Abstract
Deep brain stimulation (DBS) has emerged as a revolutionary technique for accessing and modulating brain circuits. DBS is used to treat dysfunctional neuronal circuits in neurological and psychiatric disorders. Despite over two decades of clinical application, the fundamental mechanisms underlying DBS are still not well understood. One reason is the complexity of in vivo electrical manipulation of the central nervous system, particularly in rodent models. DBS-devices for freely moving rodents are typically custom-designed and not commercially available, thus making it difficult to perform experimental DBS according to common standards. Addressing these challenges, we have developed a novel wireless microstimulation system for deep brain stimulation (wDBS) tailored for rodents. We demonstrate the efficacy of this device for the restoration of behavioral impairments in hemiparkinsonian mice through unilateral wDBS of the subthalamic nucleus. Moreover, we introduce a standardized and innovative pipeline, integrating machine learning techniques to analyze Parkinson’s disease-like and DBS-induced gait changes.
Similar content being viewed by others
Introduction
Originally developed for the symptomatic treatment of tremors, deep brain stimulation (DBS), as known today, emerged in 1987 as a groundbreaking therapeutic approach1 from ablative neurostimulation experiments as early as the 1930s2. Further development paved the way for the today-known DBS, and in 2002, DBS was granted approval by the US Food and Drug Administration (FDA)2,3. Its therapeutic scope has since broadened from essential tremor4 and Parkinson’s disease (PD)2, encompassing a range of neurological disorders, such as dystonia5, refractory epilepsy6, stroke7, and post-stroke pain7. In addition, the use of DBS is continuously expanding into the psychiatric field, finding application in disorders such as addiction8, Tourette’s syndrome9, post-traumatic stress disorder (PTSD)10, treatment-resistant depression11, and obsessive-compulsive disorder (OCD)12. This expanding use of DBS across numerous neuropsychiatric disorders highlights its substantial therapeutic potential and its growing socioeconomic role. However, neither the exact mechanism of action underlying the “make or break” effect of DBS on central brain networks, nor the molecular and cellular effects of beneficial DBS are well understood.
To better understand the complex biological mechanisms underlying DBS, translational research utilizing animal disease models is essential. However, one must decide between a wired or wireless system DBS systems in rodent DBS research, as wired and wireless systems offer distinct advantages and disadvantages, particularly in terms of ease of use and reproducibility. Wired DBS systems are commonly used in various rodent disease models13,14,15,16,17,18,19,20,21,22,23,24,25,26. While historically fundamental, they suffer from limitations, such as limited test radius and duration, as well as challenges in managing external components and potential damage with subsequent loss of data27. Conversely, albeit technically demanding to develop, wireless systems might offer greater flexibility and reproducibility, by eliminating the constraints of wired connections and external equipment28,29,30,31,32,33,34,35,36,37,38,39,40,41. Nevertheless, the design of small implantable wireless microstimulators remains a challenge, with various systems exhibiting shortcomings in size and power consumption, thus posing hurdles to long-term experiments and reproducibility27,28,29,38. Overall, while both wired and wireless DBS systems offer unique advantages, continued developments are imperative to improve usability and reproducibility in preclinical DBS research. One major technical drawback in rodent-based DBS research is the lack of standardized DBS devices27. Thus, various custom-made devices are used to apply DBS in translational research, in particular if wireless28,32,35. Furthermore, there have been reports of custom-designed devices for both unilateral and bilateral deep brain stimulation in mice32,35,42. As a result, the reproducibility of DBS effects may be compromised. Therefore, the transferability of DBS strategies across sites, for instance, in multi-site preclinical studies, becomes challenging. DBS devices should be ideally wireless, making miniaturization of wearable DBS devices technically challenging, especially for small rodents such as mice. To overcome these limitations, we present a novel standardized unilateral wireless microstimulation system designed for rodents. Given the value of unilateral PD models and the technical feasibility within the small dimensions of mice, we initially opted for a unilateral device. This device, comprising low-impedance microelectrodes and a versatile stimulation system (microstimulator), offers two key advantages: 1. It is wireless and can be implanted subcutaneously in mice and interferes minimally with their spontaneous behavior; 2. It can be programmed for frequency, current amplitude, and pulse width. To demonstrate the feasibility of the wireless microstimulation system for deep brain stimulation (wDBS), we here describe a step-by-step surgical implantation protocol. Additionally, we illustrate its application in a PD mouse model, where viral expression of human A53T-α-synuclein (hαSYN) in the substantia nigra (SN) induces PD-like phenotypes. Moreover, we demonstrate the utility of a deep learning-based gait analysis pipeline for extracting and quantifying PD-like features under wDBS conditions.
Results
Development of a wireless microstimulation system
We aimed to develop a standardized, wearable microstimulation system suitable for translational DBS research in freely moving animal models. Recognizing that DBS systems in human patients offer programmable options for key stimulation settings such as current amplitude, stimulation frequency, and pulse width, we integrated similar programmability into our wDBS microstimulation system.
The wDBS microstimulation system consists of the microstimulator, which can be connected to an implanted stimulation microelectrode (Fig. 1a). This design provides flexibility in using various types of microelectrodes, differing for instance, in length, in combination with the microstimulator.
Stimulation microelectrode
We used quartz-glass insulated platinum–tungsten fiber microelectrodes43 for stimulation (Fig. 1b–d, Table 1, Supplementary Fig. 1a). The shaft of the microelectrodes is just 100 µm in diameter and carries a small tip (Fig. 1b). These dimensions cause minimal tissue damage43,44 and allow a high spatial resolution and stimulation selectivity for the DBS process. Finally, to ensure the reproducibility of wDBS experiments, we have rigorously standardized the stimulation microelectrodes with respect to dimensions and electrical properties.
The microelectrodes feature a porous structure of microgrooves to enlarge the metal surface area in order to increase the charge transfer capacity and to reduce electrode–tissue impedance and polarization of the microelectrode. The platinum–tungsten fiber microelectrode tips are coated with a thin film of iridium metal. Due to the microgrooves in the surface structure of the microelectrodes, the iridium oxide coating is also embedded within the microgrooves, enhancing resistance to stripping during implantation. Post-coating, the iridium is activated by cycling the electrochemical potential of the metal between negative and positive potential limits within the water window to obtain an activated iridium oxide coating (AIROF). The AIROF coating is a low-impedance coating that provides a means of injecting charge into the tissue, thereby minimizing irreversible electrochemical processes at the electrode–tissue interface45.
