Abstract
The optical detection of nanoparticles, including viruses and bacteria, underpins many of the biological, physical and engineering sciences. However, due to their low inherent scattering, detection of these particles remains challenging, requiring complex instrumentation involving extensive sample preparation methods, especially when sensing is performed in liquid media. Here we present an easy-to-use, high-throughput, label-free and cost-effective method for detecting nanoparticles in low volumes of liquids (25 nL) on a disposable chip, using an acoustically actuated lens-free holographic system. By creating an ultrasonic standing wave in the liquid sample, placed on a low-cost glass chip, we cause deformations in a thin liquid layer (850 nm) containing the target nanoparticles (≥140 nm), resulting in the creation of localized lens-like liquid menisci. We also show that the same acoustic waves, used to create the nanolenses, can mitigate against non-specific, adventitious nanoparticle binding, without the need for complex surface chemistries acting as blocking agents.
Similar content being viewed by others
Introduction
The detection of nanoparticles, including the biosensing of viruses and bacteria, remains an important field in the biomedical, physical and engineering sciences, with applications as diverse as medical microbiology, cosmetics, paints, drug delivery systems and as probes for DNA sensing. Their detection, however, remains a challenge, especially during in situ liquid measurements, which often require the use of sophisticated, costly and bulky equipment coupled with advanced sample preparation methods1,2. Some of the most commonly used techniques for sensing such small particles are based around imaging systems including e.g., optical nanoscopy3, atomic force microscopy4,5, and resistive pulse sensing6. High resolution optical microscopy methods7,8,9 have also been developed for this purpose, including dynamic light scattering, nanoparticle tracking analysis10, as well as super-resolution11,12, near-field13,14 and dark-field15,16 microscopies. In many cases, these either require labeling with markers (e.g., by using fluorescent molecules, Raman active centers, or optically scattering particles), or are limited by the need for samples to be in a solid state (i.e., dry), adding additional preparation steps. As nanoparticles are not only synthesized but are also utilized in liquids17, their detection and imaging in solvents and aqueous media is increasingly important.
Many optical detection techniques for nanoparticles, including those listed above, are often limited in their resolution due to optical diffraction and typically only have a small field-of-view, FOV, (often <0.1 mm2) making measurements both time consuming and limiting, in terms of the sample volume. This in turn constrains both their sensitivity and throughput, which are key analytical parameters for many applications. Recently a microsphere-enhanced optical imaging system has been used to enable <25 nm resolution albeit with a small FOV of 125 µm × 125 μm18.
Lens-free sensor systems used in computational holographic imaging have overcome some of the limitations related to nanoparticle detection and characterization19,20, with images being computationally reconstructed from in-line holograms, resulting from the interference between the directly transmitted wave and the scattered optical wave from the sample. The resolution and detection sensitivity of the lens-free holographic sensing, however, still depend on the scattering properties of the target nanoparticles, the signal to noise ratio and the pixel size of the detector-array (where complementary metal–oxide–semiconductor, CMOS, image sensors are often chosen for their low-cost and ease-of-use)21. Although, the lens-free holographic imaging systems can be used for the reconstruction of 2D and 3D images, their resolution is fundamentally limited by the diffraction of light. Recent improvements in holographic imaging, including the use of pixel super-resolution techniques22,23 have already enabled enhanced resolution by digitally dividing individual pixels of an image sensor into much smaller virtual pixels. In other innovations, liquid polymers that self-assemble around nanoparticles have been used to create nanolenses around the nanoparticles, to mitigate against the low signal-to-noise ratio19,24, and have enabled the detection and sizing of sub-50 nm particles22.
In all these cases reported to date, where holography has been used for nanoparticle detection, the sample has been imaged and characterized in a dry state (i.e., not in a bulk liquid film), thereby constraining the throughput and range of applications. Examples of such techniques have including either draining the solvent from the chip19, using vapor condensation24, or using pyroelectric effects at high temperatures (~100 °C)25,26. Alternatively, where imaging has been achieved in a bulk film, this has involved using electrohydrodynamic methods to shape the fluids and create lensing structures. However, these methods are generally slow and require relatively high voltages at previously designed planar electrode arrays, producing lensing systems that generally have a small FOV27,28,29.
