Introduction

EPImix

A new multicontrast magnetic resonance (MR) pulse sequence has been developed, named EPImix1. This sequence utilises single-shot echo-planar imaging (EPI) and different magnetisation preparations to rapidly generate six contrasts; T1-FLAIR (fluid attenuated inversion recovery), T2-FLAIR, T2-weighted, T2*-weighted, diffusion-weighted imaging (DWI) and apparent diffusion coefficient (ADC) (for full acquisition details, see Skare et al.1). The EPImix sequence has lower signal-to-noise (SNR) ratio and resolution than standard MRI sequences1, resulting in lower image quality ratings than routinely collected corresponding MR scans2,3. However, the main benefit of this single-shot, multicontrast sequence over routinely acquired individual MRI sequences is its speed. The EPImix sequence is faster than corresponding single-contrast sequences because of its lower matrix size and because an EPI readout is more SNR-efficient than analogous fast spin echo (FSE) readouts1. This multimodal sequence can acquire full brain coverage in 78 seconds, relative to the ~750 seconds needed to collect these six contrasts using standard MRI sequences2,3. In addition to being more cost effective4, shorter scanning times improve participants’ comfort and reduce motion5, potentially resulting in less unusable data, especially for clinical populations6. The EPImix sequence also generates multiple types of MR contrasts, therefore capturing more tissue characteristics than standard parametric mapping2,3.

Quantitative analyses

To establish whether this new sequence has potential to reduce scanning time whilst producing similar derived measures, it needs to be quantitatively compared to corresponding routinely acquired MR scans. (In this context, the terms quantitative and qualitative refer to the type of analysis used, rather than the type of MR scan. Specifically, qualitative measures refer to visual inspection of scans, whereas quantitative measures refer to analyses which employ numerical computations.) Previous research established that EPImix scans are comparable to conventional scans in the context of qualitative clinical diagnosis2,3. Moreover, our previous work compared EPImix and routine T1-weighted (T1-w) scans (Note that the EPImix sequence includes a T1-FLAIR contrast, while the “conventional” single-contrast scans were acquired using an IR-FSPGR sequence. Still, as both sequences are T1-weighted, we refer to both as such (as well as simply “T1-w”).) using simple rapidly-derived quantitative measurements, including image intensity and Jacobian determinants (obtained from the registration to Montreal Neurological Institute (MNI) standard space), and found high similarity in these rudimentary quantitative measures as well as their potential to identify inter-individual differences7. (Note that the EPImix sequence includes a T1-FLAIR contrast, while the “conventional” single-contrast scans were acquired using an IR-FSPGR sequence. However, both sequences are T1-weighted and are referred to as such [as well as simply “T1-w”].) We further demonstrated the utility of the multicontrast EPImix sequence to construct morphometric similarity networks (MSNs), which provide individual estimates of anatomical connectivity8, in minutes following participants entering the MRI scanner7. However, EPImix scans have not previously been compared to standard sequences using established quantitative structural neuroimaging measures such as estimates of tissue volume9 or predicted brain-age, a putative biomarker of brain health10,11. Such quantitative measures likely reflect subtle inter-individual differences and provide a more thorough comparison between EPImix and standard sequences than qualitative measures12 or potentially noisy estimates such as tissue intensity13. Moreover, the ability to derive commonly used quantitative measures from data acquired with faster imaging sequences, such as EPImix, has the potential to reduce scanning time, increasing participant comfort and reducing costs.

Active acquisition

One potential application of the EPImix sequence is in active acquisition. Active acquisition is a proposed type of data acquisition which would utilise active learning14 to analyse MRI data as it is acquired, with results used to guide further image acquisition, in a closed-circuit sequence15,16. This would remove the necessity of making a priori decisions about scanning parameters such as the type of scan, scan resolution and/or the scan location. Instead, these would be dependent on the individual inside the scanner, driving the selection of scanning sequences towards identification of individual differences or personalised clinical diagnosis. In this way, active acquisition holds the potential for reduced scan time and improved accuracy, reliability and individualisation of (neuro)imaging16. The main feasibility obstacles to active acquisition are the speed of image collection as well as data processing and analysis; to realise the benefits of this multimodal adaptive approach, analyses need to be carried out in near real-time7,16. Due to the relative speed of the new EPImix sequence, it could contribute considerably to this process.