To reduce the implantation time and to make the implantation procedure more feasible, we have integrated a reference microelectrode wire directly into the stimulation microelectrode unit. A platinum–iridium wire was placed parallel to the stimulation microelectrode fiber extending nearly 1 mm from the electrode connection board. This design allows an electric contact of the reference microelectrode at the level of the dura (Fig. 1a, c, d). Thus, unnecessary contact between screws (stainless steel or titanium) and the microelectrode reference wire (platinum or silver) is prevented, avoiding galvanic reactions with body fluids and subsequent corrosion processes. Furthermore, to facilitate precise placement of the microelectrode, we equipped it with a specialized thread. This feature allows for accurate guidance of the microelectrode to the targeted area within the brain. After fixation of the microelectrode, the thread can be removed easily without altering the final position of the microelectrode. This approach simplifies the implantation process while ensuring the accuracy and stability of the microelectrode’s placement.
Implantable microstimulator
The microstimulator (Fig. 1e, Table 2) delivers monophasic pulses with a stepwise adjustable pulse width of tp = 60–500 µs (steps of ~50 µs until 300 µs), amplitudes in a range of ip = 10–500 µA (steps of ~5 µA until 150 µA) and frequencies between f = 1/T = 10–200 Hz (steps of ~10 Hz). These settings can be programmed individually at any time before and after implantation. During pulse duration, a constant current (ip) applies charge to the tissue. The current source is turned off between the pulses, resulting in a zero-voltage output (Fig. 1f). During this time, the accumulated current can discharge across the 1 kΩ resistor (Fig. 1f), preventing charge accumulations. A reed contact allows the investigator to turn the microstimulator on and off with a magnet. The current operating mode can be monitored through a high-power white LED, visible through the animal’s skin.
The implantable microstimulator is powered by two 1.5 V cell batteries. The battery holder is designed for quick assembly without soldering during the manufacturing process. In vitro tests showed a maximum battery life of 52 days (Supplementary Fig. 1b).
To increase its longevity, the microstimulator is embedded in biocompatible housing (Supplementary Fig. 1c). This seals the microstimulator and protects the electric device from body fluids. The housing is constructed using a visible light-curable adhesive28, known for its flexibility post-curing. This adhesive is compatible with various materials, including polyimide, polyvinyl chloride (PVC), acrylonitrile-butadiene-styrene (ABS), and other plastics, as well as dissimilar materials such as glass, stainless steel, and epoxy boards. The housing has a high tolerance for thermal cycling and is an excellent adhesive for strain relief of the microstimulator connection cables. The adhesive used has passed the testing required for United States Pharmacopeia (USP) Class VI biocompatibility approval and can withstand standard sterilization methods like gamma and ethylene oxide. An eyelet has been incorporated into the housing to allow the microstimulator to be sewn to the fascia to minimize post-implantation movements of the device if required. The outer shape of the housing is smooth and has no sharp edges to avoid injury to surrounding tissue. Two connection cables exit the housing. First, a two-pin cable connects the microstimulator output to the implanted stimulation microelectrode. Second, a four-pin cable is implanted to connect an external programmer to the implanted microstimulator (Fig. 1g), allowing stimulation setting adjustments even after implantation.
Wireless deep brain stimulation implantation is technically feasible in mice
To test the device, we performed wDBS in the subthalamic nucleus (STN) of a well-established PD mouse model46,47,48. The surgical procedure was well tolerated by all animals. The mice regained normal activity after approximately 90 min and had a stable body weight of 101% three days (day 3) after implantation compared to the weight on the day of surgery (day 0) (Fig. 2a).
We evaluated the accuracy of the microelectrode placement within the STN (Fig. 2b). Effective stimulation of STN neurons was confirmed by immunolabeling of cFos. Local de novo synthesis of cFos can be used to monitor induced neuronal activation49. The precise microelectrode positions in all tested animals are shown in Fig. 2c. In 13 of the 18 mice, the microelectrode was placed exactly within the neuroanatomical structure of the STN. In three animals, the microelectrode tip was in the functional vicinity of the STN, as indicated by the corresponding cFos signal in the STN. In these cases, the tip of the microelectrode was within 200 µm of the STN. In two mice, the microelectrode positioning was not successful. These results demonstrate that both the newly designed microelectrodes and the wDBS microstimulator are technically feasible for implantation but also meet the technical requirements for effective neuronal activation in small deep brain structures.
Wireless STN-DBS improves behavioral symptoms of PD mice
Next, we evaluated the impact of STN-DBS with our wDBS setup on motor function in the hαSYN PD mice. This model is characterized by specific motor impairments46,47,48, that have been shown to improve with high-frequency STN-DBS in hαSYN PD rats50. As PD patients demonstrate severe disease-related multimodal gait dysfunction51, we investigated gait in PD mice using a single-lane treadmill. Video recordings of freely behaving mice were analyzed with markerless pose estimation using DeepLabCut (DLC)52. We developed a novel unbiased convolutional neural network (CNN)-based scoring system (Parkinson’s disease severity score for mice, PD-SSm) to quantitatively evaluate multiple behavioral outcomes. In this study, we delineated three categories of features encompassing body pose, body rotation, and stride, each containing characteristics potentially indicative of Parkinson’s disease (PD) dysfunctions. Thus, a total of 15 distinct features were identified, representing clinical manifestations such as posture instability, reduced stride length, and gait irregularities (see methods section for details).
Initially, the gait performances of wild-type mice with implants (WTOFF) were compared to those of implanted PD hαSYNOFF mice to evaluate the PD-specific nature of our PD-SSm. The WTOFF group exhibited a mean PD-SSm score of 14 out of 45. In contrast, the hαSYNOFF animals displayed more pronounced gait impairments, as evidenced by a significantly higher average score of 25 (Supplementary Fig. 2a). The individual scores for each feature are provided in Supplementary Table 1.