As an alternative, here we present a low power, high-throughput, and a cost-effective method for detecting nanoparticles including bioparticles in a small liquid sample volume (~25 nL), using an acoustically actuated, lens-free holographic system, see Fig. 1. The technique enables the creation of on-demand nanolenses in liquids using carefully defined geometric and operational conditions to displace the liquid across the full FOV ~ 30 mm2. The acoustic pressure nodes on the glass chip form convex menisci around the nanoparticles and enhance their optical signal response, Fig. 2. We optimized the acoustic frequency and power, showing both the dispersion analysis of the guided acoustic waves in the disposable chip, as well as the power density coupling as a function of fluid height. We also analyzed the dimensions of the nanolenses as a function of nanoparticle size. By using disposable chips with immobilized antibodies, we are able to obtain the specificity required for biological sample identification. Finally, by exploiting the differential motion of the particles on the surface of the chip during acoustic actuation, which creates local streaming, we distinguish between nonspecifically bound particles (which are released from the surface upon acoustic actuation) from those which are specifically bound by antibodies (and remain fixed). The platform provides a highly adaptable, low-cost, easy-to-use, and field-portable system that has the potential to be readily integrated into different types of biosensing strategies, enabling, e.g., the high-throughput detection of viruses and bacterial cells in solutions.
Results and discussion
Detection system
The proposed detection method integrates both acoustical and optical functionalities onto the same platform, built around a reusable lithium niobate substrate, interfaced with a disposable glass chip. The acoustical system involves the excitation of Rayleigh waves, by an interdigitated transducer (IDT) patterned onto the reusable lithium niobate piezoelectric wafer, which interacts with the thin disposable chip (~145 µm) to generate dispersive Lamb-type guided waves which deform the liquid layer to enhance optical signatures from nanoparticles, Fig. 2. The generated guided waves in the disposable chip were analyzed analytically in the plate geometries, with or without liquid loading, enabling us to control wave propagation under different waveguiding conditions.
Numerical analysis
In detail, we consider how the guided wave propagates in a nonviscous liquid–solid bilayer with wavenumber k = 2πf/v, where f is the frequency and v is the guided wave phase velocity. We ignored the effect of viscosity on the phase velocity because it is negligible compared to the effect on attenuation30. The liquid layer has a thickness of 2a that can be characterized using the ultrasonic wave phase velocity in the liquid cF, and the volumetric mass density ρF. The elastic solid layer that enables propagation of ultrasonic longitudinal and transversal waves with the velocities cL and cT, respectively has a thickness of 2h and volumetric mass density ρ. The small thickness of the disposable glass chip (~145 µm, with respect to the wavelength) results in mode transformation of a surface acoustic wave (SAW) to a Lamb-type wave at 9.71 MHz as shown in the frequency–phase velocity plot in Fig. 3.
The propagation of guided waves in the liquid–solid bilayer was subsequently analyzed using the dispersion equation31
where
Eqs. (2) and (3) represent the dispersion relations for the antisymmetric and symmetric guided modes (Lamb-type wave) in the free solid plate. The remaining coefficients are given by
where \({\mathrm{\Omega }} = \frac{{4fh}}{{{\mathbf{c}}_{\mathbf{T}}}},\xi = \frac{{4fh}}{{\mathbf{v}}}.\)
An impedance mismatch causes the Lamb-type wave to reflect back from the glass–air boundary, hence creating a standing wave. Numerical results show that for a ~1 µm thick layer of liquid (with properties detailed in Supplementary Table 1) on top of the glass chip, the thin plate can support only zeroth-order (fundamental) Lamb-type wave modes A0 and S0 at the excitation frequency of 9.71 MHz (the cut-off frequency of the higher order mode is 11.70 MHz). As the excitation frequency was increased, higher-order modes could be excited and the fundamental modes converged to a surface wave at a high frequency limit of 50 MHz, Fig. 3.
Similarly, as thickness of the liquid loading was increased, the Lamb wave started to couple into the liquid and more energy leaked into the sample, as shown in the power density flow curves, Fig. 4. The direction of power flow indicates the wave propagation direction across the thickness of the material, and the length indicates the magnitude. For a thin liquid film, e.g., ~1 µm, when 2a < 2h, a small leakage of the wave energy in the liquid was possible. In contrast, as the thickness was increased, e.g., to 200 µm, such that 2a > 2h, the extent of wave energy leakage into the liquid increased, resulting in acoustic streaming in the liquid, Fig. 4. Thus, by ensuring that the sample thickness satisfied the condition 2a ≪ 2h, we could also ensure that the Lamb wave has minimal leakage into the liquid, limiting the distortions of the liquid layer on top of the glass to those induced mechanically by the deformations of the plate (the liquid–solid bilayer), rather than those caused by convection, during acoustic streaming.
