Abstract
Optofluidic devices combining optics and microfluidics have recently attracted attention for biomolecular analysis due to their high detection sensitivity. Here, we show a silicon chip with tubular microchannels buried inside the substrate featuring temperature gradient (∇T) along the microchannel. We set up an optical fluorescence system consisting of a power-modulated laser light source of 470 nm coupled to the microchannel serving as a light guide via optical fiber. Fluorescence was detected on the other side of the microchannel using a photomultiplier tube connected to an optical fiber via a fluorescein isothiocyanate filter. The PMT output was connected to a lock-in amplifier for signal processing. We performed a melting curve analysis of a short dsDNA – SYBR Green I complex with a known melting temperature (TM) in a flow-through configuration without gradient to verify the functionality of the proposed detection system. We then used the segmented flow configuration and measured the fluorescence amplitude of a droplet exposed to ∇T of ≈ 2.31 °C mm−1, determining the heat transfer time as ≈ 554 ms. The proposed platform can be used as a fast and cost-effective system for performing either MCA of dsDNAs or for measuring protein unfolding for drug-screening applications.
Similar content being viewed by others
Introduction
Temperature can significantly affect the biological systems of living organisms in terms of cellular morphology, metabolism, growth, and cell death1,2. On a molecular level, temperature influences the structure and function of biomolecules such as proteins and nucleic acid3,4. The denaturation of biomolecules under relatively high temperatures is utilized in real-time polymerase reactions (qPCR), subsequent melting curve analyses (MCA) of double-stranded deoxyribonucleic acid (dsDNA), and differential scanning fluorimetry (DSF) of proteins.
The MCA analysis of dsDNA stained with an intercalator dye such as SYBR Green I is conducted by gradually ramping the temperature while monitoring the intensity of the emitted fluorescence (F). The dsDNA melting temperature (TM) is defined as the temperature at which 50% of presented dsDNA molecules separate into single-stranded forms. This value depends on the number of base pairs and their composition. Typically, while the MCA is performed using the qPCR systems, they can be also used to determine the protein-unfolding curve. However, qPCR systems currently available are bulky, costly, and slow.
Advances in microfabrication technology have enabled the miniaturization of bioanalytical tools with precisely controlled temperatures, allowing for quick investigations into the thermal effects on biological molecules using samples with a small volume. Heating methods and the thermal conductivity of the device’s materials are essential to create thermally stable systems5 with either homogeneous distribution of temperature or temperature gradient (∇T).
Heat conduction-based microfluidic techniques, especially in a droplet configuration, have attracted attention due to their applicability in DNA amplification, protein analysis, single-cell assays, and chemical synthesis. The characteristics of the thermal behavior of water in oil droplets within microfluidic channels were studied per the generation and transport of liquid droplets within local heating at the breakup location of T-junctions and flow-focusing systems6,7 either at the downstream channel8,9 or with the heater placed at the end of the transport channel10. However, the temperature dependency of droplet physical properties represented by density, viscosity, and interfacial tension between the continuous oil phase and water-based droplet may complicate high-performance droplet microfluidics. Further, the mathematical simulation of droplet behavior under heating is complicated due to the necessity of 3D models of a droplet-based system11. In addition to the investigation of the temperature dependence of a droplet’s physical properties with localized heating on the microfluidic chip, the creation of on-chip ∇T has been challenging as one can study microtubule polymerization12 along the ∇T or MCA for single nucleotide polymorphisms13, both performed in a flow-through configuration. Moreover, the MCA in microfluidics has been conducted using either solid phase multiple analysis, which requires DNA immobilization13,14, or the immobilization of a free liquid phase restricted for a single analysis15. However, micro-/nanoscale dimensions of channels and the surrounding environment and flow-through configuration may complicate the determination of the temperature distribution in thermally driven microdevices16. Thus, heat transfer in such a system should be characterized in heat fluxes by thermal convection, thermal radiation, or the heat transfer time between the sidewall of a chip and liquid17,18.
