0.990–0.991) and a Cronbach’s alpha of 0.990. A typical acceleration curve during an upward elevator ride is shown on Fig. 1.
Feasibility of reaction time measurements
After removing outliers, normality tests showed no normal distribution of reactions times upon elevator linear accelerations (LA-RT) and elevator linear decelerations (LD-RT). Boxplots of all LA-RT and LD-RT raw data with outliers can be found in Fig. S1 of the supplementary file. The data distribution of the LA-RT and LD-RT of the upward and downward elevator rides is given in the Table 1.
Negative reaction times (premature button press) are reactions that occurred before the actual start of acceleration or deceleration (i.e., too early). Positive reaction times occurred after the actual onset of acceleration or deceleration (i.e., they likely reflected the reaction to the elevator motion). The flowchart of data analysis is shown in Fig. 2.
The five upward rides showed a significant change with ride repetition in LA-RT-up (x2 = 13.40, df = 4, p = 0.009) and a significant effect of LD-RT-up (x2 = 16.58, df = 4, p = 0.002). The five downward rides showed no effect on LA-RT-down (x2 = 6.33, df = 4, p = 0.176) and significant effect of LD-RT-down (x2 = 12.33, df = 4, p = 0.0.015). The mean reaction in all five rides and two direction times are shown in Fig. 3a–d.
Discussion
During all rides, the participants performed the tasks of indicating elevator motion, i.e., change of elevator velocity, without any difficulties. All participants confirmed the tolerability and acceptability of the repeated assessment, and the elevator rides showed excellent reliability. The vibration of the elevator at the start of the rides probably caused many premature button presses by the participants. The measures of deceleration during the upward elevator rides, however, yielded consistently valid reaction times with the least variability. This is probably related to the fact that the least noisy change in gravito-inertial acceleration (i.e., jerk) was detected during deceleration of the upward moving elevator.
Since the change in acceleration during elevator rides is slow and thus provides a low-frequency stimulus (half a period is around three seconds, or 0.16 Hz), subjects may have difficulty to distinguish derivative of elevator acceleration or deceleration (jerk) from noise due to motor vibration. We hypothesize that the reaction time upon elevator acceleration or deceleration (Fig. 4, reaction time) is composed of two parts: (1) The time that elapses from the beginning of elevator acceleration or deceleration to the moment when the perceptual threshold is reached; (2) the time that elapses between reaching the threshold and the button press. Conceivably, the second part (Fig. 4, asterixs) is equivalent to the light reaction time. The acceleration value at the time of the button press was in the range of the perception threshold found in a study using a two-interval forced-choice task when we take the light reaction time for the time elapsing between motion perception and button press (see hypothetical values in the legend of Fig. 4)18. Since we are not able to vary the acceleration and deceleration parameters of the elevator, our hypothesis can only be tested by measuring subjects with increased perceptual thresholds, i.e., patients with deficits of the sacculi or their afferents.
Over the course of the five elevator rides, the LD-RT significantly increased. This could be the result of habituation19 as well as fatigue. Habituation is an increase in reaction that can frequently be observed with repeated sensory stimulation20. Since other confounding sensory inputs were minimised and the motor output (button press) remained the same, the increase in LD-RT may have reflected vestibular threshold habituation.
An important aspect not investigated in this study was whether the vibrations of the elevator were transmitted to the chair during the accelerations. Furthermore, elevators may, of course, have different motion characteristics. Therefore, it is important to evaluate the magnitude and repeatability of acceleration/deceleration profiles generated by each individual elevator. One must note that, unlike cervical vestibular-evoked myogenic potentials (cVEMPS), an elevator test—as all vertical motion devices—cannot assess unilateral saccular function. Another limitation of our study may be that the inertial measurement unit (IMU) was placed on the floor of the elevator. Therefore, the potential confounding movement of the head was not assessed. However, subjects were instructed to keep their heads still to maximise saccular activation and to minimise any extra-saccular activation. Another weakness of the elevator procedure is that it is possible that the exact threshold is not measurable, as the velocity of the elevator may change too quickly. An ongoing study is addressing whether the acceleration/deceleration of the elevator is indeed too fast for vestibular patients with increased thresholds.