Nanoparticle detection with our in-line holographic system will depend in part upon the nature of the nanoparticle-nanolens complex and its scattering properties, as well as on the pixel size and the signal to noise ratio of the image sensor-array used. The light scattering intensity from nanoparticles that are significantly smaller than the wavelength of light used, depends on the sixth power of the size of the particle, and is governed by Rayleigh scattering accordingly32
where Io and λ are the incident intensity and the wavelength of the light source, respectively, r is the distance of the observer, Rp is the radius of the particle, n is the refractive index of the particle relative to the background, and θ is the scattering angle.
Thus, nanoparticles are difficult to detect due to their low optical scattering cross-section. In common with other coherent interferometric systems, for holographic sensors, the field signal scales as a third power of the size, thus providing a stronger signal for nanoparticle detection33. The radial coordinate of the meniscus of the nanolens (rm), as a function of the elevation above the substrate was determined by solving the Young-Laplace equation19
where
and θs and θp are the contact angles of the substrate and a spherical particle respectively, z0 is the elevation of the meniscus–particle contact line.
These equations show that the size of the nanolens meniscus depends upon the contact angle which in turn depends upon the properties of the liquid, the particles and the chemical nature of the surface of the chip. Given the very small length scales involved, it is difficult to measure the contact angles in our system34. Although, the nanoparticle itself is small, the nanoparticle–nanolens complex itself is relatively large. We therefore use approximations from previously reported simulations of nanolenses19 of similar length scales.
Particle detection
Based upon this understanding, model experiments were performed using spherical commercially sourced polystyrene nanoparticles (with diameters between 140 nm and 1 μm) with well-controlled physical and chemical properties such as the refractive index, density, and shape. The liquid sample was back-illuminated using a partially coherent light source and the in-line holograms, resulting from the interference between directly transmitted wave and the scattered object wave, were recorded using a low-cost CMOS image sensor. The sample was placed close to the image sensor at a distance of ~1 mm.
Figure 5 shows the detection of 200 nm biotinylated polystyrene nanoparticles bound onto streptavidin, itself immobilized onto the glass surface20. No nanoparticles could be detected prior to the acoustic actuation, Fig. 5a. Upon generation of the SAW, the nanoparticle–nanolens complexes were formed at the nodes of the waves, Fig. 2, and nanoparticles were detected, Fig. 5b. Figure 5b inset also shows the SEM images (as a gold standard reference imaging method) of the 200 nm particles. When the SAW excitation was switched off, the Lamb wave on the disposable chip dissipated immediately and the resulting relaxation of the liquid film resulted in the particles becoming invisible.
Similarly, Fig. 6 shows the range of nanoparticle sizes that we were able to detect between 140 nm and 1 μm, with acoustic actuation of the optofluidic lensing switched on and off. We were unable to detect nanoparticles that were smaller than 140 nm, establishing our limit of detection for the holographic sensor system at an illumination wavelength of 532 nm. In principle, tuning of the illumination wavelength makes it possible to increase the scattering cross-section of nanoparticles, although in our case, the signal to noise ratio improvement remained limited.
Our numerical calculations (Fig. 7a) show that for a 140 nm diameter spherical particle, the radius of the nanoparticle–nanolens complex Rm can range between 0.345 and 0.845 µm. To understand the capability of the detection system to resolve closely spaced objects for a transmission based optical holographic systems, however, we have to consider both the size of nanoparticle–nanolens complex (Rm) and the pixel size of the CMOS sensor, which is 1.67 µm in our case. Although it has previously proved possible to circumvent any limit imposed by the sensor’s pixel size (e.g., by using pixel super-resolution techniques)27, in this detection system we use only simple in-line holographic detection. Thus, if the inter-particle spacing is smaller than the size of the individual nanolens complex, then proximal particles are likely to merge into a single detection point. At our detection limit, observing 140 nm diameter particles, the total particle–lens complex diameter ranges from 0.690 to 1.690 µm (depending upon the contact angles), which is on the order of the pixel size of the CMOS imager in our setup.