Optical systems have been used as a detector for analyzing bio/chemical samples as well as their miniaturization utilized in lab-on-a-chip (LOC) applications19 such as portable qPCR for DNA20 or RNA diagnostics21,22.
Recently, the combination of optics and microfluidics has attracted great attention due to the combination of highly sensitive bio-detection with LOC technology23,24. By fully integrating optical functions on a chip instead of using bulky external optics, these optofluidic systems lower the cost and downsize the system, making it promising for point-of-care diagnosis25. The detection of fluorescence remains valuable in many bio-applications. Therefore, the development and characterization of a multifunctional optofluidic lab-on-a-chip was introduced for sample analysis by fluorescence and Raman spectroscopy26 or by the measurement of the absorbance and fluorescence of droplets in segmented flow27. Further, the fluorescence detection and quantification of the Ebola virus using hybrid optofluidic integration have been reported28, as have platforms for real-time visualization of viruses in complex media29 or cell phone-based imaging cytometers30.
Here, we demonstrate the MCA of dsDNA with the sensitive detection of fluorescence inside a microchannel as part of an optical microsystem exposed to ∇T value. We used a modulated laser as the light source and a photomultiplier tube (PMT) as the fluorescence detector with its output signal processed by a lock-in amplifier to suppress environmental noise. The proposed concept of a thermally regulated optofluidic platform was first characterized in flow-through configuration of dsDNA bulk. More importantly, we presented the MCA analysis in segmented flow configuration as a tool for rapid determination of a biomolecules’ stability.
Materials and Methods
Chip fabrication
We designed the chip layout with a size of (6 × 27) mm2 using the Nanolithography toolbox software31 with the aim of having all fluid/optics inputs/outputs at the chip sidewalls to provide a robust solution. The layout consisted of ≈30 µm-wide lines subsequently forming buried microchannels by a process similar to earlier ones that used two parylene-C depositions32. The first parylene-C was employed as a mask with conformal coating on the sidewalls and a second parylene-C layer to seal the created channels33. We designed our device to insert both capillaries and the optical fibers from the sides (Fig. 1A), allowing a more robust configuration than the previous one, which had its capillaries inserted vertically.
We started the fabrication process using Si wafers with a diameter of ≈ 100 mm and an unusual thickness: ≈ 1 mm, which is thicker than the ordinary wafer thickness of ≈ 450 µm to 550 µm since the targeted microchannel diameter was the same: between ≈ 450 µm and ≈ 550 µm.
First, we coated the wafers with a ≈ 10 µm-thick positive photoresist (PR), followed by a prebake at ≈ 110 °C for ≈ 165 s. The wafers were exposed to ultraviolet light with a dose of ≈ 1600 mJ·cm−2 for lithography and developed in a KOH-based developer for ≈ 300 s, patterning the shape of the trenches at a designed width of 30 µm. The Si was etched with the Bosch process34 to a target depth of ≈ 500 µm, then, we stripped PR in N-methyl-2-pyrrolidinone solution at ≈ 80 °C for ≈ 600 s (Fig. 1B).
The wafer was then coated with a ≈ 1.5 µm parylene-C layer (Fig. 1C), and this layer was photo-blasted at the trench bottoms using a femtosecond laser working at a principal wavelength of ≈ 515 nm using pulses with a duration of ≈ 300 fs and a maximum amplitude of pulse energy of ≈ 200 μJ (Fig. 1D). The wafer was exposed to XeF2 vapor to isotropically etch Si through the opening in the trench bottom, forming a buried cylindrical microchannel with a diameter of ≈ 500 µm (Fig. 1E). The parylene-C was removed using O2 plasma with a set power of 300 W for a duration of 1 h (Fig. 1F). The wafer was diced into individual chips, and the second parylene-C layer was deposited with a thickness of ≈ 30 µm to cover the microchannel with hydrophobic materials and to seal the ≈ 30 µm-wide trench (Fig. 1G). Both optical fibers and microcapillaries were inserted into the chip and sealed with epoxy resin. Details of one port are shown in Fig. 1H.