Subsequently, the gait performances under stimulation conditions were assessed in hαSYN animals. Our analysis revealed that stimulation at 60 µA does not improve gait in PD mice. However, wDBS at 100 µA significantly reduced the PD-SSm by 25% compared to control mice (hαSYNOFF, Fig. 2d, Supplementary Fig. 2b). To underline the sensitivity of the PD-SSm, we also analyzed the motor performance of the same animals using the cylinder test. This test is a well-established standard test to detect PD-related motor impairment in rodents46,48. The cylinder test did not elucidate the significant difference in ipsilateral paw use when comparing hαSYNOFF and hαSYNON(100µA) (Fig. 2e), likely due to the variance inherent in this behavioral test.
To rule out that the implantation procedure per se influences the gait of healthy mice, we compared the gait performances of untreated WT mice with those implanted with the microstimulation system (WTOFF). Our analysis revealed no significant alterations in gait induced solely by the implantation process itself (Supplementary Fig. 2c). Next, we investigated the effects of STN-DBS on gait in healthy WT mice. The animals received STN-DBS at intensities of 60 and 100 µA. Comparing the WTOFF, WTON(60µA) and WTON(100µA) groups showed no significant differences, indicating that STN-DBS at these intensities does not induce gait abnormalities in healthy animals (Supplementary Fig. 2d).
Discussion
In this study, we introduce a standardized wDBS system specifically designed for use in small animal models, including mice. This system, in combination with improved options for stereotaxic implantation to small brains, has several advantages. First, its design facilitates high-precision and reliable implantation procedures, even in exceptionally small brain regions like the mouse STN. Second, the small size and light weight of the microstimulator reduce the surgical burden on the animals. Third, the rigorous standardization of the system, from fabrication to skull attachment, minimizes the risk of adverse outcomes.
The adaptability of stimulation settings is a key feature of our system, allowing for immediate modifications as required. The combined implantation of the microstimulator and the microelectrode unit ensures high mobility for the animal, contrasting sharply with stationary stimulation methods that often necessitate restraining the animal with tethered, cable-bound stimulation systems20,27,50. Consequently, our wDBS system supports extended stimulation protocols with minimal side effects, while providing easy integration with a variety of behavioral paradigms. Since most preclinical studies utilize rodent models with hemiParkinsonism48,50,53,54,55, our microstimulation system can be readily applied in this context.
In contrast to the prior stimulator developed by Fleischer et al.28, this microstimulator exhibits a weight reduction of 0.8 g, extends battery life by 22 days (52 in total), and introduces adaptive (re-)programmability. Notably, the microelectrode is insulated with quartz glass instead of Teflon, and the core dimensions have been reduced from 50 to 30 µm.
Considering that gait disturbances are commonly observed in both PD patients and PD animal models, we have developed and applied a machine learning-based automated analysis of specific gait patterns using treadmill video recordings of mice. This innovative approach, combining established DLC markerless tracking with a custom-coded machine-learning algorithm, enabled us to extract PD-like features from mice with wDBS. Our analysis indicated that STN-stimulation at 100 µA with the wDBS system significantly improved gait in contrast to stimulation at 60 µA, likely due to a greater electrochemical infiltration of the tissue. The behavioral pipeline we developed allows us to track the influences of the standardized wDBS at the behavioral level by revealing influences on single gait characteristics summarized in the PD-SSm. The integration of our standardized wDBS with this analysis pipeline represents a significant step forward in ensuring reproducible and comparable results in mouse DBS research.
However, there are limitations to our current setup. The wDBS device is currently designed only for unilateral and unipolar one-contact stimulation, and its battery life is limited to 52 days. Therefore, more sophisticated approaches such as bilateral stimulation or long-term DBS (>50 days) are not feasible with the current design. Furthermore, stimulation was limited to 10 min before and during behavioral testing. The wDBS device is not yet fully implantable, as the programming interface remains externalized for adjusting stimulation settings. Moreover, for group housing of rodents, the externalized components of the device need to be modified to prevent harm to both the rodents and the device. Our current wDBS design theoretically allows for the recording of local field potentials (LFP) and multiunit activity (MUA), but recording cannot be performed simultaneously with stimulation and requires further development for integration into preclinical research. In addition, the reusability of once-implanted microelectrodes has not been established, as they are usually damaged during the explantation process.
Finally, the gait analysis pipeline presented here is currently limited and adjusted to Parkinsonian gait features. However, due to its design, the pipeline can easily be adapted by further developing the training data set supporting research also in other fields of movement disorders.
In summary, we present a standardized wDBS system suitable for the stimulation of small brain areas in rodents. We also introduce a pathway for standardized, automated gait analysis based on sub-categories. Together, these advancements will contribute to the standardization of translational DBS research, particularly in the study of circuitopathies.
Methods
Animals and ethics statement
All experiments were performed in accordance with the guidelines of the European Union and approved by our institutional Animal Care, the Utilization Committee and the Regierung von Unterfranken, Würzburg, Germany (license number: RUF-55.2.2-2532-2-1368). All experiments were performed at the ‘Zentrum für Experimentelle Molekulare Medizin’ Würzburg, Germany. Male C57Bl/6J mice (Charles River, Sulzfeld, Germany) were kept under standard laboratory conditions (21 °C, 12 h/12 h light/dark cycle, food and water ad libitum). For the investigation, mice were used at the age of 11–12 weeks. All experimenters carried a certificate of competence in accordance with German laws on animal welfare.
Generation of Parkinson’s disease model
Stereotaxic injection of AAV1/2 vectors expressing the A53T mutant of human synuclein was used to develop the PD mouse model48. AAV1/2 vectors were purchased from the Michael J. Fox Foundation for Parkinson’s Research (MJFF) partnered with GeneDetect (Auckland, New Zealand) and stereotaxic injection to the SN was performed as previously described48. In short, the AAV1/2 vectors were injected unilaterally into the right SN using an injector-controlled 5 µl Hamilton syringe. A volume of 2 µl of AAV1/2-A53T-αSYN (5.16 × 1012 genome copies (gc) per ml) was injected with a speed of 0.25 µl/min. Coordinates56 were (in mm from Bregma): anterior-posterior (AP) −3.1; medio-lateral (ML) 1.4; dorso-ventral (DV) 4.4. In total, 18 mice were injected with AAV1/2-A53T-αSYN of which 7 mice were included in the final behavioral analysis. Intraoperative and postoperative pain therapy was performed using carprofen (5 mg/kg body weight, subcutaneously).