A further key analytical parameter in the analysis of our detection system is an understanding of its throughput. For conventional far-field microscopy using a high magnification objective, particles of comparable size (~200 nm) can be readily detected over an FOV of 0.5–1 mm2. Our holographic detection system, however, has an image sensor area of 30 mm2, offering a potential improvement in throughput by a factor of ~50-fold. Further, and unlike conventional far-field optical detection, we can improve our FOV and hence our throughput by using even larger sensors (without any reduction in our ability to detect the particles).
The combination of size of the nano-lenses and the pixel resolution also governs the optimal density per unit area of the disposable chip, a process that can be controlled by the nature and density of the ligand-binding moieties, which in this study included both antibodies and streptavidin (together with cross linkers to ensure that an optimal distance between the particles is maintained). Following optimization of these parameters, the sensor was able to detect bioparticles in thin liquid films, including herpes simplex virus (HSV-I) (~200 nm)20 and Staphylococcus aureus (S. aureus) bacteria (800–1000 nm)35,36, Fig. 8a, b, respectively. Detecting such bioparticles in liquids is a challenging task due to the small optical scattering cross-section and refractive index difference with the surrounding liquid19. In both cases the bioparticles were captured on the disposable chip, using appropriate antibodies immobilized with silane chemistries, see Experimental Section. SEM images of the pathogens are given in the insets of Fig. 8.
The detection of these pathogens was chosen due to their importance in the field of microbiology and infectious disease. HSV is one of the most widespread sexually transmitted diseases around the globe, which has been estimated to have affected more than 50% of the adult population in the US alone37. Similarly, S. aureus is a highly infectious pathogen that causes skin infections, respiratory infections, and food poisoning.
Removal of nonspecific binding
When relying upon any ligand binding strategy within an analytical systems to improve the specificity of detection (e.g., in our case by exploiting antibody–antigen binding or streptavidin–biotin avidity) nonspecific binding events, including, the adventitious binding of beads, viruses or nanoparticles on the chip surface, can lead to false positives. Previously established methods to minimize nonspecific binding can include coating of chips with albumin, polyethylene glycol (PEG), or other blocking polymers and macromolecules38, all of which adds to the complexity of chip manufacture.
It is already known that nonspecifically bound particles39 and proteins40 can be removed using acoustic actuation with surface waves. Building upon this knowledge, we chose to integrate the acoustic actuation of the nanolens with a strategy to mitigate against nonspecific binding. Figure 9a shows superimposed fluorescent images of specifically bound fluorescently labeled 1000 nm polystyrene nanoparticles (which do not move under acoustic actuation) whilst nonspecifically bound particles showed significant movement.
The mechanism of this controlled movement will depend upon the size of the particle. For smaller particles (<500 nm), actuation will arise due to acoustic radiation pressure, whilst for larger particles (>500 nm) movement is most likely to be due to acoustically induced flows near the liquid–solid boundary layer41, Fig. 9b. We developed a simple algorithm based upon the superimposition of three images, that enabled us to readily identify specifically bound and nonspecifically bound particles, by establishing an absolute threshold for the distance moved (a parameter that we determined experimentally based upon the approximate size of the target particles, liquid’s viscosity, and acoustic power).
Our overall ability to detect particles is fundamentally determined by the effective signal to noise ratio of the imaging system, the properties of the nanoparticle and nanolens, including e.g., the size of the particle and the lens complex, the pixel size, the refractive properties of the nanoparticle and surrounding medium (as the optical contrast), the height of the liquid film, and the nature of the acoustic waves. The system itself is complex, and also includes a series of related factors, e.g., surface properties of the particle and the chip, nature of the supporting media, the residual liquid height at the nodes and the antinodes, the local temperature, the power of the SAW at that particular location and the position of the nanoparticles relative to the node and antinode.
In the present configuration, in Fig. 6, we see that nanoparticles are detected both at pressure nodes and at the antinodes, and we believe that only a very low number of false-negatives exist. We note that, by varying the excitation frequency, ultrasonic standing waves of different wavelengths can be created which spatially shift the position of nodes and antinodes in the liquid. Capturing images of the sample at different acoustic frequencies and superimposing them would uncover any blind spots. An alternative approach, to improve the coverage and mitigate against any blind spots, would be to configure two IDTs placed orthogonal to each other, to create two dimensional nanolens array thereby increasing the sensing area and throughput of the system.