Optofluidic
The Si chip had a buried microchannel system consisting of a through channel and three inlets connected via a cross junction and a single outlet. The through channel also served as a light guide with an optofluidic path with a length of ≈ 24 mm (Fig. 2B). The diameter of the buried microchannel was chosen to be ≈ 500 µm (Fig. 2A), sufficiently large for both the optical fiber and the capillary to be inserted inside the chip to form a stable, robust, and autonomous system. The chip was placed on two brass blocks with a distance of ≈ 13 mm (Fig. 2C). Each block had its own heater and sensor connected to a proportional integrative derivative temperature controller. The heaters’ temperature was set to ≈ 50 °C and ≈ 80 °C, respectively, forming a ∇T value of ≈ 2.31 °C·mm−1. We inserted two multimode optical fibers with a core and cladding diameter of (200 ± 4) µm and (220 ± 2) µm, respectively, both (mean ± standard deviation) into the through microchannel of the chip. Each of the optical fibers was placed on an opposite side of the chip and sealed with epoxy resin. One fiber was connected to laser-producing light with a principal wavelength of 471 nm and a nominal power of 1 W. Its power was electrically modulated sinusoidal AC voltage from the lock-in amplifier’s internal power supply with a frequency and amplitude set to 1.2345 kHz and 1.958 V, respectively. The laser power entering the optofluidic chip was attenuated from its original value of 1 W by a set of neutral density filters to ≈ 25 mW. The light leaving the chip was coupled into a second optical fiber connected via a bandpass filter with a center wavelength and bandwidth of ≈ 525 nm and ≈ 50 nm, respectively, to block the second harmonic laser light with a wavelength of ≈ 471 nm and harmonic of ≈ 942 nm. We used a PMT as a light detector, having its gain set to ≈ 5 × 104 by setting the control voltage to 0.67 V. The PMT output was connected to the lock-in amplifier input with sensitivity and time constant set to 50 mV per full range and 300 ms, respectively.
The high power of the laser interacting with the fluid inside the microchannel could affect the measurement as the power corresponds to the power density of ≈ 127 mW·mm−2. Nevertheless, the photothermal effect should not be very severe. First, the oil/water interacts with light at a rather low nominal wavelength of 471 nm, but there is, of course, an interaction with fluorescence-producing dyes, such as SYBR Green I in presence of dsDNA, which warms the water-based sample. The photothermal effect was determined earlier using 500 times more powerful laser per unit of area, which only warmed the water by ≈ 6 K35. Finally, we performed a relative measurement, not absolute, and thus heating of the sample by photothermal effect is compensated.
The fluid was connected to the chip via silica capillaries with internal and external diameters of ≈ 100 µm and 360 µm, respectively, to inputs and output, which were sealed with epoxy resin (Fig. 2D). We used a pressure-controlled system to control the fluid flow rate (ν). The chip was even designed to be able to generate segmented flow using a cross-junction; we formed it externally using a double T-junction as before36, since the externally generated segmented flow was more stable than the one formed internally.
Chemicals
Synthetically prepared dsDNA has a length of 17 bps with a sequence of 5’-TCT GCT GTC ACA ACT AA-3’. The TM value of the dsDNA is ≈ 60.6 °C as verified by a commercial qPCR system. We prepared the test solution by mixing this dsDNA at a concentration of ≈ 20 nM with SYBR-Green I diluted 10,000 times in a Tris-EDTA buffer (1×). We also used this buffer with no dsDNA as a reference. An oil phase of hexadecane supplemented with 2% of SPAN-80 surfactant was used as an immiscible continual phase for segmented flow analysis.
Results and Discussion
Mathematical calculation of heat distribution
Devices operating at elevated air temperatures with the sample flowing through are subject to heat losses by convection, radiation, and heat flux due to sample flow, possibly affecting desired temperature distribution as well as its ∇T37,38. We will only briefly describe the physical analysis, its numerical modeling, and experimental verification by infrared (IR) imaging. Details of those three methods were recently described39.