In a recently published study, the otolith system’s ability to perceive self-motion was tested by investigating adaptive responses to asymmetric off-axis vertical rotation. The results revealed that after conditioning in the upright position, adaptive perceptual errors increased by 50%21. This raises the possibility of further perceptual errors that should be explored in subsequent studies.
Our approach to saccular stimulation is inexpensive, not time consuming, and easy to apply, with little requirement for operator training22. Indeed, for clinical purposes, rapid—even if imperfect—can greatly assist clinical diagnosis without complex and time-consuming motion platform-based assessment11,12. Deceleration of the elevator (upward) yielded the most valid measurements, as jerk motion was minimal. In addition, short-duration assessment is advantageous to minimise the risk of vertigo or dizziness and to minimise any effect of attention deficits, which have been reported in vestibular patients12.
Conclusion
The reaction time to earth-vertical deceleration elicited by an elevator provides a consistent indicator of linear vestibular motion perception in healthy humans. The testing procedure is inexpensive and easy to use. Deceleration of the elevator during upward rides yielded the most robust measurements, with no premature indication of perceived deceleration by the subjects. Further evaluation of repeatability and threshold responses is warranted, in addition to comparison with the responses of patients with relevant vestibular dysfunctions.
Methods
Participants
Twenty (7 male) healthy young adults aged between 18 and 30 years (mean age 22.6 ± 1.2 years; mean height: 176.4 ± 11.2 cm; mean weight: 67.1 ± 12.4 kg; mean BMI: 21.4 ± 2 kg/m2) were recruited having provided written informed consent to participate in the study approved by the ethics committee of the Canton of Zurich under BASEC 2019-01759 and registered at ClinicalTrials.gov Identifier: NCT04200820 at 16/12/2019. Testing was carried out in accordance with relevant guidelines and regulations. The exclusion criteria were as follows: (1) acute pain, (2) chronic neck pain, (3) previous neck surgery, (4) any vestibular disorder, (5) fear of elevators or claustrophobia, (6) gait disorders putatively attributed to causes other than vestibular dysfunction, and (7) a score of > 30 on the dizziness handicap inventory (DHI). The DHI is a validated, 25-item, self-reported questionnaire used to evaluate the self-perceived handicapping effects caused by dizziness23. The higher the score, the greater the degree of handicap. All testing was performed at the Department of Neurology, University Hospital Zurich, Switzerland.
Study procedures
Participants visited the department on a single occasion for approximately 30 min. Each participant was exposed to five upward and five downward elevator rides. Linear acceleration reaction times (LA-RT) were recorded on the upward (A-up) and downward (A-down) accelerations, and linear deceleration reaction times (LD-RT) were recorded on the upward (D-up) and downward ride decelerations (D-down). Following completion of the elevator rides, participants were asked if they had experienced any adverse events or symptoms.
Prior to the lift rides, the light reaction time was recorded. During the lift ride, the participants were asked to close their eyes and wear a (white) noise-cancelling headset. To minimise lower limb proprioceptive sensory information, participants were seated on foam, and their feet were placed on foam blocks (Airex balance pads, 3A Composites Holding AG, Switzerland)24 (Fig. 5). The vestibular vertical movement perception measurements were performed in a 16-floor bed elevator in a hospital building (Nord1, University Hospital Zurich, Switzerland). Data collection was instigated prior to the beginning of the elevator ride and terminated once the elevator had made a full stop on the respective floors (1 and 16). The acceleration and deceleration duration was 3.15 s, with a peak acceleration of 0.07 m/s2.
Elevator rides are clearly not optimal for inducing linear acceleration or deceleration due to accompanying vibrations at the start and end of the elevator motion. Therefore, the reliability of the acceleration curve of elevator rides and the variability of acceleration or deceleration of the bevor at the start and end were tested to determine which of the measurements was feasible. This study followed the STROBE checklist.
Linear velocity reaction time (LVRT)
LVRT was recorded via an inertial measurement unit (IMU) (FXOS8700 CQ from NXP mounted on an adafruit (“NXP Precision 9DoF” Breakout board) to record vertical acceleration connected to two Bluetooth remote controls (Satechi ST-ARCM) (Fig. 5). The master remote control was operated by the experimenter, instigating and terminating the recording. The second remote control was used by the log the state of the participant’s trigger button (red), which they pressed with their thumb when they sensed motion (Fig. 6). A Raspberry Pi CPU was used to record linear acceleration and trigger/controller status.