We also note that the technique is sensitive both to the shape and size of the target nanoparticles. The technique is therefore able to differentiate between clusters of particles and the same particles as individual objects, but it is not able to differentiate between small particles aggregated in clusters as this will be detected as a single (larger) particle, when using the raw hologram for rapid detection. In any mixed sample of different nanoparticles, particularly when they are biological, we propose that the strategy for detection be best performed with an immobilized ligand, such as an antibody as demonstrated in Figs. 8 and 9.
In conclusion, we presented an acousto-optic sensing method that enables the detection of nanoparticles in liquid samples using a low-cost, mass producible, wide FOV, high-throughput, and portable lens-free optical microscope coupled with ultrasonic actuation on a disposable chip. We demonstrated the creation of on-demand dynamic nanolenses using an active substrate without moving parts and thus favouring the miniaturization of the optical systems. We showed that a thin meniscus forms around the nanoparticles, which acts as a nanolens, by generating standing waves in the liquid. As a practical application, the detection of the HSV-I virus (200 nm diameter) as well as S. aureus bacterial cells (800–1000 nm diameter) in liquid was also demonstrated.
In principle, the technique described requires only simple sample preparation. However, to enable detection outside of the laboratory environment, as used in this study, where the sample composition (including the opaque nature of many matrices), the stage in the life cycle of microbes, and-or the state of bioparticle aggregation (due to matrix effects), may all represent challenges that could affect the analytical metrics, particularly around sensitivity, more detailed optimization will be necessary. In order to mitigate against these, we demonstrated that the technique can be combined with controlled antibody immobilization methods to enable specifically bound particles to be detected, whilst nonspecifically bound particles can be removed through acoustic streaming39,40.
Methods
Materials
Biotin and m-PEG-silane (PJK-1915 and PLS-2013) was purchased from Creative PEGWorks (NC). The fluorescently labeled nanoparticles (F-8767, F8768, and F13081), the biotin tagged HSV type 1/2 polyclonal antibody (PA1-7488), biotin tagged S. aureus polyclonal antibody (PA1-73174), PEG-300 and PEG-400 were acquired from Thermo Fischer Scientific. The viral culture (NATtrol Herpes Simplex Virus Type 1 Strain: MacIntyre, 50,000 copies/ml) was purchased from ZeptoMetrix Corporation. The S. aureus samples were bought from ATCC (ATCC 27660). The washing buffer comprised 50 mM phosphate buffer, pH 7.4. Unless otherwise stated, reagents were purchased from Sigma-Aldrich. All the materials were used as received.
SAW device fabrication and characterization
Single crystal lithium niobate piezoelectric wafers were used to generate SAW but were not required for their propagation. In this case, the Rayleigh wave, excited on a 128° Y-cut X-propagating lithium niobate wafer with thickness of 1 mm (Roditi, UK), was coupled into a disposable glass chip (a thin microscope coverslip) as a Lamb wave. This excitation of SAW was achieved by means of an IDT with a resonant frequency of 9.71 MHz (the resonant frequency is dependent on the width of the IDT electrode, was one-fourth of device pitch, calculated using a simple frequency–wavelength relationship, f = c/λ, where, f is the frequency, and c is the sound wave propagation velocity in lithium niobate). For a 9.71 MHz IDT, D was 411 µm. The IDT aperture was 20 mm, and 40 titanium–gold finger pairs were fabricated on the wafer. The IDT was manufactured by standard photolithographic techniques, as has been reported previously42. To characterize the frequency response, i.e., the resonant frequency and the harmonics, a network analyzer (E507C ENA series, Agilent Technologies) was used.
Lens-free detection
A schematic of the lens-free holographic sensor is shown in Fig. 1. The system was constructed from 3D-printed parts using a Dimension Elite 3D printer (Stratasys). The illumination chamber consisted of 24 fiber-coupled LEDs, centered around 532 nm, controlled by a microcontroller. The light from the fibers passed through a band pass filter with a center wavelength of 532 nm (10 nm bandwidth) before illuminating the sample. The sample was placed at a distance of ~15 cm from the light source and ~1 mm from the CMOS chip (10-megapixel CMOS image sensor (UI-1492LE-M, Imaging Development Systems) with USB readout, with an active sensing area of ~5 mm × 6 mm). It was noted that for detection of particles with PEG-based samples, the optical contrast between the polystyrene bead (n = 1.59) and the surrounding medium (n = 1.47), was small, underlining the analytical challenge.