Implantation and stimulation
All experimenters need a certificate of competence in accordance with the laws on animal welfare. The welfare of the animal must be considered a high priority for the entire experiment. During surgical procedures, constant monitoring of the animal’s respiration, temperature, and reflexes is necessary. The experimenter must be aware of the use of chemicals and tools that may be hazardous to health if not performed properly. For the stereotaxic surgery, a semi-automatic device (Neurostar Robot) was used for virus injection and microelectrode implantation.
The stimulation microelectrode was stereotaxically implanted into the right STN and fixed with dental restorative and sutures. Coordinates56 were (in mm from Bregma): AP −2.03; ML 1.60; DV 4.7. The microstimulator was placed subcutaneously over the spine into a skin pocket of 2.5 × 2 cm. Microstimulator and microelectrode were connected, and stimulation settings were programmed. The microstimulator was switched off until behavioral investigation on the treadmill after 4 days. Preoperative and intraoperative pain therapy was performed using carprofen (5 mg/kg body weight, subcutaneously) and bupivacaine 0.25% locally. The implantation procedure is visualized in the Supplementary Fig. 3 and described in detail in the following protocol:
Initiation and implantation of the microstimulator:
-
1.
Inject 5 mg/kg bodyweight carprofen in NaCl subcutaneously (s.c.).
-
2.
Place the mouse in the anesthesia box until the pedal reflex is negative. Input = Isoflurane/O2-pump (3 bar O2 and 4 vol% isoflurane)
Output = Absorber.
-
3.
After the initial anesthesia, carefully place the head of the anesthetized mouse into the isoflurane mask (Supplementary Fig. 3a).
Input = Isoflurane/O2-pump (1 bar O2 and 2 vol% isoflurane)
Output = Absorber.
-
4.
Insert the rectal temperature probe and tape it to the tail. Place the heating pad below the animal and run the pre-installed mouse settings.
-
5.
Shave the head and back of the animal and remove any shed hair.
-
6.
Apply Bepanthen eye protection ointment to the eyes to prevent corneal damage and subsequent pain.
-
7.
Inject 0.25% bupivacaine locally.
-
8.
Make a horizontal central incision (1 cm length) about 1 cm posterior from the ears and another vertical incision on the head until you see Bregma and Lambda (Supplementary Fig. 3b).
-
9.
Head treatment: Stop bleeding with Q-tips, disinfect the area and dry the skull with ethanol.
-
10.
Preparation for microstimulator implantation: Create an artificial space (“back pocket”) between the skin and muscle fascia by carefully sliding the surgical scissors under the skin and opening the scissors (Supplementary Fig. 3c). Cave: Do not cut after the initial skin opening—use only the blunt outer surface of the scissors to create an artificial space. The space should have an area of 2.5 cm × 2 cm (length × width). Disinfect the incision area and the created space with a Q-tip soaked in disinfectant. Repeat the disinfection procedure at least twice with a new Q-tip.
-
11.
Use the scissor to create a subcutaneous tunnel connecting the opened skull area to the artificial space on the back; the cable will later pass through this tunnel.
-
12.
Carefully disinfect the microstimulator and insert it into the ‘back pocket’. Leave the two sets of cables outside (Supplementary Fig. 3d). The microstimulator control light should be facing up.
-
13.
Sew the skin between the cable sets to secure the microstimulator device inside the mouse (Supplementary Fig. 3e).
-
14.
Insert the ear bars properly into the mouse’s ear and ensure that the head is aligned with the mask.
Skull treatment:
-
15.
Take a dummy electrode and attach it to the holder of the stereotaxic frame using the thread and metal rod. The dummy electrode has the same dimensions as the stimulation electrode but with a metal wire. It is, therefore less fragile and can be used to touch the skull for skull calibration.
-
16.
Set the tip of the dummy electrode to Bregma.
(‘Synchronize syringe and drill’ → ‘Syringe’ → ‘Set Bregma’)
-
17.
Go to Lambda and check the DV difference. Change the position of the head and repeat steps 13 and 14 until Bregma and Lambda are planar (maximum deviation ± 0.03 mm). Do it in this comparative way, depending on the stereotaxic frame you use.
-
18.
Move the drill to Bregma.
(‘Synchronize syringe and drill’ → ‘Drill’ → ‘Set Bregma’).
-
19.
Go to coordinates on the contralateral side, where the fixation screw will be positioned: e.g. AP −0.9; ML −1.8; DV −1.
-
20.
Place the tip of the drill bit on the cranial surface and drill a notch (not a hole) in 50 µm increments (approximately 250 µm in total).
-
21.
Go to: AP −2.03; ML 1.6; DV −1.
-
22.
Place the drill on the surface of the skull and drill a hole in 50 µm increments (~500 µm) (Supplementary Fig. 3f).
-
23.
Stop any bleeding with a Q-tip and clean the skull as described above.
-
24.
Use the sharp tip of a cannula and check if the dura has been removed. If not, carefully pierce the dura with the cannula. The intact dura may damage the electrode or affect the insertion direction.
-
25.
Insert a previously disinfected (70% ethanol) screw into the notch until half of the shaft is fixed (Supplementary Fig. 3g). Do not screw any further as the risk of loss is high due to the thin architecture of the mouse skull bones.
-
26.
Reposition the needle and verify that bregma and lambda are still at the same planar level. If not, repeat step 16.
Electrode implantation:
-
27.
Take the stimulation electrode and screw the metal rod into the thread of the electrode assembly; the following steps must be carried out with great care. The electrode itself is fragile. Set the tip of the dummy electrode to Bregma.
-
28.
Insert the metal rod into the stereotaxic frame holder.
-
29.
Go to Bregma, close above the skull surface (do not touch!)
(‘Synchronize syringe and drill’ → ‘Syringe’ → ‘Set Bregma’)
-
30.
Go to: AP −2.03; ML 1.60; DV −1.
-
31.