The system comprised a silicon chip with a thicknesses (t) and widths (w) supported by a pair of heaters (Fig. 2B), resulting in a heat flux P1 between the heaters through the microfluidic chip as:
where λSi is the thermal conductance of Si, (L) is the chip length, and ΔT is the T difference between the heaters. The resulting value of P1 was calculated to be ≈ 2.42 W.
The convection (P2) in air was:
where D is the distance between the heaters, h is the convection coefficient and T1 ambient temperature, giving an amplitude of P2 as ≈ 259 mW, which is 10.8% resulting in minor non-uniformity of the ∇T along the microchannel.
The power dissipated (P3) due to a water-based sample (sample) ν influence can be calculated by:
where ρ is the specific mass of the sample and c is the sample heat capacitance.
Finally, there is also a radiation power (P4) emitted influencing the ∇T value. The P4 amplitude is defined by the Stefan-Boltzmann law:
where ε is surface emissivity, and σ is the Stefan-Boltzmann constant. We calculated the values of P1, P2, P3, and P4 as before39 and determined the total values of (P2 + P3 + P4) are only 9% of the P1; thus, their influence on the ∇T can be neglected.
We neglected different temperatures along the chip to make the analytical estimation simpler. It should be done more complexly, using the integral of the function along the gradient. Nevertheless, the finite element modelling (FEM) does take the local temperature into account.
FEM and IR imaging
We modeled the microfluidic chip as well as simplified heaters in CAD software and transferred them to the finite element modeling software COMSOL Multiphysics. Then we modeled temperature distribution along the microchannel due to convection, radiation, and sample ν using Heat Transfer in Solids and Fluids and Creeping Flow modules in a fashion similar to before39. Here, we only show the model mesh (Fig. 3A) and calculated temperature distribution within the chip and the heaters (Fig. 3B). The simplified detail of the chip edge showing the buried microchannel is shown in Fig. 3C, where the red line shows the location of temperature data extraction for subsequent analysis. We performed FEM of the system with no external influence, with convection, radiation, and both to extract the temperature values inside the channel (Fig. 3D). We also showed the effect of convection and radiation in the inset. Finally, we checked the influence of the sample ν (Fig. 3E), shown in detail in the inset.
Then we assembled the system and imaged the surface temperature using an IR camera after attaching the carbon tape on the chip surface to achieve a uniform emissivity close to unity. The chip was mounted on two heaters with the temperature set at ≈ 50 °C and ≈ 80 °C in order to create the ∇T along the microchannel inside the chip. The first set of IR images was captured without ν of the water sample (Fig. 4A), and the second set of images was taken with the ν set to 0.5 µL·s−1 (Fig. 4B), all in a monochromatic fashion. Here, we used false colors for contrast enhancement. The temperature profile alongside the center microchannel—indicated by the black line in Figs. 4A and 4B—was extracted from the IR images and plotted in Fig. 4C. We found that the value of ∇T between the two heaters was constant, confirming the negligible influence of convection and radiation on heat losses as simulated by FEM. The ν values of up to 0.5 µL·s−1 had only a marginal difference on the setup system shifting the T amplitude by (−2.8 ± 4.5) 0.001·A.U. (mean ± measurement error) (Fig. 4C inset). This temperature shift does show the induced influence, albeit only of a marginal value with a relatively large fitting error.