Quantitative comparison of EPImix and standard T1-w contrasts

Here, we focused on the T1-w EPImix contrast and compared it to a standard T1-w contrast at two levels. We first compared T1-w global and voxel-wise estimates of grey matter (GM), white matter (WM) and cerebrospinal fluid (CSF) tissue volumes9. These tissue volumes vary as a function of age and are typically abnormal in many psychiatric and neurological populations17,18.

We then compared predicted brain-ages from EPImix and standard T1-w scans to assess whether EPImix can reliably detect the well-established relationship between healthy aging and changes in tissue volumes17,19. Predicted brain-ages are derived from a validated multivariate regression model that estimates the biological age of an individual brain directly from the imaging data. This estimate provides a useful summary of the large voxel-wise spaces generated by standard high-resolution T1-w scans into a single interpretable value. The difference between the predicted brain-age and the true chronological age is a versatile measure that has been associated with cognitive ability11,20 as well as clinical status and severity, with significantly “older” brain-ages associated with traumatic brain injury (TBI)21, mild cognitive impairment (MCI) and Alzheimer’s disease (AD)22. Furthermore, predicted brain-age could be a better predictor of disease risk than chronological age20,23. For example, brain-age has been shown to be a reliable predictor of which individuals will develop AD24,25.

We expected quantitative measurements derived from EPImix T1-w scans, including global and voxelwise estimates of tissue volume as well as predicted brain-age, to be broadly comparable to analogous estimates derived from standard T1-w scans. Given the notable differences in the field of view (FoV) between EPImix and standard T1-w scans, the tissue volume and brain-age analyses were repeated with standard T1-w scans that had their FoV artificially reduced to match the FoV of the EPImix T1-w contrast. Finally, we assessed the within-session test-retest reliability of the EPImix-derived estimates of tissue volume and predicted brain-age.

Methods

Participants

EPImix scans (including a T1-w contrast) were acquired for 95 healthy participants across three studies that used the same MRI scanner; one participant with particularly reduced cortical coverage (due to reduced FoV of EPImix scans, discussed below) was excluded, resulting in EPImix scans from 94 participants (F: 47, M: 47) used for analyses. The age range was 18-59 years. The mean age of participants was 28.2 years, with a standard deviation of 9.2 years (female mean age: 27.5 ± 8.6 years; male mean age 28.9 ± 9.7 years). Of those 94 participants, 64 (F: 32, M: 32) were also scanned using a standard high-resolution T1-w sequence. The age range of this subset of participants was 18-59 years; their mean age was 28.9 years, with a standard deviation of 10.1 years (female mean age: 28.6 ± 9.6 years; male mean age: 29.3 ± 10.5 years); for details, see Supplementary Information (SI) Fig. S1. All three studies received ethical approval from King’s College London’s Psychiatry, Nursing and Midwifery Research Ethics Committee (KCL Ethics References: HR-18/19-9268, HR-18/19-11058 and HR-19/20-14585). All participants gave written informed consent to take part in their respective study. All experiments were performed in accordance with relevant guidelines and regulations.

MRI acquisition

All scans were collected on the same General Electric (GE) MR750 3T scanner (Waukesha, WI). EPImix scans were acquired from 94 participants, consisting of six contrasts (T2*, T2-FLAIR, T2, T1-FLAIR, DWI, ADC). Only the T1-FLAIR contrast was used in this study (referred to as the EPImix T1-w contrast), which was acquired with the following parameters: TE = 16.5 ms, TR = 1300 ms, TI = 582 ms, flip angle = 90°, matrix size = 180 × 180, FoV 240 mm, 32 slices, slice thickness = 3 mm, voxel resolution = 0.975 × 0.975 × 3 mm, acquisition time 1 min 10 s. The EPImix sequence includes an in-scanner motion correction step1 and we used the motion-corrected images for analyses. For details regarding the EPImix sequence, see Skare et al.1. Additionally, for 10 participants a second EPImix scan was acquired during the same session, which was used to quantify test-retest reliability.