The glass chip was placed on a lithium niobate piezoelectric wafer which had a photolithographically patterned IDT generating a 9.71 MHz acoustic wave, as described above. A 20 μL drop of PEG-400 was used to couple the ultrasonic signal between the piezoelectric wafer and the glass chip. The IDT was connected to a 50 MHz function generator (TG5011, TTi) in conjunction with an amplifier (ZHL-5W-1, Mini Circuits) and a 3A, ±24 V dc power supply. The particle detection systems and the illumination module were controlled with a software application described elsewhere27.
Sample preparation on disposable chips
The chips were prepared according to previously established protocols20. A standard microscopy glass coverslip served as the low-cost disposable chip, which was first washed in acetone and then thoroughly rinsed, with isopropanol and deionized water. The chips were then plasma-treated for 5 min to make their surface hydrophilic, and able to wet when either reagents for immobilization or the liquid sample were added.
In those experiments where a surface functionalization was performed, this involved first adding a silane–PEG–biotin solution [1 mg/ml in a solution containing 95% ethanol (1.875 ml) and acetic acid (125 µl)] to the disposable chip (glass). After 1 hour of incubation, the surface was again washed with ethanol and water to remove unbound PEG. We then added a solution of streptavidin (100 µg/ml in water) and again incubated it for 1 hour to allow it to bind effectively to biotin–PEG. Unattached streptavidin was washed away with washing buffer. The substrate was either used immediately or stored at −20 °C for future use.
In a series of experiments exploring potential applications, we chose fluorescent biotinylated native (unmodified) nanoparticles (with diameters of 140–1000 nm, (F-8767, F8768, and F13081)), which we used as model systems to characterize the detection process. HSV-I virus particles bound to a biotin-tagged antibody (PA1-7488)) or S. aureus bacteria bound to a biotin-tagged antibody (PA1-73174) were also used to demonstrate the detection of important microbes. In both cases particles were incubated on the substrate for an hour. The samples were subsequently washed gently to remove unattached particles and then covered with a droplet of washing buffer.
The antibodies were bound to the virus particles or bacterial cells by mixing the biotin-tagged antibody (volume ratio 1:10) with the respective pathogen and by incubating the mixture for 2 hours; the unbound antibodies were washed away by centrifugal filtration. Here, we used 50 µL solution of bacteria/virus on each glass coverslip.
Immediately prior to imaging, 50 µL of PEG-400 was added onto the coverslip and spin-coated at 12,000 rpm enabling formation of a thin liquid film. The height of the liquid film (2a), was characterized using ellipsometry. Chips (18 × 18 mm) were plasma treated in an oxygen asher (PlasmaFab 505, Electroetch) and coated with PEG-400 at a range of spin-speeds, using a spinner (PWM32, Headway Research). An ellipsometer (M-2000, J.A. Woollam) was used to scan the sample at an incidence angle of 75° with polarized light. The data were analyzed using CompleteEASE software with a Cauchy model.
Acoustic actuation of particles and their subsequent movement involved fluorescence imaging (×20 Objective size, NA = 0.4, with an Olympus, BX51 microscope). Both videos and images were acquired using a charge-coupled device camera (QIMAGING, Retiga 2000R) with a 20× objective lens (λex = 488 nm, λem = 532 nm). The images were analyzed in ImageJ43 (v1.5) using built-in image calculator function where the time-series images were superimposed onto each other. This simple method enabled us to readily identify specifically bound and nonspecifically bound particles, by establishing thresholds for the distance moved (a parameter determined experimentally based on the approximate size of the particles, liquid’s viscosity and acoustic power).
Electron microscopy
Electron microscopy was used for validation and calibration purposes. The samples were imaged using an SEM (Nova 600 SEM/FIB System). Prior to holographic detection, we placed identifiable calibration marks on the glass chip using marker pens, which are easily recognizable in the lens-free images as well as in SEM images. Easily identifiable (large) objects, located close to the region of interest, were also used as secondary reference points. The horizontal and vertical distances of the particles of interest from such features were calculated using the lens-free image, to generate a map20. While imaging, the samples were aligned such that the angular orientation of each slide matched the lens-free image. The particles of interest were identified by locating the primary and secondary markers using the generated map. This step-by-step particle searching strategy enabled us to register our images accurately, which was used for comparison purposes.
