Abstract
Iron deficiency is the most common and widespread nutritional disorder worldwide. The automated haematology analyser XN-30 (Sysmex, Kobe, Japan) was developed to detect malaria-infected red blood cells (RBCs) in human blood samples using flow cytometry. The optical system of the analyser detects autofluorescence (AF)-emitting RBCs containing iron-deficient haem groups and would aid in the diagnosis of anaemia resulting from iron deficiency. Here, an RBC-optical (RBO) channel was devised and implemented on the analyser. In vitro analyses showed that the analyser detected AF-emitting RBCs treated with 5-aminolevulinic acid. Furthermore, the analyser detected AF-emitting RBCs in mice fed a low iron diet and infected with a rodent malaria parasite; it could also be effectively used in humans. This study demonstrates that the analyser can quantitatively and reproducibly detect AF-emitting RBCs and measure other haematological parameters, suggesting its usefulness for the initial evaluation of latent iron deficiency anaemia in conjunction with the diagnosis of malaria.
Similar content being viewed by others
Introduction
Iron deficiency anaemia (IDA) is the most common and widespread nutritional disorder and is estimated to affect approximately two billion people worldwide1. In nutritionally poor regions, the diagnosis of IDA is often complicated with infectious diseases such as malaria, parasitic worms, and HIV/AIDS1. Because iron deficiency is easily treated with iron supplementation2,3, the World Health Organization recommends iron supplementation in conjunction with effective malaria prevention and treatment strategies4,5. However, iron administered to women and children without prophylaxis or access to adequate health care is often accompanied by an increase in malaria incidence, thus suggesting that IDA protects against Plasmodium falciparum malaria infection6,7,8,9,10,11. This contradiction generates a dilemma for policy makers in dealing with public health management in malaria-endemic areas12,13 and makes further clinical studies necessary. Therefore, a rapid and simple assay to detect latent IDA in malaria patients who require additional treatment would be valuable in routine laboratory testing, especially in areas where malaria is endemic and iron deficiency is common.
The automated haematology analyser XN-30 (Sysmex, Kobe, Japan) was developed to quickly detect malaria-causing parasites and to calculate parasitaemia in human blood samples through flow cytometric analysis14. In this system, the nucleic acids of malaria-infected RBCs (iRBCs) are stained with a Fluorocell M solution and excited by a semiconductor 405 nm laser beam. Therefore, we expected the optical system of the XN-30 analyser to detect RBCs containing iron-deficient haem (IDH), which is excited with a 402 nm laser and emits at approximately 631 nm15.
This study aimed to demonstrate the applicability of the XN-30 analyser for the initial evaluation of IDA in routine laboratory testing. We devised and implemented an RBC-optical (RBO) channel on the analyser and validated detection of autofluorescence (AF)-emitting RBCs in vitro and in vivo. Finally, we assessed the ability of the analyser to detect AF-emitting RBCs in healthy human blood samples.
Results
Implementation of the RBO channel on the XN-30 analyser
To demonstrate that the optical system of the XN-30 analyser can detect AF-emitting RBCs, an RBO chamber/channel was implemented on the XN-30 analyser (Fig. 1). In the analyser, aspirated test samples were separated into two reaction chambers, which feed the analyser’s M and RBO channels. In the M chamber, the samples were treated with both Lysercell M and Fluorocell M14; in the RBO chamber, the samples were treated with CELLPACK DFL, which enables the precise determination of RBC size. The treated samples were sequentially measured in the same flow cell in the order of the M and RBO channels. The flow cell was completely washed using CELLPACK DCL after the detection of iRBCs from the M chamber and was used for the detection of AF from RBCs carried from the RBO chamber. Similarly, the flow cell was completely washed using CELLPACK DCL after the detection of AF. In addition, the M and RBO chambers were also completely washed using CELLPACK DCL every measurement. Therefore, the acquired fluorescence was the background level after every measurement (data not shown). The detected signals were automatically displayed on the M and RBO scattergrams as parasitaemias (MI-RBC%) and AF rate (AF%) (Fig. 1, see also Fig. 2a).