Move down slowly in the DV direction. The electrode should fit into the hole without skull fragments in the vicinity.
-
32.
Set the DV frame speed to 1 mm/min.
-
33.
Go to AP −2.03; ML 1.60; DV 4.7 (STN coordinates) (Supplementary Fig. 3h).
Wound closure and suture:
-
34.
Apply restorative around the electrode and build a connection to the screw. Use the UV lamp 2 times for 20 s.
-
35.
Gradually apply more restorative around the electrode, screw, and on top of the dry skull and expose it to UV light.
-
36.
Create a multiple-layered mounting covering up two-thirds of the electrode device (Supplementary Fig. 3i). Cave: Check thoroughly that the restorative is well hardened before proceeding to the next step.
-
37.
Gently unscrew the metal rod from the holder and move the holder away (anterior).
-
38.
Gently unscrew the metal rod from the thread. Due to the length of the metal rod, there is a significant leverage effect that can completely remove the implanted electrode if the procedure is not performed carefully.
-
39.
Remove the ear bars.
-
40.
Create the first suture (single button suture) anterior.
-
41.
Take the electrode connector cable and pass it through the tunnel to the front (clean it with a Q-tip).
-
42.
Connect the two cables (Supplementary Fig. 3i) and fix them by applying UV glue around the port (UV light).
-
43.
Push the connected cable to the back as far as possible.
-
44.
Pull the two sides of the skin together (over the cables) and sew them together.
-
45.
Fill in the holes between the stitches with restorative (UV light).
-
46.
Secure the second set of cables remaining outside the animal by stitching around the cable (Supplementary Fig. 3j). In addition, close any remaining holes on the back of the animal by stitching.
-
47.
Disinfect all treated areas again.
-
48.
Turn off the isoflurane pump, remove the temperature probe, release the animal, and place it on a heating plate until the animal wakes up.
Stimulation programming:
-
1.
Turn on the laptop and plug in the memory stick containing ‘Silicon Labs integrated development environment (IDE)’ and the file ‘Pulse.wsp’ (Thomas RECORDING GmbH).
-
2.
Run ‘Silicon Labs IDE’.
-
3.
Open a new project and select the file ‘Pulse.wsp’. Then the code appears.
-
4.
Define the settings frequency, amplitude and pulse width by entering in the corresponding values.
-
5.
Click ‘Rebuild all‘.
-
6.
Confirm that your code is successful (0 errors and 0 warnings). The manual supplied by Thomas RECORDING GmbH provides a guideline for troubleshooting in case of errors or warnings.
-
7.
Connect the yellow programming adapter to the laptop.
-
8.
Confirm the ON state of the control lamp on the device.
-
9.
Connect the programming adapter to the microstimulator implanted in the animal (pay attention to the aligning colors of the wires, representing the polarity).
-
10.
Click ‘Connect’ to couple the software with the microstimulator.
-
11.
Confirm the ON state of the second control lamp on programming adapter device.
-
12.
Click ‘Download code‘.
-
13.
In case of a successful download (read the message), click ‘Disconnect’.
Finally, unplug the devices and use the magnet to turn the programmed stimulation ON or OFF.
Experimental design
The experiment started with unilateral AAV1/2-expressing human mutated A53T-αSyn injection into the SN of the right hemisphere (day 0). Five weeks later, on day 35, the microstimulator and microelectrode were implanted in a single session. On day 39, animals were recorded under hαSYNOFF, hαSYNON(60µA) and hαSYNON(100µA) conditions on a treadmill. After a one-day washout period, the cylinder test was performed under hαSYNOFF and hαSYNON(100µA) conditions on day 40. Before the behavior recordings under hαSYNON conditions, the stimulation was active for 10 min. Finally, on day 41, the animals were perfused and brain tissue was harvested.
Tissue processing and immunofluorescence
Animals were stimulated with 140 µA for 10 min to induce cFos expression. 60–90 min after stimulation, animals were perfused transcardially with 0.34% heparin sodium-25000 (Ratiopharm) in 0.1 M phosphate-buffered saline (PBS). The tissue was post-fixed in 4% paraformaldehyde (PFA) in 0.1 M PBS for 48 h and immersed in 30% sucrose/0.1 M PBS solution for another 72 h. The brains were embedded in TissueTek O.C.T. compound using methylbutane and dry ice. Afterward, the brains were sectioned into 40 µm-thick coronal slices.
For immunofluorescence staining of cFos, sections were first washed in 0.1 M PBS and incubated in a quenching solution (100 mM glycine, buffered with Tris-base, pH 7.4) for 1 h at room temperature (RT). Sections were incubated in a blocking solution containing 10% goat serum, 0.3% Triton X-100, and 0.1% Tween20 for 1 h at RT. Rabbit anti-cFos primary antibody (c = 0.25 µg/ml) was used in the blocking solution for 24 h at RT. Slices were washed in 0.1 M PBS, 0.1% Triton X100, 0.1% Tween20. As a secondary antibody, donkey anti-rabbit Cy5 was used at a final concentration of 1 µg/ml. After washing, nuclei were stained with 4′,6-diamidino-2-phenylindole (DAPI) for 20 min at RT. After a final washing step in 0.1 M PBS, slices were embedded in Aqua-Poly/Mount.
Immunofluorescence staining for tyrosine hydroxylase (TH) and αSYN was performed as described before48 to verify the αSyn expression and the degeneration of dopaminergic neurons (TH+) in the SN. Therefore, primary antibodies chicken anti-TH and rabbit anti-αSyn and secondary antibodies goat anti-chicken AF488 and goat anti-rabbit Cy3 were applied (catalog numbers and RRIDs are described in Supplementary Table 2).
Treadmill analysis
For gait analysis, seven hαSYN animals were investigated without prior training. Gait behavior was recorded on day 39 post-hαSYN injection using a forced walking paradigm, on a single-unit treadmill with transparent sidewalls and a transparent treadmill track (DigiGait treadmill). Animals were simultaneously recorded from a ventral and lateral (left side exposed to the side camera) angle. The treadmill recording was performed at a speed of 20 cm/s. Initially, animals were recorded in the hαSYNOFF condition to capture the current gait impairment in the disease state. Subsequently, the wDBS device was switched on, and animals were recorded with 60 µA followed by 100 µA (frequency = 130 Hz; pulse width = 60 µs). This stimulation sequence (starting with low amplitude) remained identical for all animals used. Each recording took about 3–5 s. Gait behavior was recorded with 160 frames/s. A similar procedure was conducted on WT mice. Recordings were performed on six untreated WT mice and eight WT mice with implants under stimulation OFF, 60 and 100 µA conditions.