Additionally, conventional T1-w scans were acquired for 64 participants within the same session, using an inversion recovery-prepared fast spoiled gradient recalled-echo (IR-FSPGR) sequence. Of these, 12 scans were acquired with the following parameters: TE = 3.172 ms, TR = 8.148 s, flip angle = 12°, matrix size = 256 × 256, FoV = 256mm, 164 slices, slice thickness = 1 mm, voxel resolution 1 × 1 × 1 mm, acquisition time 2 min 54 s; and 52 scans were acquired with the following parameters : TE = 3.016 ms, TR = 7.312 ms, flip angle = 11°, matrix size = 256 × 256, FoV = 270 mm, 196 slices, voxel resolution = 1.05 × 1.05 × 1.2 mm, acquisition time 5 min 37 s.

Pre-processing and brain-age estimation

All brain scans were pre-processed using a SPM12 DARTEL processing pipeline26; including bias field correction, segmentation, registration to standard space (MNI152, 6th generation) via an intermediate study-specific template, and smoothing using a 4mm full width at half maximum (FWHM) kernel. This process generated voxelwise estimates of tissue volumes, including grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). Global tissue volumes were calculated by summing across voxels in each tissue class.

The above processing pipeline was applied as part of the brainageR software, which was additionally used to obtain brain age predictions using voxel-wise estimates of GM, WM and CSF tissue volume. The brain-age model was previously trained to predict chronological age from conventional T1-w scans in 3377 healthy people aged 18-92 years, using Gaussian Processes Regression. For additional information about the brain-age model, see Cole27 and https://github.com/james-cole/brainageR.

The reduced FoV of the EPImix acquisition used in this study resulted in imperfect cortical coverage in some participants; in particular, portions of the inferior temporal and/or superior parietal lobes were not consistently scanned (SI Fig. S2). To investigate the potential impact of this reduced FoV on quantitative analyses, we repeated analyses using standard T1-w scans with artificially reduced FoV, to match the reduced EPImix FoV (within-participants). This was performed using affine registration (FSL FLIRT)28 of the standard T1-w scan to an upsampled version of the same participant’s EPImix T1-w contrast (resampled from the EPImix resolution of 0.975 x 0.975 x 3 mm, to match the standard T1-w resolution of 1 x 1 x 1 mm, or 1.05 x 1.05 x 1.2 mm). Standard T1-w scans with artificially reduced FoV (T1-w FoVEPI) were then processed using the steps described above, for both tissue volume estimation and brain age prediction.

Processing was run on a Dell workstation (16-CPU 3.6GHz Intel Xeon, 128Gb RAM). Total processing time was recorded for each scan, including data processing using SPM12 and brain-age estimation. The processing time of EPImix T1-w scans was: median [1st, 3rd Quartile] (Md [Q1, Q3]) = 4.82 [4.75, 4.9] min; for standard T1-w scans, Md [Q1, Q3] = 7.61 [7.50, 8.17] min; for standard T1-w scans with reduced FoV, Md [Q1, Q3] = 5.42 [5.37, 5.66] min (SI Fig. S3).

Figure 1
figure 1

Global tissue volume estimation across contrasts. Comparison of tissue volumes (grey matter, white matter and cerebrospinal fluid) between T1-w and EPImix (T1-w) scans with full field of view (AC) and between EPImix (T1-w) scans and T1-w scans with reduced field of view (DF). (Spearman’s correlation coefficient rs, p-value, r2 derived from Pearson’s correlation).

Statistical analysis

Quantitative measures derived from EPImix T1-w scans, standard T1-w scans and T1-w scans with reduced FoV from 64 participants were compared using the Spearman’s correlation coefficient (rs). We first compared tissue volumes (for GM, WM and CSF) between sequences, both at the global and voxelwise levels. Voxelwise comparisons were restricted to voxels contained within the FoV of EPImix scans in the majority of participants (i.e. voxels with at least 0.001 mm3 tissue volume in at least 95% of participants).