Definition of AF-emitting RBCs on the RBO scattergram
To define AF-emitting RBCs on the RBO scattergram, the signals of both total and AF-emitting RBCs were gated. The forward-scattered light (FSC), shown as the vertical axis, refers to the cell size. Therefore, the upper and lower populations represent RBCs and platelets, respectively (Fig. 2a). In addition, the side fluorescent light (SFL), shown on the horizontal axis, indicates the intensity of the AF emitted from RBCs. The basal gating for counting the number of total RBCs was also determined as the "RBC gate" (Fig. 2a, black rectangle). The vertical axis range for the RBC gate was determined to include most of the RBCs. Specific gating for AF-emitting RBCs (“AF gate”) is indicated on the scattergram (Fig. 2a, red rectangle). The lower limit (LT) of the AF gate along the horizontal axis was determined using the population of RBCs lacking AF in 5-aminolevulinic acid (ALA)-untreated samples as a reference. To determine the LT, 10 ALA-untreated RBC samples were measured, and the average and standard deviation (SD) of the RBC population along the horizontal axis were calculated. The average value plus three-fold SD was defined as the LT of the AF gate. On the vertical axis, the range for the AF gate was identical to that for the RBC gate. The AF% was calculated according to these gates. Fluorescence microscopy confirmed that the observed AF originated from RBCs (Fig. 2b). The spectrometer showed that the emission fluorescence of the RBC lysate excited at 405 nm was most intense at 617 nm (Fig. 2c). This wavelength was nearly identical to that reported in a previous study on PPIX (ca. 631 nm)15. In contrast, the generation of AF was disturbed by the addition of succinylacetone (SA), which is an inhibitor of aminolevulinic acid dehydratase in the PPIX synthesis pathway (Fig. 2d). These results suggest that AF was emitted from IDH, and especially from coproporphyrinogen III (CPP) that was generated in the PPIX synthesis pathway (see Discussion). Further, the SD values between measurements were considerably low (SD = 0.080 to 0.93) in each sample (Fig. 2e).
Characterisation of AF obtained from RBCs
To characterise the AF obtained from the RBO channel, RBCs were treated with ALA. The AF% of iRBCs was higher than that of RBCs and increased according to ALA concentration and incubation time (Fig. 3a). In addition, the XN-30 analyser also revealed that a high concentration (over 50 µM) of ALA was toxic for parasites (Supplementary Fig. S1). Fluorescence microscopy showed that AF was enhanced not only in iRBCs but also in RBCs (Fig. 3b). This observation indicated that the AF of non-iRBCs was enhanced by co-cultivation with iRBCs, suggesting that changes in culture conditions associated with the growth of the parasite enhanced AF in non-iRBCs.
To determine the factors affecting the enhancement of AF, we examined culture conditions. The analyser showed that stored RBCs increased AF% (Fig. 3c), whereas stored human sera supplemented with culture medium did not affect AF% (Supplementary Fig. S2). Furthermore, the addition of glucose decreased AF% (Fig. 3d), suggesting that a decrease in glucose concentration enhanced AF. In addition, Ko143, an inhibitor of ABCG2 (an exporter of IDH), increased AF% (Fig. 3e), suggesting that ABCG2 exported IDH from the RBCs.
Evaluation of AF-emitting RBCs and haematological parameters in mice fed a low iron diet (LID)
To confirm the availability of the RBO channel for the detection of IDA, mice were fed an LID and their blood samples were analysed (Fig. 4a). Mouse AF% increased during feeding (Fig. 4b(i), days 7 and 14); however, AF amounts returned to near basal levels when mice were fed a normal diet (ND) (Fig. 4b(i), day 21). The RBC count remained constant throughout the experiment (Fig. 4b(ii)). Other haematological parameters including haemoglobin (HGB), haematocrit (HCT), and mean corpuscular volume (MCV) values were decreased, whereas platelet (PLT) counts were increased and mean platelet volume (MPV) value was slightly increased at days 7 and 14. These changes were reversed by feeding the ND (Fig. 4b(iii) to (viii) excluding (vi)). In contrast, the RBC distribution width-coefficient of variation (RDW-CV) rate (which shows variation in circulating RBC size) was significantly increased after switching to the ND (Fig. 4b(vi), p < 0.001), suggesting the recovery of RBC size (see also Fig. 4b(v)). To confirm that the changes in AF% and haematological parameters were related to haematopoietic responses, we compared mRNA expression levels of hepcidin (Hamp) in the liver, erythropoietin (Epo) in the kidney, and erythroferrone (Erfe) in the spleen. Two weeks after ND- and LID-feeding, the AF% of the LID-fed group was higher than that of the ND-fed group (Fig. 4c(i)). The expression level of Hamp mRNA was significantly decreased, whereas the expression levels of Epo and Erfe mRNA were increased (Fig. 4c(ii) to (iv)). These results suggest that the AF% increase is related to enhanced iron uptake and erythropoiesis16. Taken together, these observations suggest that the analyser detects IDA caused by diet, as well as AF-emitting RBCs. In addition, the RBO scattergram showed that AF was mostly emitted from smaller RBCs on day 14 (Fig. 4d), implying that newly generated RBCs are smaller and contain IDH. In contrast, fluorescence microscopy showed that AF was not necessarily emitted from smaller RBCs (Fig. 4e). These results indicate that RBCs treated with CELLPACK DFL, which enables the precise determination of RBC size, are displayed on the scattergram according to their size.