To evaluate the gait of the animals, we performed the PD-SSm evaluation. For this CNN-based evaluation, the open-source machine learning framework DLC was used for markerless detection of critical anatomical structures52. It leverages the power of deep neural networks, employing a modified version of the ResNet architecture, which is pre-trained on the ImageNet dataset. The framework consists of two main stages: training and analysis. During the training phase, a user labeled a subset of frames with the body parts of interest, and the model is fine-tuned on this labeled training data. The training employs a multi-stage process, including the use of transfer learning, to achieve accurate pose estimation. This process is described elsewhere in detail52. In the analysis phase, the trained network predicts the body part positions in new videos. DLC utilizes a bottom-up approach, where the network first detects body parts and then assembles them into full-body poses. This approach allows for tracking multiple animals or body parts simultaneously. The model’s predictions are refined through a part affinity field, ensuring spatial coherence between body parts. DLC’s architecture enables efficient training and inference, making it suitable for both laboratory and real-world applications. Its open-source nature and compatibility with low-cost hardware further contribute to its widespread use in behavioral research and clinical settings.
The markerless tracking was first cleaned by removing low-likelihood marker data points (p < 0.25, likelihood defined by DLC during prediction). The signals were not band pass filtered prior to analysis. We used these data to define specific gait-related traits to be included in the PD-SSm. First, we defined the three feature classes, body pose, body rotation, stride, and assigned features that captured structural aspects of the rodent's gait, including distances and angles between markers (Table 3): (I) the angle deviation of the line between the hip and tailbase (tb) in the lateral plane from horizontal, (II) the angle deviation of the line between the snout and tb in the ventral plane from horizontal, and (III) the Euclidean distance between the left paws in the ventral plane (Fig. 3a). Next, we engineered features that captured the temporal dynamics of the gait. Here, we calculated the power spectral density of each paw in the ventral plane using Welch’s method (Scipy 1.10.1), extracting (IV, V) the dominant frequency of the power spectrum and (VI, VII) measuring the amplitude of the dominant frequency (Fig. 3b). We limited the frequency domain to a maximum of 10 Hz and linear interpolation was used where low-likelihood data-points had been removed. Then, we identified each stride in the signal using peak detection (Scipy 1.10.1) allowing us to measure (VIII, IX) the swing and (X, XI) stance phases (Fig. 3c). We could further compute the time period T of each stride (which is related to frequency as 1/T) allowing us to capture (XII, XIII) the frequency and (XIV, XV) amplitude of each stride (Fig. 3c).
The same data was collected from a control group containing 13 healthy animals (C57Bl/6J mice), which were injected with an empty vector instead of the hαSYN viral vector. The confidence intervals 68%, 95%, and 99% were calculated for each feature from these control animals. The feature values within the interval 0–68% were scored with ‘0’, 68–95% with ‘1’, 95–99% with ‘2’, and above 99% with ‘3’. We defined this score as the PD-SSm. Since we included 15 behavioral features with a severity rating range of 0–3, the maximum score to be achieved was 45.
Cylinder test
Spontaneous forepaw use was assessed by the cylinder test 6 weeks after hαSYN injection as described earlier46,47. Data was collected from one cohort.
Statistics
Graph Pad Prism version 9.1.1 was used for statistical analysis. Normality was determined using Q–Q-plots. Multiple non-normal distributed data sets were analyzed by using a Friedman test followed by Dunn’s multiple comparisons test. Normal distributed data sets comparing multiple groups were statistically analyzed by one-way ANOVA and Tukey’s multiple comparisons test. In the case of two normally distributed data sets, we used a two-tailed t-test. (*) p < 0.05, (**) p < 0.01, (***) p < 0.001, and (****) p < 0.0001 were considered significant p values.
Data availability
The experimental datasets supporting the findings of this study are available upon reasonable request from the corresponding author. Supplementary Table 1 provides raw data on the PD-SSm.
Code availability
The code used for the gait feature analysis can be found at https://github.com/peach-lucien/rodent_dbs.
References
Benabid, A. L., Pollak, P., Louveau, A., Henry, S. & de Rougemont, J. Combined (thalamotomy and stimulation) stereotactic surgery of the VIM thalamic nucleus for bilateral Parkinson disease. Appl. Neurophysiol. 50, 344–346 (1987).
Gardner, J. A history of deep brain stimulation: technological innovation and the role of clinical assessment tools. Soc. Stud. Sci. 43, 707–728 (2013).
Miocinovic, S., Somayajula, S., Chitnis, S. & Vitek, J. L. History, applications, and mechanisms of deep brain stimulation. JAMA Neurol. 70, 163–171 (2013).
Deuschl, G., Raethjen, J., Hellriegel, H. & Elble, R. Treatment of patients with essential tremor. Lancet Neurol. 10, 148–161 (2011).
Kupsch, A. et al. Pallidal deep-brain stimulation in primary generalized or segmental dystonia. N. Engl. J. Med. 355, 1978–1990 (2006).
Tykocki, T., Mandat, T., Kornakiewicz, A., Koziara, H. & Nauman, P. Deep brain stimulation for refractory epilepsy. Arch. Med. Sci. 8, 805–816 (2012).
Baker, K. B. et al. Cerebellar deep brain stimulation for chronic post-stroke motor rehabilitation: a phase I trial. Nat. Med. 29, 2366–2374 (2023).
Muller, U. J. et al. Deep brain stimulation of the nucleus accumbens for the treatment of addiction. Ann. N. Y. Acad. Sci. 1282, 119–128 (2013).
Viswanathan, A., Jimenez-Shahed, J., Baizabal Carvallo, J. F. & Jankovic, J. Deep brain stimulation for Tourette syndrome: target selection. Stereotact. Funct. Neurosurg. 90, 213–224 (2012).