Acceleration/deceleration reaction time (LA-RT or LD-RT)
LA-RT or LD-RT was defined as the time elapsed from the start of the elevator acceleration or deceleration to the time that the subjects indicated a perception of a change of velocity by pressing their button. Figure 7 explains how LD-RT would be become longer as the perception threshold increased due to saccular deficits. The same mechanism would also apply to vestibular habituation. Future studies should investigate whether further habituation occurs and at what point responses plateau.
Light reaction time test (LRTT)
LRTT is defined as the time that elapses from the random switch-off of a blue LED until the participant reacts by pressing the red button with the thumb as quickly as possible25. After a trial run, the test was performed twice and averaged. The literature reports that LRTT is typically shorter than LA-RT/LD-RT26. For confirmation, LRTT was compared with the LA-RT/LD-RT results.
Data processing
LVRT data (150 Hz) were processed with a script in Python 3 (Python Software Foundation, http://www.python.org) using the Pandas Package. To reduce the impact of shot noise, an average running filter with a 300 ms window was applied. Subsequently, the acceleration and declaration peaks were identified using the find_peak algorithm from the scipy signal package. The start and end times of accelerations and decelerations were defined as the time points when the denoised signal reached 95% of its peak value, respectively. The time points of the participants’ reaction times were identified by the button press with respect to the start/end times.
Statistical analysis
Feasibility elevator
The reliability of the acceleration curve of elevator rides (upward and downward) was assessed with an intraclass correlation coefficient (ICC) with a 95% CI. Cronbach’s alpha was assessed to evaluate internal consistency.
Feasibility LA-RT and LD-RT
Data normality was tested using the Shapiro–Wilk test after removing outliers (× 3 interquartile range). The effect of trial repetition (N = five trials) was evaluated using the Friedman test, with a significance level of α = 0.05. Two LRTTs were performed; the first two LA-RT/LD-RT measurements were used for comparison. Correlation of mean was assessed via Spearman testing. Correlation coefficients were interpreted as follows: r < 0.50, weak relationship; 0.50 ≤ r ≤ 0.74, moderate relationship; and r ≥ 0.75, strong relationship27. All statistics were performed with SPSS Statistics 25 for Windows (IBM Inc., USA).
Data availability
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Britton, Z. & Arshad, Q. Vestibular and multi-sensory influences upon self-motion perception and the consequences for human behavior. Front. Neurol. 10, 63 (2019).
Ferre, E. R., Walther, L. E. & Haggard, P. Multisensory interactions between vestibular, visual and somatosensory signals. PLoS ONE 10(4), e0124573 (2015).
Valko, Y., Lewis, R. F., Priesol, A. J. & Merfeld, D. M. Vestibular labyrinth contributions to human whole-body motion discrimination. J. Neurosci. 32(39), 13537–13542 (2012).
Jamali, M., Sadeghi, S. G. & Cullen, K. E. Response of vestibular nerve afferents innervating utricle and saccule during passive and active translations. J. Neurophysiol. 101(1), 141–149 (2009).
Clarke, A. H., Schonfeld, U. & Helling, K. Unilateral examination of utricle and saccule function. J. Vestib. Res. 13(4–6), 215–225 (2003).
Kingma, H. Function tests of the otolith or statolith system. Curr. Opin. Neurol. 19(1), 21–25 (2006).
Rosengren, S. M., Colebatch, J. G., Young, A. S., Govender, S. & Welgampola, M. S. Vestibular evoked myogenic potentials in practice: Methods, pitfalls and clinical applications. Clin. Neurophysiol. Pract. 4, 47–68 (2019).
Hosli, S. & Straumann, D. Independent measures of utricular function: ocular vestibular evoked myogenic potentials do not correlate with subjective visual vertical or fundus photographic binocular cyclorotation. Front. Neurol. 12, 658419 (2021).
Curthoys, I. S., Uzun-Coruhlu, H., Wong, C. C., Jones, A. S. & Bradshaw, A. P. The configuration and attachment of the utricular and saccular maculae to the temporal bone. New evidence from microtomography-CT studies of the membranous labyrinth. Ann. N. Y. Acad. Sci. 1164, 13–18 (2009).