Evaluation of AF-emitting RBCs and haematological parameters in mice infected with rodent malaria parasite
To investigate the relationship between malaria infection and IDA, blood samples collected from parasite-infected mice were analysed. The AF% fluctuated according to the degree of parasitaemia, with changes in AF% appearing after changes in parasitaemia (Fig. 5a), particularly at days 7, 10, and 28. AF% changes resulted from fluctuations not only in RBC counts but also in AF-emitting RBC counts (Fig. 5a and Supplementary Fig. S3). Fluorescence microscopy showed that AF was emitted from non-iRBCs, as well as iRBCs (Fig. 5b). In addition to evaluating changes in AF% that occurred during parasite infection, we also evaluated AF changes after treatment with artemisinin (Fig. 5c). The AF% rapidly increased and then decreased after treatment, in contrast with only an AF% increase observed in the non-treatment group (Fig. 5d and Supplementary Fig. S4). No incremental change in AF% was observed in non-infected mice treated with artemisinin. These results suggest that latent IDA occurs temporarily in parasite-infected mice due to the stimulation of immune/inflammatory responses elicited by artemisinin-induced parasite death.
Evaluation of AF-emitting RBCs and haematological parameters in healthy humans
Blood samples from 21 healthy human volunteers were analysed using the XN-30 analyser. We found that the XN-30 analyser could detect AF-emitting RBCs (Fig. 6a). A comparison of AF% allowed for the categorisation of nine subjects as AF-positive (with a threshold of 0.05%). The AF% means were 0.016 and 0.15 in the negative (< 0.05%) and positive (> 0.05%) groups, respectively (Fig. 6b, p < 0.01). In comparison with other haematological parameters, the AF-emitting RBC counts of the positive group was significantly higher than that of the negative group (Fig. 6c(i), p < 0.01). In addition, the MCV and mean corpuscular haemoglobin (MCH) values of the positive group were significantly lower than those of the negative group (Fig. 6c(v) and (vi), p < 0.01), whereas the RDW-CV value of the positive group was significantly higher than that of the negative group (Fig. 6c(viii), p < 0.05). Although other parameters did not differ significantly (Fig. 6c(ii) to (iv), (vii), (ix) and (x)), the HGB and HCT values of the positive group were lower than those of the negative group (Fig. 6c(iii) and (iv)). In addition, all four subjects with an HGB value < 12 g/L, which is a criterion of anaemia, were categorised as AF-positive (Fig. 6c(iii)). These data suggest that the subjects in the AF-positive group were anaemic. However, no subjects had an MCV value < 80 fL, which is a criterion of microcytic anaemia with IDA (Fig. 6c(v)). It is not known if IDA patients are included in this tested group, but these observations—using the analysers—indicated a possibility that latent IDA might be present. To corroborate the ability of the XN-30 analyser to suggest the possibility of latent IDA, we evaluated other haematological parameters with the XN-1000 analyser. The XN-1000 analyser revealed that although reticulocyte percentage (RET%) was not significantly different between AF-negative and AF-positive groups (Fig. 6d(i)), the reticulocyte haemoglobin equivalent (RET-He) value of the positive group was significantly lower than that of the negative group (Fig. 6d(ii), p < 0.05), suggesting that reticulocytes or immature RBCs contain low haemoglobin content in the positive group. Similarly, the RBC-He value of the positive group was significantly lower than that of the negative group (Fig. 6d(iii), p < 0.05), whereas the percentage of hypochromic RBCs (%Hypo-He) and the percentage of microcytic RBCs (%Micro-R) in the positive group were higher than those in the negative group (Fig. 6d(iv) and (v)). These data indicate that reticulocytes and RBCs remarkably contain insufficient haemoglobin and that RBCs are small in the AF-positive group. Taken together, our results suggest that the XN-30 analyser implemented with the RBO channel can help detect latent IDA.
Discussion
Here, we implemented the RBO chamber/channel on the XN-30 analyser to detect AF-emitting RBCs (Fig. 1). The RBO channel detected AF-emitting RBCs treated with ALA (Figs. 2 and 3). Whereas cancer cells treated with ALA synthesise PPIX in mitochondria, RBCs lacking mitochondria cannot synthesise PPIX17,18. Previous studies demonstrated that iRBCs predominantly synthesise CPP, whereas PPIX and coproporphyrin I were synthesised to a lesser extent17,19. These findings suggest that AF detected with the RBO channel is predominantly emitted by CPP synthesised from ALA in the cytoplasm of RBCs in the in vitro parasite culture.
RBCs in the parasite culture displayed evident AF (Fig. 3a). Similar enhancement was also observed using stored RBCs (Fig. 3c), in the absence of glucose (Fig. 3d), and upon the addition of an ABCG2 inhibitor Ko143 (Fig. 3e). Although the reason underlying this enhanced accumulation of CPP in RBCs is not fully understood, further studies of this phenomenon would be useful to design applications involving photodynamic diagnosis (PDD) and photodynamic therapy (PDT), as discussed as follows.