Novakovic, V. et al. Brain stimulation in posttraumatic stress disorder. Eur. J. Psychotraumatol. 2, https://doi.org/10.3402/ejpt.v2i0.5609 (2011)
Taghva, A. S., Malone, D. A. & Rezai, A. R. Deep brain stimulation for treatment-resistant depression. World Neurosurg. 80, S27 e17–24 (2013).
de Koning, P. P., Figee, M., van den Munckhof, P., Schuurman, P. R. & Denys, D. Current status of deep brain stimulation for obsessive-compulsive disorder: a clinical review of different targets. Curr. Psychiatry Rep. 13, 274–282 (2011).
Apetz, N. et al. Effects of subthalamic deep brain stimulation on striatal metabolic connectivity in a rat hemiParkinsonian model. Dis. Model. Mech. 12, https://doi.org/10.1242/dmm.039065 (2019)
Fischer, D. L. et al. Subthalamic nucleus deep brain stimulation employs trkB signaling for neuroprotection and functional restoration. J. Neurosci. 37, 6786–6796 (2017).
Fischer, D. L. et al. Subthalamic nucleus deep brain stimulation does not modify the functional deficits or axonopathy induced by nigrostriatal alpha-synuclein overexpression. Sci. Rep. 7, 16356 (2017).
Reese, R. et al. High frequency stimulation of the entopeduncular nucleus sets the cortico-basal ganglia network to a new functional state in the dystonic hamster. Neurobiol. Dis. 35, 399–405 (2009).
He, Z. et al. High frequency stimulation of subthalamic nucleus results in behavioral recovery by increasing striatal dopamine release in 6-hydroxydopamine lesioned rat. Behav. Brain Res. 263, 108–114 (2014).
Leblois, A. et al. Deep brain stimulation changes basal ganglia output nuclei firing pattern in the dystonic hamster. Neurobiol. Dis. 38, 288–298 (2010).
Lee, H. J. et al. Change of extracellular glutamate level in striatum during deep brain stimulation of the entopeduncular nucleus in rats. J. Korean Neurosurg. Soc. 62, 166–174 (2019).
Spieles-Engemann, A. L. et al. Stimulation of the rat subthalamic nucleus is neuroprotective following significant nigral dopamine neuron loss. Neurobiol. Dis. 39, 105–115 (2010).
Spieles-Engemann, A. L. et al. Subthalamic nucleus stimulation increases brain derived neurotrophic factor in the nigrostriatal system and primary motor cortex. J. Parkinsons Dis. 1, 123–136 (2011).
Temel, Y. et al. Protection of nigral cell death by bilateral subthalamic nucleus stimulation. Brain Res. 1120, 100–105 (2006).
van der Plasse, G., Schrama, R., van Seters, S. P., Vanderschuren, L. J. & Westenberg, H. G. Deep brain stimulation reveals a dissociation of consummatory and motivated behaviour in the medial and lateral nucleus accumbens shell of the rat. PLoS ONE 7, e33455 (2012).
Visanji, N. P. et al. Deep brain stimulation of the subthalamic nucleus preferentially alters the translational profile of striatopallidal neurons in an animal model of Parkinson’s disease. Front. Cell. Neurosci. 9, 221 (2015).
Paap, M. et al. Deep brain stimulation by optimized stimulators in a phenotypic model of dystonia: effects of different frequencies. Neurobiol. Dis. 147, 105163 (2021).
Schmuckermair, C. et al. Behavioral and neurobiological effects of deep brain stimulation in a mouse model of high anxiety- and depression-like behavior. Neuropsychopharmacology 38, 1234–1244 (2013).
Knorr, S. et al. Experimental deep brain stimulation in rodent models of movement disorders. Exp. Neurol. 348, 113926 (2022).
Fleischer, M., Endres, H., Sendtner, M. & Volkmann, J. Development of a fully implantable stimulator for deep brain stimulation in mice. Front. Neurosci. 14, 726 (2020).
Pinnell, R. C., Pereira de Vasconcelos, A., Cassel, J. C. & Hofmann, U. G. A miniaturized, programmable deep-brain stimulator for group-housing and water maze use. Front. Neurosci. 12, 231 (2018).
Adan, İ., Noor, S., Kiss, Z. & Murari, K. A lightweight discrete biphasic current stimulator for rodent deep brain stimulation. In 2015 IEEE Biomedical Circuits and Systems Conference (BioCAS), Atlanta, GA (2015). https://www.researchgate.net/publication/288496651_A_lightweight_discrete_biphasic_current_stimulator_for_rodent_deep_brain_stimulation.
Adams, S. D., Bennet, K. E., Tye, S. J., Berk, M. & Kouzani, A. Z. Development of a miniature device for emerging deep brain stimulation paradigms. PLoS ONE 14, e0212554 (2019).
Alpaugh, M. et al. A novel wireless brain stimulation device for long-term use in freely moving mice. Sci. Rep. 9, 6444 (2019).
Angotzi, G. N., Boi, F., Zordan, S., Bonfanti, A. & Vato, A. A programmable closed-loop recording and stimulating wireless system for behaving small laboratory animals. Sci. Rep. 4, 5963 (2014).
Badstuebner, K., Gimsa, U., Weber, I., Tuchscherer, A. & Gimsa, J. Deep brain stimulation of hemiParkinsonian rats with unipolar and bipolar electrodes for up to 6 weeks: behavioral testing of freely moving animals. Parkinsons Dis. 2017, 5693589 (2017).
de Haas, R. et al. Wireless implantable micro-stimulation device for high frequency bilateral deep brain stimulation in freely moving mice. J. Neurosci. Methods 209, 113–119 (2012).
Ewing, S. G., Lipski, W. J., Grace, A. A. & Winter, C. An inexpensive, charge-balanced rodent deep brain stimulation device: a step-by-step guide to its procurement and construction. J. Neurosci. Methods 219, 324–330 (2013).
Farahmand, S., Vahedian, H., Abedinkhan Eslami, M. & Sodagar, A. M. Wearable, battery-powered, wireless, programmable 8-channel neural stimulator. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2012, 6120–6123 (2012).