TM determination by a flow-through system
Previously, we showed a method of precise TM determination, temperature calibration, or heat transfer rate measurement using dsDNA in the presence of a fluorescent intercalator such as SYBR Green I or Eva Green for stationary droplets40, as well as a flow-through system39. For the latter, we used a microscope to capture and analyze fluorescence images from the microchannel having ∇T. Here, we replaced a bulky microscope with a thermally regulated optofluidic platform to characterize the proposed detection system in a bulk sample without ∇T. First, we filled the microchannel with a dsDNA sample. We set T1 = T2 and gradually increased their temperature values, starting from ≈ 30 °C to ≈ 85 °C with ≈ 5 K increments while monitoring the VL amplitude (Fig. 5A). The measurement at a determined temperature was followed by washing the solution that interacted with laser light with a fresh solution; then, we waited for ≈ 10 s for the VL signal to stabilize the photobleaching effect. The VL value for each temperature was read when the temperature changed on both heaters, and the signal dropped and stabilized as indicated by the arrows in Fig. 5A. Then we plotted the VL values as a function of temperature and performed a nonlinear curve fitting using the sigmoidal Boltzmann function (Fig. 5B black line). Figure 5B also shows negative numerical derivative of VL with respect to temperature (Fig. 5B blue line). The maximum of this curve is the value of TM as (59.9 ± 0.2)°C (mean ± fitting error), which was close to the TM value measured earlier using the commercial qPCR system.
TM and heat transfer determination using segmented flow
Further, we conducted an experiment using segmented flow to demonstrate a rapid MCA of the dsDNA at ∇T. The segmented flow was generated off-chip using a double T-junction, as it was found that the uniformity of produced segments was more stable than one generated within a heated chip.
We controlled the droplet generation as well as the ratio between oil and water phases using external p values set from ≈ 25 kPa to ≈ 75 kPa, respectively. We recorded the VL signal for ≈ 250 s, getting its value to alternate between ≈ 300 mV and ≈ 3 V for reference droplets and the sample, respectively (Fig. 6A), with the length of the droplet estimated to be ≈ 200 µm (Fig. 6A inset).
The VL amplitude increases above the baseline corresponding to the droplet entering the optofluidic path, reaching the maximum when the whole droplet is inside the optofluidic microchannel (Fig. 6B). The droplet there was exposed to ∇T inside the microchannel; thus, the value of VL dropped. This decrease of the VL value is nonlinear due to the nature of dsDNA’s melting characteristic. We extracted the part of the curve corresponding to ∇T exposure (red arrowed line) and performed a non-linear curve fitting using the Boltzmann (sigmoid) function with respect to time, subsequently converting time to location and temperature39. The fitted curve was processed by derivation -dV/dT (Fig. 6C) with flow speed (r) values (Table 1) as a parameter. The values of r were determined by the time (t) required for a droplet to pass the whole length of ≈ 24 mm of microfluidic microchannel and then r = t/24. Subsequently, we expressed the TM as a function of r values (Table 1 and Fig. 6D). Figure 6D shows how the TM values increase with increasing r corresponding to the heat transfer rate. From this curve, the slope can be extracted to calculate the time required for heat to reach the center of the microchannel from the channel sidewalls.
Thus, we performed linear fitting to obtain the slope (S) as (1.28 ± 0.04) K·s·mm−1 (mean ± fitting error). The S value divided by ∇T of ≈ 2.31 K mm−1 produces a heat transfer time of ≈ 554 ms. Knowledge of this value can be used to correct the system results when they are collected at different values of r.
The MCAs are typically performed using commercial real-time PCR systems with a typical temperature ramp rate between ≈ 0.1 and ≈ 0.5 K·s−1, thus corresponding to 500 s to 100 s per test (50 K temperature ramping), excluding sample loading41. An ultrafast method to perform the MCA in an astonishing ≈ 50 ms has been reported42. It was based on a droplet placed on a cold Cu substrate with the sample heated by laser irradiation that required a complex setup and careful sample preparation. Our optofluidic platform enables rapid MCA of the sample containing biomolecules such as DNA or proteins. The time needed for the analysis is less than ≈ 30 s, which makes this platform faster than other systems. That can significantly speed up the MCA and make it suitable for high-throughput screening.