Smith and Kain applied ALA to PDT for malaria treatment and found that P. falciparum parasite growth was completely inhibited by the application of 200 µM ALA followed by white light exposure or by a higher dose (2 mM) of ALA alone20. The current study also demonstrated that the addition of 500 µM ALA partially inhibited parasite growth (Supplementary Fig. S1). However, the direct adoption of ALA-PDT and the use of a high-dose ALA therapy for malaria patients are clinically unrealistic strategies. Recently, Sigala and colleagues showed that the accumulation of CPP in iRBCs could be harnessed for antimalarial chemotherapy using luminol-based chemiluminescence and combinational stimulation with low-dose artemisinin to photoactivate PPIX to produce cytotoxic reactive oxygen17. Furthermore, other studies found this pathway to be a promising target for malaria chemotherapy when parasites were treated with ALA supplemented with ferrous sodium citrate in the in vitro P. falciparum culture18 and in the in vivo rodent malaria model21. These studies suggest that RBO technology is useful for follow-up studies involving ALA treatment.
Our feasibility study using the mouse model demonstrated that the analyser was capable of detecting AF-emitting RBCs, as well as haematological parameters and parasitaemia caused by the LID (Fig. 4 and Supplementary Fig. S3) and malaria infection (Fig. 5 and Supplementary Fig. S4). The clinical study demonstrated that the analyser could detect AF-emitting RBCs, as well as haematological parameters (Fig. 6). The recommended initial evaluation of suspected IDA is based on haematological parameters. Especially, RBC-related parameters, including HGB and MCV, are important for the initial evaluation of IDA22,23,24. As the analyser directly detects RBCs containing IDH in addition to these parameters, it is expected that the analyser would not only be useful for the initial evaluation of IDA but would also provide a novel parameter for the diagnosis of IDA. Therefore, blood samples from IDA patients would be required for more direct evidence in further analyses. In addition, two recent studies reported that the XN-30 analyser is available for the study of clinical malaria25,26. The XN-30 system can evaluate iRBCs, as well as haematological parameters, in approximately 1 min without the requirement for technical expertise14. For the detection of iRBCs, the XN-30 analyser evaluates from approximately 10 thousand to 8 million RBCs/µL every measurement depending on samples. In addition to this, the XN-30 analyser implemented with the RBO channel does not need extra time and specialised handling for the detection of AF-emitting RBCs. These facts suggest that the XN-30 analyser implemented with the RBO channel is useful for the initial evaluation of latent IDA in conjunction with the diagnosis of malaria.
A rapid increase followed by a decrease in the AF% was observed after artemisinin treatment in vivo, suggesting that IDA is also caused by artemisinin treatment (Fig. 5d and Supplementary Fig. S4). It is possible that this rapid change in AF% was caused by a biological response, most likely an inflammatory response elicited by killed parasites. Although previous clinical studies reported delayed haemolysis after artemisinin therapy27,28,29,30, more in-depth studies regarding the relationship between IDA and antimalarial medicine would also be of clinical importance.
The insertion of iron into PPIX is a pivotal step in the production of haem during erythropoiesis31,32. If iron is unavailable, divalent zinc is incorporated into PPIX instead of iron, producing zinc protoporphyrin (ZnPP), which is excited by 405 nm beams and emits at approximately 630 nm, similarly to PPIX33. ZnPP persists throughout the life span of RBCs as a biochemical indicator of functional iron deficiency34,35. Because our methodology does not distinguish ZnPP from PPIX, the XN-30 analyser implemented with the RBO channel possibly detected RBCs containing ZnPP but not PPIX in the in vivo studies using mouse and human samples.
In the current study, AF% was defined as the ratio of RBCs having an intensity higher than the LT (Fig. 2a). Although this definition distinguished AF-positive from AF-negative RBCs, the difference in AF intensity was not considered (see Supplementary Fig. S5). Utilisation of the AF intensity data, as well as the AF% value, might provide new applications for RBO technology.
In cancer cells, treatment with high concentrations of ALA results in IDH accumulation36,37. This accumulation leads to the development of PDD and PDT38,39,40,41,42. The principle of AF generation in cancer cells is the same as that in RBCs. Therefore, RBO technology could also have an application in cancer diagnosis and therapy.
In conclusion, the RBO channel-implemented XN-30 analyser quantitatively and reproducibly detected AF-emitting RBCs in vitro. In addition, the analyser detected AF-emitting RBCs in mice fed an LID and infected with rodent malaria parasites; it was also effective for detection in healthy human subjects. Collectively, these results suggest that the XN-30 analyser implemented with the RBO channel would be useful for the initial evaluation of latent IDA in conjunction with the diagnosis of malaria in malaria-endemic and/or non-endemic areas.
Methods
Ethics, consent, and permissions
The use of blood samples from healthy Japanese volunteers was approved by the institutional review committees of the Research Institute for Microbial Diseases (RIMD), Osaka University (Approval Number: 22-3) and Sysmex corporation (Approval Number: 2019-07). Informed consent was obtained from all participants. All experimental protocol and methods in this experiment were performed in accordance with the relevant guidelines and regulations from the Declaration of Helsinki. All animal experiments were conducted in accordance with the guidelines of “Animal experiment rules” established by the Research Institute for Microbial Diseases, Osaka University, and were approved by the Animal Care and Use Committee of the Research Institute for Microbial Diseases, Osaka University (Approval Numbers: Biken-AP-H26-06-0 and Biken-AP-R01-03-0).
