Abstract
Percutaneous intramuscular electrodes and a portable multichannel system were used to restore the function of the paralyzed lower extremities in six patients with complete paraplegia. The total number of inserted electrodes was 168. All of the patients could stand, two could walk in parallel bars, and two could walk with a walker. The rate of breakage of electrodes was only 0.6% in our series. There were 10 (6.0%) superficial infections, and 10 (6.0%) movement of electrodes which required reimplantation. The results suggest that the ultrafine intramuscular electrode is practical for long term use with paraplegic patients. Although the system can be used for paraplegic patients during the activities of daily living, it will be necessary to develop a closed-loop controller to reduce the amount of stimulation to the extensor muscles and extend the endurance of upright activity to reduce fatigue.
Similar content being viewed by others
Article PDF
References
Andrews B J et al. Hybrid FES orthosis incorporating closed and sensory feedback. J Biomed Eng 1988; 10: 189–195.
Bajd T et al. Assessment of functional characteristics relevant to FES application in complete SCI patients. Proc 9th Intern Symp ECHE pp 129–140, 1987.
Handa Y et al. Studies on ultrafine intramuscular electrode and skin button for FES. Proc 10th Ann Intern Conf IEEE/EMBS pp 1686–1687, 1988.
Handa Y, Ohkubo K, Hoshimiya N . A portable multichannel FES system for restoration of motor function of the paralyzed extremities. Automedica 1989; 11: 221–231.
Handa Y, Hoshimiya N, Iguchi Y, Oda T . Development of percutaneous intramuscular electrode for multichannel FES system. IEEE Trans Biomed Eng 1989; 36: 705–710.
Hoshimiya N, Handa Y . A master-slave type multichannel functional electrical stimulation (FES) system for the control of the paralyzed upper extremities. Automedica 1989; 11: 209–220.
Ichie M . Restoration of locomotion by FES. Proc Rehabil Intern Seminar pp 1–12, 1990.
Jaeger R J, Yarkony G M, Roth E J, Lovell L . Estimating the user population of a simple electrical stimulation system for standing. Paraplegia 1990; 28: 505–511.
Kagaya H . Therapeutic and functional electrical stimulation for paraplegics. J Jpn Ortop Assoc 1994; 68: 751–762.
Kantrowitz A . Electronic physiologic aids. A report of Maimonides Hospital pp 4–5, 1960.
Marsolais E B, Kobetic R . Functional electrical stimulation for walking in paraplegia. J Bone Joint Surg 1987: 69-A: 728–733.
Marsolais E B, Kobetic R, Chizeck H J, Jacobs J L . Orthoses and electrical stimulation for walking in complete paraplegia. J Neuro Rehab 1991; 5: 13–22.
Scheiner A, Poland G, Marsolais B . Design and clinical application of a double helix electrode for functional electrical stimulation. IEEE Trans Biomed Eng 1994; 41: 425–431.
Shimada Y et al. Restoration of locomotion in paraplegics by functional electrical stimulation. Proc 1st Intern FES Symp pp 110–114, 1992.
Solomonow M et al. Gait performance of paraplegics ambulating with the reciprocating gait orthosis powered by electrical muscle stimulation. Proc 11th Intern Conf IEEE/EMBS p 1013, 1989.
Yarkony G M et al. Functional neuromuscular stimulation for standing after spinal cord injury. Arch Phys Med Rehabil 1990; 71: 201–206.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Shimada, Y., Sato, K., Abe , E. et al. Clinical experience of functional electrical stimulation in complete paraplegia. Spinal Cord 34, 615–619 (1996). https://doi.org/10.1038/sc.1996.110
Issue Date:
DOI: https://doi.org/10.1038/sc.1996.110
Keywords
This article is cited by
-
Functional electrical stimulation
Artificial Life and Robotics (2000)