Conclusion
We proposed an optofluidic chip having ∇T along the microchannel to perform the MCA of dsDNA or protein unfolding. We first numerically analyzed the influence of convection, radiation, and a sample ν on the temperature distribution, as well as ∇T deviation from a constant value. The optofluidic chip was equipped with a buried microchannel with a diameter of ≈ 500 µm. We inserted both optical fibers and microcapillaries inside the chip, forming an integrated and robust autonomous system. The microfluidic channel served as a light guide for sensitive fluorescence detection along the ∇T inside the chip. We demonstrated this concept by determining the dsDNA TM in a continuous-flow configuration. We also conducted the MCA in a droplet-based configuration using segmented flow. This concept of fluorescence monitoring in a microchannel exposed to ∇T represents a fast and cost-effective approach for the characterization of thermal properties of biomolecules. It could be crucial in many areas of biology and chemistry, including the stability of proteins, by exposing them to different environments such as pH, buffer composition and ionic strength, purity control, and protein-ligand interaction. Those applications could have a great impact on drug discoveries as well as the molecular diagnostics of infectious diseases or genotyping.
References
Dolan, E. B., Haugh, M. G., Tallon, D., Casey, C. & McNamara, L. M. Heat-shock-induced cellular responses to temperature elevations occurring during orthopaedic cutting. J R Soc 9, 3503–3513, https://doi.org/10.1098/rsif.2012.0520 (2012).
Vergara, M. et al. Differential Effect of Culture Temperature and Specific Growth Rate on CHO Cell Behavior in Chemostat Culture. PLOS ONE 9, e93865, https://doi.org/10.1371/journal.pone.0093865 (2014).
Katava, M. et al. Critical structural fluctuations of proteins upon thermal unfolding challenge the Lindemann criterion. Proc. Natl. Acad. Sci. USA, 201707357, https://doi.org/10.1073/pnas.1707357114 (2017).
Tanaka, J. et al. KRAS genotyping by digital PCR combined with melting curve analysis. Sci Rep 9, 2626, https://doi.org/10.1038/s41598-019-38822-1 (2019).
Miralles, V., Huerre, A., Malloggi, F. & Jullien, M.-C. A Review of Heating and Temperature Control in Microfluidic Systems: Techniques and Applications. Diagnostics 3, 33–67, https://doi.org/10.3390/diagnostics3010033 (2013).
Sohel Murshed, S. M., Tan, S.-H. & Nguyen, N.-T. Temperature dependence of interfacial properties and viscosity of nanofluids for droplet-based microfluidics. J. Phys. D 41, 085502, https://doi.org/10.1088/0022-3727/41/8/085502 (2008).
Murshed, S. M. S., Tan, S. H., Nguyen, N. T., Wong, T. N. & Yobas, L. Microdroplet formation of water and nanofluids in heat-induced microfluidic T-junction. Microfluid Nanofluid 6, 253–259, https://doi.org/10.1007/s10404-008-0323-3 (2009).
Verneuil, E., Cordero, M., Gallaire, F. & Baroud, C. N. Laser-Induced Force on a Microfluidic Drop: Origin and Magnitude. Langmuir 25, 5127–5134, https://doi.org/10.1021/la8041605 (2009).
Yesiloz, G., Boybay, M. S. & Ren, C. L. Effective Thermo-Capillary Mixing in Droplet Microfluidics Integrated with a Microwave Heater. Anal Chem 89, 1978–1984, https://doi.org/10.1021/acs.analchem.6b04520 (2017).
Khater, A., Mohammadi, M., Mohamad, A. & Nezhad, A. S. Dynamics of temperature-actuated droplets within microfluidics. Sci Rep 9, 3832, https://doi.org/10.1038/s41598-019-40069-9 (2019).
Ho, P. C. & Nguyen, N.-T. Numerical study of thermocoalescence of microdroplets in a microfluidic chamber. Phys. Fluids 25, 082006, https://doi.org/10.1063/1.4819134 (2013).
Liu, J.-H., Hsia, K.-C., Yokokawa, R. & Lu, Y.-W. Microtubule polymerization in alignment by an on-chip temperature gradient platform. Sens. Actuator B-Chem. 298, 126813, https://doi.org/10.1016/j.snb.2019.126813 (2019).