Furthermore, we quantified the relationship between GM volume and participant chronological age, to ascertain whether EPImix T1-w scans are equally sensitive as standard T1-w acquisitions to known decreases in GM volume in ageing17,19. To investigate whether associations between GM volume and age differ across contrasts, we additionally investigated differences in the strength of this association between pairs of contrasts using bootstrap. Specifically, we re-estimated Spearman’s rs, and r2, in 10’000 samples of 64 participants, sampled with replacement from the original group. We subsequently obtained a distribution of the difference in Spearman’s rs, and r2, between all three pairs of contrasts (i.e. EPImix VS T1-w, EPImix VS T1-w FoVEPI, T1-w VS T1-w FoVEPI), as well as a corresponding 95% confidence interval. Inspecting whether the confidence interval includes 0 allows us to determine whether pairs of associations differ.

We next compared estimates of predicted brain-age to true (chronological) age, across sequences, using Spearman’s correlation coefficient (rs), the proportion of variance explained (r2) as well as the median absolute error (MAE). We also compared the predicted brain-ages between scan types.

As the correspondence between brain age predictions from standard and EPImix T1-w scans was imperfect (likely due to the brain-age model being trained on standard T1-w scans; see Results & Discussion), we explored the use of an additional regression step to optimise brain age prediction from EPImix T1-w scans. We used leave-one-out cross-validation to regress T1-w predicted brain-age on EPImix-predicted brain-age (in 63 participants), leading to an adjusted estimate of predicted brain-age in the remaining (left-out) participant. Repeating this approach by iterating across all participants enabled us to quantify the MAE (relative to chronological age) of this adjusted brain-age estimate.

Finally, to benchmark the relative ability of each scan type (T1-w, EPImix T1-w, T1-w FoVEPI) to predict brain-age, we compared each MAE estimate to the MAE from the worst possible brain-age model. Our “null model” MAE estimate was based on an assumed prediction of the same brain-age for every participant, equivalent to the mean age of participants within the training dataset of the brainageR model (40.6 years); this led to a null MAE of 15.6 years. We then calculated the ratio of this null MAE to the MAE of each scan type, to estimate the extent of predictive improvement of each scan type relative to the worst possible model.

Most analyses were carried out using the sample of 64 participants for whom both EPImix and standard T1-w scans were available; additionally, some analyses focused on EPImix T1-w scans were repeated in the full sample of 94 participants – including the relationship between EPImix-derived GM volume and chronological age, and the correlation between EPImix-predicted age and chronological age (SI Fig. S4).

Test-retest reliability of EPImix T1-w scans

Test-rest reliability of quantitative measures derived from EPImix T1-w scans was assessed using 10 within-session test-retest EPImix scans. Test-retest reliability was evaluated using the intraclass correlation coefficient; specifically, we used the one-way random effects model for the consistency of single measurements, i.e., ICC(3,1), hereafter referred to as ICC29. We quantified the test-retest reliability of global tissue volumes (GM, WM and CSF), corresponding voxel-wise tissue volumes, as well as predicted brain-age.

All statistical analyses were carried out in Python 3.7.

Figure 2
figure 2

Voxel-wise tissue volume estimation across contrasts. Correlation between voxel-wise tissue volume estimates from T1-w and EPImix (T1-w) scans, and corresponding probability density plots, for grey matter (A, B), white matter (C, D) and CSF (E, F). Only voxels with at least 0.001 mm3 tissue volume in at least 95% of participants are shown.

Results

Tissue volume

We first compared global brain volumes of conventional and EPImix-derived T1-w scans. We observed strong positive correlations between standard T1-w and EPImix T1-w scans, in both GM volume (rs = 0.84, p < 0.001; Fig. 1A) and WM volume (rs = 0.84, p < 0.001; Fig. 1B). Measures of CSF volume demonstrated a weaker positive correlation (rs = 0.56, p < 0.001; Fig. 1C). These results were qualitatively consistent when using T1-w scans with reduced FoV (Fig. 1D–F).

We next compared voxelwise estimates of GM, WM and CSF volume between conventional and EPImix-derived T1-w scans, in voxels with at least 0.001 mm3 tissue volume in at least 95% of participants. Predominantly positive correlations across participants were observed for all three tissue types. Similarly to global tissue volumes, correlations were strongest in the GM (median [1st, 3rd Quartile] (Md [Q1, Q3]) = 0.70 [0.52, 0.82]; Fig. 2A,B) and WM (Md [Q1, Q3] = 0.77 [0.62, 0.88]; Fig. 2C,D), followed by CSF (Md [Q1, Q3] = 0.53 [0.36, 0.70]; Fig. 2E,F).