Compounds
ALA was purchased from Cosmo Bio (Tokyo, Japan) and prepared as a 100 mM stock solution in saline (0.9 g/L NaCl). SA was obtained from Cayman Chemicals (Ann Arbor, MI, USA) and prepared as a 10 mM stock solution in saline. Ko143 was purchased from Sigma-Aldrich (St. Louis, MO, USA) and prepared as a 10 mM stock solution in dimethyl sulfoxide (DMSO).
The automated haematology analyser, XN-30
An XN-30 analyser (Sysmex) equipped with a prototype algorithm for cultured P. falciparum parasites (software version: 01-03, (build 16)) was used for the detection of malaria parasites. In brief, the XN-30 analyser aspirated and diluted blood samples in a CELLPACK DCL solution at a specific dilution ratio (1:50). Subsequently, nucleic acids were stained with a staining solution (Fluorocell M) along with a lysis solution (Lysercell M), and iRBCs and WBCs were excited by a 405 nm laser beam. Approximately 100 µL of the culture suspension diluted with 100 µL of phosphate-buffered saline was added to an EDTA-K2 microtube (BD Microtainer MAP microtube for automated process; Becton Dickinson and Co., Franklin Lakes, NJ, USA) and loaded in the XN-30 analyser with an auto-sampler, as described in the instrument manual (Sysmex). The XN-30 analyser generated an M scattergram, which showed the developmental stages of the parasites according to DNA content and iRBC size14. The parasitaemias (total, MI-RBC%; ring-form, RNG-RBC%; trophozoite, TRPZ-RBC%; and schizont, SCHZ-RBC%) were reported automatically.
Implementation of the RBO channel on the XN-30 analyser
The RBO channel was implemented on the XN-30 analyser. The test samples were injected into the RBO chamber and then treated with CELLPACK DFL. The treated samples were measured in the flow cell in order of the M and RBO channels. For the RBO channel, approximately 60 thousand RBCs were evaluated in this study of the human peripheral blood sample. The analyser simultaneously provided an RBO scattergram, in which the dots were presented according to AF intensity and cell size (see Fig. 1). Horizontal and vertical axes represented the intensities of RBO-side fluorescent light (SFL, indicating the intensity of the AF) and forward scattered light (FSC, indicating RBC size), respectively. “RBC gate” and “AF gate” were configured according to the respective intensities (see Fig. 2a). The AF% was calculated according to Eq. (1):
Treatment of RBCs with ALA, SA, and/or Ko143 in vitro
For in vitro analyses, RBCs and ring-form-synchronized iRBCs were prepared in parasite culture medium or in PBS with the indicated concentration of ALA, SA, and/or Ko143 and incubated under defined culture conditions (see the section: Parasite strain and culture in vitro). In in vivo analyses, the addition of ALA is not required because IDH is generated in response to diet and malaria infection.
Parasite strain and culture in vitro
P. falciparum laboratory strain 3D7 was obtained from Prof. Masatsugu Kimura (Osaka City University, Osaka, Japan). For the assessment of antimalarial activity of the compounds in vitro, the parasites were cultured in RPMI 1,640 medium (containing 2.0 g/L glucose) supplemented with 0.5 g/L L-glutamine, 5.96 g/L HEPES, 2 g/L sodium bicarbonate (NaHCO3), 50 mg/L hypoxanthine, 10 mg/L gentamicin, 10% heat-inactivated human serum, and RBCs at a 3% haematocrit level in an atmosphere of 5% CO2, 5% O2, and 90% N2 at 37 °C, as previously described43. RBCs infected with ring-form parasites were collected using the sorbitol synchronisation technique44. Briefly, the RBCs were collected by centrifugation at 840 g for 5 min at room temperature, suspended in a fivefold volume of 5% D-sorbitol (Nacalai Tesque) for 10 min at room temperature, and then washed twice with RPMI 1,640 medium to remove the D-sorbitol.
Diet-induced IDA
Experimental IDA was induced by a LID16. Briefly, female ICR mice (21 days-old) were purchased from Japan SLC (Shizuoka, Japan). For induction of IDA, 7 mice were fed a LID containing 3.6 ppm Fe. As a control, CLEA Rodent Diet CE-2, which contains 310.2 ppm Fe, was used as the ND. Both diets were obtained from CLEA Japan (Tokyo, Japan). The animals were fed a LID for 14 days, after which a ND was given for 7 days.
Quantitation of messenger RNA (mRNA) levels
Total RNA was harvested from mouse liver, kidney, and spleen with the RNeasy Mini Kit (Qiagen, Hilden, Germany). Complementary DNA was synthesized using the High Capacity cDNA Reverse Transcription Kit (Applied Biosystems, Foster City, CA, USA). Quantitative polymerase chain reaction (qPCR) was performed with Luna Universal qPCR Master Mix (NEB, Hitchin, UK) and run in triplicate on the Illumina qPCR Eco system (Illumina, San Diego, CA, USA). Expression levels of Hamp, Epo, and Erfe mRNA were normalized to the reference 60S ribosomal protein L4 (Rpl4), hypoxanthine phosphoribosyl transferase (Hprt), and Rpl4 mRNA, respectively45. Data were expressed as relative quantification, i.e. comparison between ND and LID mice. Primer sequences are indicated in Table 1.