Fluri, F. et al. Development of a head-mounted wireless microstimulator for deep brain stimulation in rats. J. Neurosci. Methods 291, 249–256 (2017).
Forni, C. et al. Portable microstimulator for chronic deep brain stimulation in freely moving rats. J. Neurosci. Methods 209, 50–57 (2012).
Paulat, R., Meissner, W., Morgenstern, R., Kupsch, A. & Harnack, D. Development of an implantable microstimulation system for chronic DBS in rodents. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2011, 660–662 (2011).
Kölbl, F. et al. An embedded deep brain stimulator for biphasic chronic experiments in freely moving rodents. IEEE Trans. Biomed. Circuits Syst. 10, https://doi.org/10.1109/TBCAS.2014.2368788 (2014)
Pol, S., Temel, Y. & Jahanshahi, A. A custom made electrode construct and reliable implantation method that allows for long-term bilateral deep brain stimulation in mice. Neuromodulation 24, 212–219 (2021).
Reitboeck, H. J. Fiber microelectrodes for electrophysiological recordings. J Neurosci. Methods 8, 249–262 (1983).
Mountcastle, V. B., Reitboeck, H. J., Poggio, G. F. & Steinmetz, M. A. Adaptation of the Reitboeck method of multiple microelectrode recording to the neocortex of the waking monkey. J. Neurosci. Methods 36, 77–84 (1991).
Cogan, S. F. et al. Sputtered iridium oxide films for neural stimulation electrodes. J. Biomed. Mater. Res. B Appl. Biomater. 89, 353–361 (2009).
Ip, C. W. et al. AAV1/2-induced overexpression of A53T-alpha-synuclein in the substantia nigra results in degeneration of the nigrostriatal system with Lewy-like pathology and motor impairment: a new mouse model for Parkinson’s disease. Acta Neuropathol. Commun. 5, 11 (2017).
Karikari, A. A. et al. Neurodegeneration by alpha-synuclein-specific T cells in AAV-A53T-alpha-synuclein Parkinson’s disease mice. Brain Behav. Immun. 101, 194–210 (2022).
Grotemeyer, A. et al. Inflammasome inhibition protects dopaminergic neurons from alpha-synuclein pathology in a model of progressive Parkinson’s disease. J. Neuroinflamm. 20, 79 (2023).
Dragunow, M. & Faull, R. The use of c-fos as a metabolic marker in neuronal pathway tracing. J. Neurosci. Methods 29, 261–265 (1989).
Musacchio, T. et al. Subthalamic nucleus deep brain stimulation is neuroprotective in the A53T alpha-synuclein Parkinson’s disease rat model. Ann. Neurol. 81, 825–836 (2017).
Hausdorff, J. M. Gait dynamics in Parkinson’s disease: common and distinct behavior among stride length, gait variability, and fractal-like scaling. Chaos 19, 026113 (2009).
Mathis, A. et al. DeepLabCut: markerless pose estimation of user-defined body parts with deep learning. Nat. Neurosci. 21, 1281–1289 (2018).
Lessner, G. et al. Differential proteome of the striatum from hemiParkinsonian rats displays vivid structural remodeling processes. J. Proteome Res. 9, 4671–4687 (2010).
Simola, N., Serra, M., Marongiu, J., Costa, G. & Morelli, M. Increased emissions of 50-kHz ultrasonic vocalizations in hemiParkinsonian rats repeatedly treated with dopaminomimetic drugs: a potential preclinical model for studying the affective properties of dopamine replacement therapy in Parkinson’s disease. Prog. Neuropsychopharmacol. Biol. Psychiatry 108, 110184 (2021).
Koprich, J. B., Johnston, T. H., Reyes, M. G., Sun, X. & Brotchie, J. M. Expression of human A53T alpha-synuclein in the rat substantia nigra using a novel AAV1/2 vector produces a rapidly evolving pathology with protein aggregation, dystrophic neurite architecture and nigrostriatal degeneration with potential to model the pathology of Parkinson’s disease. Mol. Neurodegener. 5, 43 (2010).
Paxinos, G. & Franklin, K. B. J. The Mouse Brain in Stereotaxic Coordinates (Academic Press, 2013).
Acknowledgements
This project was funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation) Project-ID 424778381-TRR 295 (A01 to J.V., A02 to R.B. J.V., A05 to M.S., A06 to C.W.I., Gerok position to A.G.). C.W.I. receives funding from the IZKF at the University of Würzburg (S-506, N-362) and from the VERUM foundation.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
A.G. methodology, validation, formal analysis, investigation, data curation, writing—original draft, writing—review and editing, supervision, project administration, funding acquisition. T.P. methodology, validation, formal analysis, investigation, data curation, writing—original draft, writing—review and editing. R.P. methodology, software, validation, formal analysis, data curation. D.H. methodology, validation, formal analysis, investigation, data curation, writing—original draft. T.K. methodology, writing—review and editing. U.T. methodology, validation, writing - review and editing. H.E. stimulator circuit and programming—review and editing. R.B. methodology, writing—review and editing. M.S. methodology, writing—review and editing. J.V. Resources, writing—review and editing C.W.I. conceptualization, methodology, validation, resources, writing—original draft, review and editing, supervision, project administration, funding acquisition.
Corresponding author
Ethics declarations
Competing interests
A.G., T.P., R.P., H.E., R.B., M.S., J.V., and C.W.I. declare that there are no conflicts of interest. D.H., T.K., and U.T. are directly related to Thomas Recording GmbH. The implantable microstimulation system used in this article was manufactured by Thomas RECORDING GmbH, Germany, as a prototype for testing. After the here reported tests had been finished, the product was commercialized by the private industry company Thomas RECORDING GmbH. The author Uwe Thomas (CEO) is owner-manager; the author Dirk Hoehl is technical director and the author Torsten Knauer is head of the electrode department of Thomas RECORDING GmbH. They all may financially benefit from this circumstance if the company is successful in marketing its products that are related to this research.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Grotemeyer, A., Petschner, T., Peach, R. et al. Standardized wireless deep brain stimulation system for mice. npj Parkinsons Dis. 10, 153 (2024). https://doi.org/10.1038/s41531-024-00767-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41531-024-00767-2