Figure 3
figure 3

Grey matter volume as a function of chronological age, using volume estimates derived from (A) T1-w scans, (B) EPImix T1-w scans and (C) T1-w scans with reduced field of view.

Additionally, we quantified the relationship of global GM volume with age. GM volume decreased as a function of age, for estimates derived from standard T1-w scans (rs = −0.28, p = 0.027; Fig. 3A), EPImix T1-w scans (rs = −0.44, p < 0.001; Fig. 3B) and T1-w scans with reduced FoV (rs = −0.27, p = 0.032; Fig. 3C). Bootstrap sampling of participants demonstrated that the association between GM volume and age for EPImix T1-w scans is marginally stronger than corresponding associations estimated using GM volumes derived from T1-w scans, and T1-w scans with reduced FoV. For details, see Supplementary Information.

Brain age

Following tissue volume analyses, we used a pre-trained Gaussian Processes Regression model27 to quantify the relative ability of EPImix T1-w and standard T1-w scans to predict brain-age. As a first step, we compared the chronological (true) age of participants with their brain-age prediction from each type of scan. Standard T1-w scans showed a high correspondence between chronological and predicted age, including both a high correlation and low error (rs = 0.73, p < 0.001, MAE = 3.72 years; Fig. 4A). EPImix T1-w scans also showed a high correlation between chronological and predicted age (rs = 0.61, p < 0.001) but with a substantially higher error (MAE = 14.24 years), due to a systematic prediction offset (Fig. 4B). Additional analysis of standard T1-w scans with reduced FoV (to match EPImix) showed a similarly reduced correspondence between predicted and chronological age (rs = 0.36, p = 0.004, MAE = 13.05 years; Fig. 4C), suggesting that the poorer predictive ability of EPImix T1-w scans was likely caused by their reduced brain coverage.

Figure 4
figure 4

Age prediction across contrasts. Predicted age as a function of chronological age for (A) T1-w scans, (B) EPImix T1-w scans and (C) T1-w scans with reduced FoV. (D) Predicted age of EPImix T1-w scans compared to standard T1-w scans, and E) EPImix T1-w scans compared to T1-w scans with reduced FoV. (Spearman’s correlation coefficient rs, p-value, r2 derived from Pearson’s correlation).

Table 1 Correlation and Median Absolute Error between chronological age and predicted brain-age.

Furthermore, we directly evaluated the correspondence between brain-age estimates from different sequences. There was a strong positive correlation between predicted brain-age derived from EPImix T1-w and standard T1-w scans (rs = 0.71, p < 0.001), although the same systematic offset (relative to the identity line) described above was apparent (Fig. 4D). The correlation decreased when comparing brain-age estimates from T1-w scans with identically reduced FoV (rs = 0.54, p < 0.001), although data shifted closer to the identity line (Fig. 4E).

Figure 5
figure 5

Test-retest reliability of EPImix (T1-w) tissue volume estimates. Reliability of voxel-wise volume estimates, and corresponding probability density plots, in grey matter (A, B), white matter (C, D) and CSF (E, F). Test-retest reliability was evaluated using the one-way random effects model for the consistency of single measurements (ICC(3,1)29). Only voxels with at least 0.001 mm3 tissue volume in at least 95% of participants are shown.

We next benchmarked the brain-age MAE of each contrast type relative to the worst possible MAE. Assuming an identical brain-age prediction for each participant, equivalent to the mean age of the brainageR training dataset (40.6 years), gives rise to an MAE of 15.6 years. While brain-age prediction using single-contrast T1-w scans is 4.20× better than this null benchmark, the performance of both EPImix T1-w scans and T1-w scans with reduced FoV is minimally improved (respectively 1.10×  and 1.20×).

To improve brain-age prediction from EPImix T1-w scans, we used leave-one-out regression to adjust EPImix T1-w brain-age estimates. We regressed T1-w brain-age on EPImix brain-age (in 63 participants), leading to an adjusted estimate of predicted brain-age in the remaining (left-out) participant. Following iteration across all participants, this led to an adjusted MAE of 3.70 years (across left-out participants).