Parasite, infection, and drug treatment in a mouse model
Female C57BL/6 mice (6 weeks-old) purchased from Japan SLC were injected intraperitoneally with 3 × 105 iRBCs with the non-lethal strain P. yoelii 17XNL. Three mice were analysed for analysing the parasite infection (Fig. 5a). For the analysis of parasite infection and treatment (Fig. 5d), 6 parasite-infected mice (3 mice per group) and 3 non-infected mice were analysed; 3 infected mice and 3 non-infected mice were treated with artemisinin. A stock solution of 50 mg/mL of artemisinin (TCI, Tokyo, Japan) was prepared (65% DMSO and 35% Tween-80) and then further diluted to 5 mg/mL (1:10 dilution) with saline. The mice were subcutaneously administrated with artemisinin (25 mg/kg body weight) 4 days after infection with the P. yoelii 17XNL parasite46. Blood samples were collected before and after drug treatment and analysed with the XN-30 analyser. Briefly, row data (saved as FCS file) were exported from the XN-30 analyser and analysed by the Flowing Software 2.5.1 (Turku Centre for Biotechnology, University of Turku, Turku, Finland) using three-dimensional analysis40. RBC count, HGB value, PLT count, and WBC count were calculated according to the indicated dilution ratio (1:50), while HCT, MCV, and MPV values were used directly40.
Fluorescence microscopy
Nucleic acids were stained with 20 nM coriphosphine O (CPO) (TCI, Tokyo, Japan) for 5 min at room temperature and then washed with saline once by centrifugation at 840 g for 5 min at room temperature. Images were captured using a BZ-X710 fluorescence microscope (Keyence, Osaka, Japan). A specific filter (5ALA-405UF1-BLA; excitation = 405/20 nm, emission = 640/30 nm and dichroic mirror = 425 nm) and a BZ-X GFP filter (OP-87763; excitation = 470/40 nm, emission = 525/50 nm and dichroic mirror = 495 nm) were used for detection of AF and CPO, respectively.
Measurement of the fluorescence spectrum
RBCs were treated with 500 µM ALA (dissolved in DMSO) or with DMSO in RPMI 1,640 medium under the parasite culture condition for three days. Treated RBCs were collected by centrifugation at 200 g for 5 min at 4 °C. The collected RBCs were disrupted in water and centrifuged at 200 g for 5 min at 4 °C. The supernatants were measured with an F-4500 fluorescence spectrophotometer (Hitachi, Tokyo, Japan). The samples were excited at 405 nm (bandpass width, 5.0 nm) and their emission spectra were recorded between 200 and 800 nm.
Measurement of human blood samples
Twenty-one healthy Japanese female volunteers participated in the RBO analysis. Peripheral blood samples were collected by EDTA-K2 tube (Terumo, Tokyo, Japan) and were analysed in the XN-30 and XN-1000 analysers within 4 h after blood collection without any dilution. The XN-1000 analyser provided data of RET%, RET-He, RBC-He, %Hypo-He, and %Micro-R values for evaluation of anaemia47,48.
Statistical analyses
The statistical significance of differences between groups of the LID-fed mice was evaluated through a one-way analysis of variance followed by Tukey’s multiple comparison tests using GraphPad Prism version 5.0 (GraphPad Software, San Diego, CA, USA). Statistical significance of differences between two groups in the other experiments was calculated by an unpaired two-tailed Student's t test using GraphPad Prism version 5.0.
Data availability
The datasets used and/or analysed during the current study are available from the corresponding author upon reasonable request.
References
World Health Organisation. Micronutrient deficiencies. https://web.archive.org/web/20170713171156/http://www.who.int/nutrition/topics/ida/en
Pasricha, S. R., Hayes, E., Kalumba, K. & Biggs, B. A. Effect of daily iron supplementation on health in children aged 4–23 months: a systematic review and meta-analysis of randomised controlled trials. Lancet Glob. Health 1, e77–e86 (2013).
Peña-Rosas, J. P., De-Regil, L. M., Garcia-Casal, M. N. & Dowswell, T. Daily oral iron supplementation during pregnancy. Cochrane Database Syst. Rev. 22, CD004736. https://doi.org/10.1002/14651858.CD004736.pub5 (2015).
World Health Organization. Guideline: daily iron supplementation in infants and children (WHO, Geneva, Switzerland, 2016).
World Health Organization. World malaria report 2019. https://www.who.int/publications-detail/world-malaria-report-2019
Nyakeriga, A. M. et al. Iron deficiency and malaria among children living on the coast of Kenya. J. Infect. Dis. 190, 439–447 (2004).