Merfeld, D. M., Priesol, A., Lee, D. & Lewis, R. F. Potential solutions to several vestibular challenges facing clinicians. J. Vestibul. Res.-Equil. 20(1–2), 71–77 (2010).
Bermudez Rey, M. C. et al. Vestibular perceptual thresholds increase above the Age of 40. Front. Neurol. 7, 162 (2016).
Dupuits, B. et al. A new and faster test to assess vestibular perception. Front. Neurol. 10, 707 (2019).
Wright, W. G., DiZio, P. & Lackner, J. R. Vertical linear self-motion perception during visual and inertial motion: more than weighted summation of sensory inputs. J. Vestib. Res. 15(4), 185–195 (2005).
Kelly, K. M. et al. Oculomotor, vestibular, and reaction time effects of sports-related concussion: Video-oculography in assessing sports-related concussion. J. Head Trauma Rehabil. 34(3), 176–188 (2019).
Christ, B. U., Combrinck, M. I. & Thomas, K. G. F. Both reaction time and accuracy measures of intraindividual variability predict cognitive performance in Alzheimer’s disease. Front Hum Neurosci 12, 1–10 (2018).
Swanenburg, J., Buchi, F., Straumann, D., Weber, K. P. & de Bruin, E. D. Exergaming with integrated head turn tasks improves compensatory saccade pattern in some patients with chronic peripheral unilateral vestibular hypofunction. Front. Neurol. 11, 601 (2020).
King, S. et al. Self-motion perception is sensitized in vestibular migraine: pathophysiologic and clinical implications. Sci. Rep. 9, 803 (2019).
Nesti, A., Barnett-Cowan, M., Macneilage, P. R. & Bulthoff, H. H. Human sensitivity to vertical self-motion. Exp. Brain Res. 232(1), 303–314 (2014).
Binder, M. D. Habituation/Dishabituation Test (Springer, 2009).
Norre, M. E. & De Weerdt, W. Treatment of vertigo based on habituation. 1. Physio-pathological basis. J. Laryngol. Otol. 94(7), 689–696 (1980).
Faralli, M. et al. Adaptive perceptual responses to asymmetric rotation for testing otolithic function. Exp. Brain Res. 240(7–8), 2017–2025 (2022).
Swanenburg, J., Gruber, C., Brunner, F. & Wirth, B. Patients’ and therapists’ perception of change following physiotherapy in an orthopedic hospital’s outpatient clinic. Physiother. Theory Pract. 31(4), 293–298 (2015).
KurreAnnette, B. C. H., van GoolChristel, J. A. W., Thomas, G.-J., De BruinEling, D. & Dominik, S. Exploratory factor analysis of the Dizziness Handicap Inventory (German version). BMC Ear Nose Throat Disord. 10(3), 1–10 (2010).
Asslander, L. & Peterka, R. J. Sensory reweighting dynamics following removal and addition of visual and proprioceptive cues. J. Neurophysiol. 116(2), 272–285 (2016).
Amini Vishteh, R., Mirzajani, A., Jafarzadehpour, E. & Darvishpour, S. Evaluation of simple visual reaction time of different colored light stimuli in visually normal students. Clin. Optom. 11, 167–171 (2019).
Barnett-Cowan, M. & Harris, L. R. Perceived timing of vestibular stimulation relative to touch, light and sound. Exp. Brain Res. 198(2–3), 221–231 (2009).
Portney, L. G. & Watkins, M. P. Foundations of Clinical Research: Applications To Practice (Pearson/Prentice Hall, 2015).
Funding
This study was supported by the Dr. M.K. Dabbous Foundation.
Author information
Authors and Affiliations
Contributions
J.S. developed the research question. J.S. and D.S. established the concept and design parts. S.S. performed the data acquisition. S.S., J.S., P.S., and D.S. performed the analysis and interpretation of the results. S.S. produced an early version of the manuscript. J.S., D.G., P.S., Y.Z., and D.S. edited the manuscript. All authors have read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
David Green is employed by KBR GmbH, Cologne, Germany. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Schellenberg, S., Straumann, D., Green, D.A. et al. Earth-vertical motion perception assessment using an elevator: a feasibility study. Sci Rep 13, 9450 (2023). https://doi.org/10.1038/s41598-023-36655-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-36655-7
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.