For all summary statistics related to predicted brain-age, see Table 1.

Test-retest reliability

We used a subsample of 10 participants with two (within-session) EPImix acquisitions each to quantify the test-retest reliability of all EPImix-derived quantitative measures evaluated in this study, using the intraclass correlation coefficient (ICC).

Estimates of global tissue volume showed high reliability, for GM (ICC [95% confidence interval (CI95)] = 0.99 [0.99,1]), WM (ICC [CI95] = 0.99 [0.98, 1] and CSF (ICC [CI95] = 0.92 [0.70,0.98]) (all p < 0.001).

Voxelwise estimates of tissue volume showed equally high test-retest reliability, for GM (Md [Q1, Q3] = 0.95 [0.90,0.97]; Fig. 5A,B), WM (Md [Q1, Q3] = 0.96 [0.90,0.98]; Fig. 5C,D) and CSF (Md [Q1, Q3] = 0.92 [0.87,0.97]; Fig. 5E,F).

Finally, brain-age estimates were equally reliable (ICC [CI95] = 0.99 [0.95,1]).

Discussion

We investigated whether T1-w scans from the recently-developed rapid multicontrast EPImix sequence1 are quantitatively comparable with routinely collected single-contrast T1-w scans. For these comparisons, we relied on interpretable and widely used quantitative measures, including global and local estimates of tissue volume as well as predicted brain-age. We found a strong correspondence between tissue volumes derived from both sequences, at both the global and local levels. Moreover, estimates derived from EPImix scans showed the expected decrease in grey matter volume as a function of age. Additionally, both types of scan significantly predicted participant brain-age, although the reduced FoV of EPImix scans led to a systematic offset and a commensurate increase in prediction error. However, we demonstrated that this can potentially be corrected using additional leave-one-out regression of T1-w-predicted brain-age onto EPImix-predicted brain-age. Finally, we used a subset of participants to show high test-retest reliability of all EPimix-derived quantitative measures evaluated in this study.

Tissue volume

Both global and voxelwise analyses showed that grey matter, white matter and cerebrospinal fluid volumes are comparable between T1-w and EPImix T1-w scans. This extends our previous findings of correspondence between tissue intensities and Jacobian determinants (derived from registration of T1-w scans to MNI standard space), between EPImix and standard T1-w scans7. While Jacobian determinants can be derived relatively rapidly from T1-w scans, the primary constraint being the speed of non-linear registration, they are both less interpretable and less widely used than estimates of tissue volume. For example, recent work employing the increasingly popular approach of normative modeling30,31 leveraged a large sample of scans to construct normative charts of variation in brain tissue volume across the lifespan32,33. Our results suggest that even volume estimates derived from rapid sequences such as EPImix could be used to anchor individual measures of brain morphology relative to such reference datasets. Our results suggest that global GM and WM volumes show higher correspondence than CSF, between EPImix and T1-w scans, for both the full and reduced FoV (Fig. 1). Consistently, the correspondence of CSF volumes is also reduced at the voxel level (Fig. 2E). However, some locations show high correlations between EPImix and standard T1-w scans, including the ventricles, indicating value in future parsing of CSF volumes into components such as ventricular volume. Such measures are potentially more relevant to clinical translation, including demonstration of dementia progression34 and differentiation of dementia variants35. However, it would be preferable to increase the FoV of EPImix scans to cover the whole brain, at the cost of a small increase in acquisition time (e.g., ~1.5 minutes instead of ~1 minute).

In addition, healthy ageing is well known to be associated with grey matter volume reduction17,19. Despite our modest sample size and non-uniform distribution of participants as a function of age, we found this signature of decreasing grey matter volume in EPImix T1-w scans. Bootstrap resampling of participants suggests that the association between GM volume and age in EPImix T1-w scans is marginally stronger than corresponding associations estimated using standard T1-w scans or T1-w scans with reduced FoV. However, potential differences between these associations should be re-evaluated in a larger sample of participants who are uniformly distributed across age.