Sazawal, S. et al. Effects of routine prophylactic supplementation with iron and folic acid on admission to hospital and mortality in preschool children in a high malaria transmission setting: community-based, randomised, placebo-controlled trial. Lancet 367, 133–143 (2006).
Kabyemela, E. R., Fried, M., Kurtis, J. D., Mutabingwa, T. K. & Duffy, P. E. Decreased susceptibility to Plasmodium falciparum infection in pregnant women with iron deficiency. J. Infect. Dis. 198, 163–166 (2008).
Gwamaka, M. et al. Iron deficiency protects against severe Plasmodium falciparum malaria and death in young children. Clin. Infect. Dis. 54, 1137–1144 (2012).
Senga, E. L., Harper, G., Koshy, G., Kazembe, P. N. & Brabin, B. J. Reduced risk for placental malaria in iron deficient women. Malar. J. 10, 47 (2011).
Jonker, F. A. et al. Iron status predicts malaria risk in Malawian preschool children. PLoS ONE 7, e42670 (2012).
Prentice, A. M., Verhoef, H. & Cerami, C. Iron fortification and malaria risk in children. JAMA 310, 914–915 (2013).
Neuberger, A., Okebe, J., Yahav, D. & Paul, M. Oral iron supplements for children in malaria-endemic areas. Cochrane Database Syst. Rev. 2, CD006589. https://doi.org/10.1002/14651858.CD006589.pub4 (2016).
Tougan, T. et al. An automated haematology analyzer XN-30 distinguishes developmental stages of falciparum malaria parasite cultured in vitro. Malar. J. 17, 59 (2018).
de Oliveira Silva, F. R. et al. Intrinsic fluorescence of protoporphyrin IX from blood samples can yield information on the growth of prostate tumours. J. Fluoresc. 20, 1159–1165 (2010).
Kobayashi, M. et al. Iron-heme-Bach1 axis is involved in erythroblast adaptation to iron deficiency. Haematologica 102, 454–465 (2017).
Sigala, P. A., Crowley, J. R., Henderson, J. P. & Goldberg, D. E. Deconvoluting heme biosynthesis to target blood-stage malaria parasites. Elife 4, e09143. https://doi.org/10.7554/eLife.09143 (2015).
Nagaraj, V. A. et al. Malaria parasite-synthesized heme is essential in the mosquito and liver stages and complements host heme in the blood stages of infection. PLoS Pathog. 9, e1003522 (2013).
Komatsuya, K. et al. Synergy of ferrous ion on 5-aminolevulinic acid-mediated growth inhibition of Plasmodium falciparum. J. Biochem. 154, 501–504 (2013).
Smith, T. G. & Kain, K. C. Inactivation of Plasmodium falciparum by photodynamic excitation of heme-cycle intermediates derived from delta-aminolevulinic acid. J. Infect. Dis. 190, 184–191 (2004).
Suzuki, S. et al. In vivo curative and protective potential of orally administered 5-aminolevulinic acid plus ferrous ion against malaria. Antimicrob. Agents Chemother. 59, 6960–6967 (2015).
Short, M. W. & Domagalski, J. E. Iron deficiency anemia: evaluation and management. Am. Fam. Phys. 87, 98–104 (2013).
Siu A. L. (2015) US Preventive Services Task Force Screening for Iron Deficiency Anemia in Young Children: USPSTF Recommendation Statement. Pediatrics. 136, 746‒752 (2015)
Siu A. L. U.S. Preventive Services Task Force. Screening for Iron Deficiency Anemia and Iron Supplementation in Pregnant Women to Improve Maternal Health and Birth Outcomes: U.S. Preventive Services Task Force Recommendation Statement. Ann. Intern. Med. 163, 529‒536 (2015).
Pillay, E., Khodaiji, S., Bezuidenhout, B. C., Litshie, M. & Coetzer, T. L. Evaluation of automated malaria diagnosis using the Sysmex XN-30 analyser in a clinical setting. Malar. J. 18, 15 (2019).
Post, A. et al. The XN-30 hematology analyzer for rapid sensitive detection of malaria: a diagnostic accuracy study. BMC Med. 17, 103 (2019).
Centers for Disease Control and Prevention (CDC). Published reports of delayed hemolytic anemia after treatment with artesunate for severe malaria--worldwide, 2010–2012. MMWR Morb. Mortal Wkly. Rep. 62, 5–8 (2013).
Rehman, K., Lötsch, F., Kremsner, P. G. & Ramharter, M. Haemolysis associated with the treatment of malaria with artemisinin derivatives: a systematic review of current evidence. Int. J. Infect. Dis. 29, 268–273 (2014).
Rolling, T., Agbenyega, T., Krishna, S., Kremsner, P. G. & Cramer, J. P. Delayed haemolysis after artesunate treatment of severe malaria - review of the literature and perspective. Travel Med. Infect. Dis. 13, 143–149 (2015).