Li, K.-C., Ding, S.-T., Lin, E.-C., Wang, L. A. & Lu, Y.-W. Melting analysis on microbeads in rapid temperature-gradient inside microchannels for single nucleotide polymorphisms detection. Biomicrofluidics 8, 064109–064109, https://doi.org/10.1063/1.4902907 (2014).
Dodge, A., Turcatti, G., Lawrence, I., de Rooij, N. F. & Verpoorte, E. A Microfluidic Platform Using Molecular Beacon-Based Temperature Calibration for Thermal Dehybridization of Surface-Bound DNA. Anal Chem 76, 1778–1787, https://doi.org/10.1021/ac034377+ (2004).
Neuzil, P., Pipper, J. & Hsieh, T. M. Disposable real-time microPCR device: lab-on-a-chip at a low cost. Mol. Biosyst. 2, 292–298, https://doi.org/10.1039/B605957K (2006).
Davies, E., Christodoulides, P., Florides, G. & Kalli, K. Microfluidic Flows and Heat Transfer and Their Influence on Optical Modes in Microstructure Fibers. Materials 7, 7566–7582, https://doi.org/10.3390/ma7117566 (2014).
Toh, K. C., Chen, X. Y. & Chai, J. C. Numerical computation of fluid flow and heat transfer in microchannels. Int. J. Heat Mass Transf 45, 5133–5141, https://doi.org/10.1016/S0017-9310(02)00223-5 (2002).
Hajmohammadi, M. R., Alipour, P. & Parsa, H. Microfluidic effects on the heat transfer enhancement and optimal design of microchannels heat sinks. Int. J. Heat Mass Transf 126, 808–815, https://doi.org/10.1016/j.ijheatmasstransfer.2018.06.037 (2018).
Novak, L., Neuzil, P., Pipper, J., Zhang, Y. & Lee, S. An integrated fluorescence detection system for lab-on-a-chip applications. Lab Chip 7, 27–29 (2007).
Ahrberg, C. D., Ilic, B. R., Manz, A. & Neužil, P. Handheld real-time PCR device. Lab Chip 16, 586–592 (2016).
Ahrberg, C. D., Manz, A. & Neužil, P. Palm-sized device for point-of-care Ebola detection. Anal Chem 88, 4803–4807 (2016).
Neuzil, P. et al. Rapid detection of viral RNA by a pocket-size real-time PCR system. Lab Chip 10, 2632–2634 (2010).
Ozcelik, D., Cai, H., Leake, K. D., Hawkins, A. R. & Schmidt, H. Optofluidic bioanalysis: fundamentals and applications. Nanophotonics 6, 647–661, https://doi.org/10.1515/nanoph-2016-0156 (2017).
Fan, X. & White, I. M. Optofluidic microsystems for chemical and biological analysis. Nat Photonics 5, 591–597, https://doi.org/10.1038/nphoton.2011.206 (2011).
Song, C. & Tan, S. H. A Perspective on the Rise of Optofluidics and the Future. Micromachines (Basel) 8, 152, https://doi.org/10.3390/mi8050152 (2017).
Persichetti, G., Grimaldi, I. A., Testa, G. & Bernini, R. Multifunctional optofluidic lab-on-chip platform for Raman and fluorescence spectroscopic microfluidic analysis. Lab Chip 17, 2631–2639, https://doi.org/10.1039/C7LC00460E (2017).
Yang, T., Stavrakis, S. & deMello, A. A High-Sensitivity, Integrated Absorbance and Fluorescence Detection Scheme for Probing Picoliter-Volume Droplets in Segmented Flows. Anal Chem 89, 12880–12887, https://doi.org/10.1021/acs.analchem.7b03526 (2017).
Cai, H. et al. Optofluidic analysis system for amplification-free, direct detection of Ebola infection. Sci Rep 5, 14494–14494, https://doi.org/10.1038/srep14494 (2015).
Scherr, S. M. et al. Real-Time Capture and Visualization of Individual Viruses in Complex Media. ACS Nano 10, 2827–2833, https://doi.org/10.1021/acsnano.5b07948 (2016).