Tissue volumes derived from EPImix showed high within-session test-retest reliability. Further work should investigate between-session test-retest reliability, and additionally compare test-retest reliability of EPImix to standard high-resolution T1-w scans. To our knowledge, the impact of scan resolution on reliability has previously only been investigated in the context of simple statistics (e.g. mean and variance) and texture features (e.g. autocorrelation and contrast)36. Further work is needed on the impact of scan resolution on reliability of more interpretable and clinically relevant quantitative measures, such as global and local tissue volumes. Translation to the clinical domain should take into account differences in reliability of tissue volume between healthy controls and clinical groups, such as individuals with MCI and AD37.

Predicted brain-age

Previous research has established predicted brain-age as a putative biomarker of brain health10,11. An increased predicted brain-age can be a strong predictor of risk of neurodegenerative diseases, such as Alzheimer’s disease, and neuropsychiatric disorders, such as schizophrenia10,20,23. We used a pre-trained Gaussian Processes Regression model27 to derive estimates of predicted brain-age from EPImix as well as conventional T1-w scans. EPImix-derived brain-age showed a strong correspondence (as quantified using Spearman’s rs, and r2) to chronological age – of a similar magnitude to previous studies (e.g. Cole20) – as well as to brain-age estimated from standard T1-w scans.

However, a systematic offset led to a large median absolute error of the prediction. Re-analysis of T1-w scans with reduced FoV led to a commensurate drop in performance, confirming that this systematic error is likely caused by incomplete brain coverage of EPImix scans, combined with the fact that the brainageR model we used27 was pre-trained on conventional single-contrast T1-w scans (with full FoV). However, we showed that this systematic offset can potentially be corrected using leave-one-out regression of T1-w-predicted brain-age on EPImix-predicted brain-age, to yield an adjusted estimate of EPImix-derived brain-age. Such approaches could be used to adjust brain age predictions derived from newly-acquired EPImix scans with reduced FoV.

In future, multicontrast data produced by the EPImix sequence could also be used to improve the accuracy of brain age predictions and help compensate for the reduced resolution of EPImix scans, as demonstrated by prior work on the added value of multimodal information in the context of brain age prediction20. Beyond improving brain age estimates, multicontrast information provided by EPImix could help enhance its detection of inter-individual differences.

Active acquisition

An exciting potential application for the EPImix sequence is the active acquisition approach proposed by Cole et al.16. Active acquisition involves the online analysis of MR scans, aiming to use active learning14 to analyse scans as they are being acquired, in turn guiding subsequent acquisition steps16. As our study has found comparable quantitative measures between EPImix and routinely collected T1-w scans, whilst scanning times are considerably faster, there is potential for EPImix to be utilised in the online collection and analysis of brain scans, in the process of active acquisition.

Cole et al.16 proposed three examples of active acquisition scenarios, with EPImix potentially being suitable for use in all three. Firstly, due to its relative speed, EPImix could be used to rapidly acquire low resolution data to inform whether higher resolution data needs to be acquired (more slowly), depending on detection of abnormalities in the initial EPImix scan. The other two scenarios propose to leverage multi-modal data, to classify participants and/or to identify modalities in which participants deviate the most from the norm. Due to the multicontrast nature of the EPImix sequence, it could also be used in such scenarios. However, a larger normative sample would be required to robustly model inter-individual variability, and clinical data would be useful to test the translational relevance of such models30,31.

The acquisition time of the EPImix sequence is considerably faster than analogous single-contrast scans1; however, the current analysis pipeline (median = 4 min 49 seconds) is too slow for real-time analysis required for active acquisition16. As it currently stands, a small amount of analyses could be carried out whilst the participant is in the scanner, to inform several additional acquisition steps. Although not sufficient for the near-real-time process of active acquisition, this could still reduce the need for participants to be recalled for follow-up scans.

In the future, there is potential to run the data analysis on more powerful computers, to speed up the processing and analysis pipeline. Alternatively, deep learning tools hold the promise of massively accelerating processing. While such models are very costly to train – in terms of time, computing power and energy – inference is very fast. Examples of relevant recent tools include SynthSeg for scan segmentation38, voxelmorph for scan registration39 or FastSurfer, a deep learning analogue of FreeSurfer which is particularly suitable for surface-based measures40. Additionally, EPImix scans could be super-resolved using image quality transfer tools, to improve scan resolution and likely increase correspondence to conventional single-contrast acquisitions41.