Kurth, F. et al. Hemolysis after oral artemisinin combination therapy for uncomplicated Plasmodium falciparum malaria. Emerg. Infect. Dis. 22, 1381–1386 (2016).
Tsiftsoglou, A. S., Vizirianakis, I. S. & Strouboulis, J. Erythropoiesis: model systems, molecular regulators, and developmental programs. IUBMB Life 61, 800–830 (2009).
Chiabrando, D., Mercurio, S. & Tolosano, E. Heme and erythropoieis: more than a structural role. Haematologica 99, 973–983 (2014).
Dailey, H. A. & Meissner, P. N. Erythroid heme biosynthesis and its disorders. Cold Spring Harb. Perspect. Med. 3, a011676 (2013).
Enko, D. et al. Assessment of human iron status: a cross-sectional study comparing the clinical utility of different laboratory biomarkers and definitions of iron deficiency in daily practice. Clin. Biochem. 48, 891–896 (2015).
Hennig, G. et al. Non-invasive detection of iron deficiency by fluorescence measurement of erythrocyte zinc protoporphyrin in the lip. Nat. Commun. 7, 10776 (2016).
Schoenfeld, N. et al. The heme biosynthetic pathway in lymphocytes of patients with malignant lymphoproliferative disorders. Cancer Lett. 43, 43–48 (1988).
Kondo, M., Hirota, N., Takaoka, T. & Kajiwara, M. Heme-biosynthetic enzyme activities and porphyrin accumulation in normal liver and hepatoma cell lines of rat. Cell Biol. Toxicol. 9, 95–105 (1993).
Kennedy, J. C., Pottier, R. H. & Pross, D. C. Photodynamic therapy with endogenous protoporphyrin IX: basic principles and present clinical experience. J. Photochem. Photobiol. B. 6, 143–148 (1990).
Kennedy, J. C., Marcus, S. L. & Pottier, R. H. Photodynamic therapy (PDT) and photodiagnosis (PD) using endogenous photosensitization induced by 5-aminolevulinic acid (ALA): mechanisms and clinical results. J. Clin. Laser Med. Surg. 14, 289–304 (1996).
Kelty, C. J., Brown, N. J., Reed, M. W. & Ackroyd, R. The use of 5-aminolaevulinic acid as a photosensitiser in photodynamic therapy and photodiagnosis. Photochem. Photobiol. Sci. 1, 158–168 (2002).
Zhang, J. et al. An updated overview on the development of new photosensitizers for anticancer photodynamic therapy. Acta Pharm. Sin. B. 8, 137–146 (2018).
Mahmoudi, K. et al. 5-aminolevulinic acid photodynamic therapy for the treatment of high-grade gliomas. J. Neurooncol. 141, 595–607 (2019).
Trager, W. & Jensen, J. B. Human malaria parasites in continuous culture. Science 193, 673–675 (1976).
Lambros, C. & Vanderberg, J. P. Synchronization of Plasmodium falciparum erythrocytic stages in culture. J. Parasitol. 65, 418–420 (1979).
Latour, C. et al. Erythroferrone contributes to hepcidin repression in a mouse model of malarial anemia. Haematologica 102, 60–68 (2017).
Tougan, T. et al. Application of the automated haematology analyzer XN-30 in an experimental rodent model of malaria. Malar. J. 17, 165 (2018).
Urrechaga, E., Borque, L. & Escanero, J. F. Clinical value of hypochromia markers in the detection of latent iron deficiency in nonanemic premenopausal women. J. Clin. Lab. Anal. 30, 623–627 (2016).
Levy, S. & Schapkaitz, E. The clinical utility of new reticulocyte and erythrocyte parameters on the Sysmex XN 9000 for iron deficiency in pregnant patients. Int. J. Lab. Hematol. 40, 683–690 (2018).
Acknowledgements
We thank Ms. Toshie Ishisaka for her technical assistance. A patent application has been filed relating to the RBO channel technology (Japan Patent JP2016-50931A).
Funding
This research was partially supported by the Ministry of Education, Culture, Sports, Science, and Technology, Grants-in-Aid for Scientific Research (C) (25460754 and 16K08759) awarded to T.T.
Author information
Authors and Affiliations
Contributions
T.T., K.U., and T.H. conceived and designed the study. Y.T. set up the XN-30 analyser. S.I. performed the in vitro ALA assay. Y.T. and K.U. carried out the clinical study. T.T. conducted all experiments and analysed the resulting data. T.T. drafted the manuscript with input from the other authors. All authors reviewed and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
Y.T. and K.U. are employees of Sysmex Corporation. K.U. holds a patent on Lysercell M. T.T., S.I., and T.H. declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Tougan, T., Itagaki, S., Toya, Y. et al. Implementation of a red blood cell-optical (RBO) channel for detection of latent iron deficiency anaemia by automated measurement of autofluorescence-emitting red blood cells. Sci Rep 10, 15605 (2020). https://doi.org/10.1038/s41598-020-72382-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-72382-z
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.