Zhu, H., Mavandadi, S., Coskun, A. F., Yaglidere, O. & Ozcan, A. Optofluidic fluorescent imaging cytometry on a cell phone. Anal Chem 83, 6641–6647, https://doi.org/10.1021/ac201587a (2011).
Balram, K. C. et al. The Nanolithography Toolbox. J. Res. Natl. Inst. Stand. 121, 464–475, https://doi.org/10.6028/jres.121.024 (2016).
Gablech, I. et al. Fabrication of buried microfluidic channels with observation windows using femtosecond laser photoablation and parylene-C coating. Microfluid Nanofluid 22, 105 (2018).
Ilic, B., Czaplewski, D., Zalalutdinov, M., Schmidt, B. & Craighead, H. Fabrication of flexible polymer tubes for micro and nanofluidic applications. Journal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena 20, 2459–2465 (2002).
Laerme, F., Schilp, A., Funk, K. & Offenberg, M. In Technical Digest. IEEE International MEMS 99 Conference. Twelfth IEEE International Conference on Micro Electro Mechanical Systems (Cat. No. 99CH36291). 211–216 (IEEE).
Reismann, M., Bretschneider, J. C., Plessen, G. V. & Simon, U. Reversible Photothermal Melting of DNA in DNA–Gold-Nanoparticle Networks. Small 4, 607–610, https://doi.org/10.1002/smll.200701317 (2008).
Castro, E. R. et al. Determination of dynamic contact angles within microfluidic devices. Microfluid Nanofluid 22, 51 (2018).
Svatos, V., Gablech, I., Ilic, B. R., Pekarek, J. & Neuzil, P. In situ observation of carbon nanotube layer growth on microbolometers with substrates at ambient temperature. J. Appl. Phys 123, 7, https://doi.org/10.1063/1.5016465 (2018).
Svatoš, V., Gablech, I., Pekárek, J., Klempa, J. & Neužil, P. Precise determination of thermal parameters of a microbolometer. Infrared Phys Techn 93, 286–290 (2018).
Zhu, H. et al. Heat transfer time determination based on DNA melting curve analysis. Microfluidics and Nanofluidics 24, 7 (2020).
Neuzil, P., Cheng, F., Soon, J. B. W., Qian, L. L. & Reboud, J. Non-contact fluorescent bleaching-independent method for temperature measurement in microfluidic systems based on DNA melting curves. Lab Chip 10, 2818–2821, https://doi.org/10.1039/c005243d (2010).
Bartsch, M. S. et al. The rotary zone thermal cycler: a low-power system enabling automated rapid PCR. PloS one 10, e0118182–e0118182, https://doi.org/10.1371/journal.pone.0118182 (2015).
Baaske, P., Duhr, S. & Braun, D. Melting curve analysis in a snapshot. Applied Physics Letters 91, 133901 (2007).
Acknowledgements
The authors wish to acknowledge the financial support of the W099109 grant from P.R. China, the GACR Project Number GA16-11140S from the Czech Republic, and the technical support provided by Nano+ and by Centrum SIX of Brno University of Technology, Czech Republic. LY and SN acknowledge the funding from the Research Grant Council of Hong Kong, received under Grant Number 16209316. The authors would like to thank Peter Fecko and Jakub Sommer for their help with chip fabrication.
Author information
Authors and Affiliations
Contributions
Z.F. performed analysis, intepretation of data and wrote the main manuscript text H.Z. was responsible for chip design and fabrication J.H. review & editing S.N. and L.Y. were responsible for FEM simulations and participation on writing the manuscript P.P. and A.O. were responsible for optofluidic setup and measurements P.N. writing – review & editing All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Fohlerova, Z., Zhu, H., Hubalek, J. et al. Rapid Characterization of Biomolecules’ Thermal Stability in a Segmented Flow-Through Optofluidic Microsystem. Sci Rep 10, 6925 (2020). https://doi.org/10.1038/s41598-020-63620-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-63620-5