Methodological considerations

There are a number of methodological limitations to be considered, to maximise potential practical utility of the rapid EPImix sequence in the context of quantitative analyses.

One limitation of the present study is the reduced FoV of EPImix scans, particularly in outer areas of cortex, such as the inferior temporal lobe and superior parietal lobe. We investigated the impact of this reduced coverage on our results by re-analysing single-contrast T1-w scans with identically reduced FoV (on an individual participant basis). This enabled us to conclude that the reduced FoV is likely the primary hindrance in the ability of EPImix T1-w scans to predict brain-age. However, this issue can potentially be addressed using an additional leave-one-out regression step to correct predicted brain-age values, as demonstrated here. We relied on leave-one-out cross-validation – despite its inflated performance relative to k-fold or hold-out validation – due to small sample size, and to demonstrate the utility of even a small paired (EPImix and single-contrast T1-w scan) dataset for the correction of brain-age predictions in a new (EPImix-only) scan. An improved approach in future studies would be to increase the FoV of EPImix scans; despite increased scanning time, the multicontrast EPImix sequence would still remain considerably faster than routinely collected (higher-resolution) single-contrast scans. A larger FoV could be particularly important in clinical settings, where accurate diagnosis is dependent on comprehensive brain coverage42.

In addition, the image analysis pipeline used here remains too slow to realistically be used in an active acquisition setting, particularly if image processing needs to occur in near-real-time to drive scanning in a closed loop16. Our previous work developed a custom minimal processing pipeline which is sufficiently fast to realistically operate in near-real-time; however, this is limited to scan registration and Jacobian extraction, and does not enable quantification of tissue volume or brain age prediction. Increasing the speed of processing would be particularly valuable to enable the EPImix sequence to be used as a quantitative screening test, given that its speed is one of its main advantages over standard structural MR imaging1,2. One potential avenue for the creation of custom rapid processing pipelines is Bayesian optimisation, which has previously been used to customise processing tools for brain-age prediction43.

Moreover, additional research is needed to quantitatively compare the other contrasts generated by the multicontrast EPImix sequence, including T2-weighted, T2-FLAIR, T2*-weighted, diffusion-weighted contrasts and the apparent diffusion coefficient, with analogous single-contrast MR scans. Similarly, it would be interesting to compare EPImix to other similar sequences on scanners from other manufacturers44, using quantitative tools applied here as well as the aforementioned deep-learning approaches.

Finally, further research should focus on clinical groups. Beyond the translational value of such studies, they would help to further investigate the correspondence between EPImix and standard T1-w scans. In particular, a key question is whether group differences are similar within EPImix and standard T1-w scans. Regarding future translation of our work to a clinical setting, it should be noted that global brain-age is a general measure, which is not disease specific. While the aim of active acquisition is to ultimately obtain a personalised diagnosis, any information regarding potential abnormality – including global deviation from expected brain-age – could help inform clinical predictions. In future, this should additionally be supplemented by more specific markers of brain health, including local estimates of brain-age45,46,47,48 as well as local deviations of brain anatomy from the norm31,32,33. Accordingly, further work should compare both of these local candidate biomarkers between EPImix and standard T1-w scans.

Conclusion

In summary, we used popular quantitative measures derived from brain MRI to compare T1-w scans derived from the new rapid EPImix sequence with routinely collected single-contrast T1-w scans. We found that both global and voxelwise tissue volume estimates derived from EPImix T1-w scans were comparable with analogous measures extracted from single-contrast high-resolution T1-w scans. Brain age predictions from EPImix T1-w scans showed a strong correlation with both chronological age and brain-age estimates from single-contrast T1-w scans, although a systematic offset led to a high prediction error. However, this could be corrected using additional leave-one-out regression.

Taken together, our findings underline the potential of the EPImix sequence to reduce scanning time, increasing participant comfort and reducing cost, and further highlight its relevance and applicability to quantitative MR analysis routines. Future extension of this work includes the development of additional customised and deep-learning-based processing tools, as well as the analysis of other contrasts generated by EPImix; this will enable researchers to fully harness the potential of this multicontrast sequence to drive innovative paradigms in MRI acquisition and processing.