Numerical analysis
The dispersion curves were numerically calculated in MATLAB. The plate geometry was chosen such that x3 < 0 corresponded to a glass plate of thickness 2 h, and x3 > 0 corresponded to a liquid layer of thickness 2a. For layer x3 > 0, the liquids have a thickness (2a).
Data availability
The data including numerical simulation codes of this study are available at https://doi.org/10.5525/gla.researchdata.908.
References
Martin, C. R. Nanomaterials: a membrane-based synthetic approach. Science 266, 1961–1966 (1994).
Betzig, E., Trautman, J. K., Harris, T. D., Weiner, J. S. & Kostelak, R. L. Breaking the diffraction barrier: optical microscopy on a nanometric scale. Science 251, 1468–1470 (1991).
Li, L., Guo, W., Yan, Y., Lee, S. & Wang, T. Label-free super-resolution imaging of adenoviruses by submerged microsphere optical nanoscopy. Light Sci. Appl. 2, e104–e104 (2013).
Lee, J. S. et al. Multifunctional hydrogel nano-probes for atomic force microscopy. Nat. Commun. 7, 1–14 (2016).
Ebenstein, Y., Nahum, E. & Banin, U. Tapping mode atomic force microscopy for nanoparticle sizing: tip–sample interaction effects. Nano Lett. 2, 945–950 (2002).
Song, Y., Zhang, H., Chon, C. H., Pan, X. & Li, D. Nanoparticle detection by microfluidic resistive pulse sensor with a submicron sensing gate and dual detecting channels-two stage differential amplifier. Sens. Actuators B Chem. 155, 930–936 (2011).
Pol, Evander et al. Particle size distribution of exosomes and microvesicles determined by transmission electron microscopy, flow cytometry, nanoparticle tracking analysis, and resistive pulse sensing. J. Thrombosis Haemost. 12, 1182–1192 (2014).
Pol, E., van der, Coumans, F. A. W., Sturk, A., Nieuwland, R. & van Leeuwen, T. G. Refractive index determination of nanoparticles in suspension using nanoparticle tracking analysis. Nano Lett. 14, 6195–6201 (2014).
Hendrickson, O. D., Safenkova, I. V., Zherdev, A. V., Dzantiev, B. B. & Popov, V. O. Methods of detection and identification of manufactured nanoparticles. Biophysics 56, 961–986 (2011).
Cho, E. J. et al. Nanoparticle characterization: state of the art, challenges, and emerging technologies. Mol. Pharmaceutics 10, 2093–2110 (2013).
Huang, B., Babcock, H. & Zhuang, X. Breaking the diffraction barrier: super-resolution imaging of cells. Cell 143, 1047–1058 (2010).
Cognet, L., Leduc, C. & Lounis, B. Advances in live-cell single-particle tracking and dynamic super-resolution imaging. Curr. Opin. Chem. Biol. 20, 78–85 (2014).
Betzig, E. & Chichester, R. J. Single molecules observed by near-field scanning optical microscopy. Science 262, 1422–1425 (1993).
Kalkbrenner, T., Ramstein, M., Mlynek, J. & Sandoghdar, V. A single gold particle as a probe for apertureless scanning near-field optical microscopy. J. Microsc. 202, 72–76 (2001).
Liu, X. et al. Single gold nanoparticle-based colorimetric detection of picomolar mercury ion with dark-field microscopy. Anal. Chem. 88, 2119–2124 (2016).
Ray, A., Kopelman, R., Chon, B., Briggman, K. & Hwang, J. Scattering based hyperspectral imaging of plasmonic nanoplate clusters towards biomedical applications. J. Biophotonics 9, 721–729 (2016).
Park, J. et al. 3D structure of individual nanocrystals in solution by electron microscopy. Science 349, 290–295 (2015).
Chen, L., Zhou, Y., Li, Y. & Hong, M. Microsphere enhanced optical imaging and patterning: from physics to applications. Appl. Phys. Rev. 6, 021304 (2019).
Mudanyali, O. et al. Wide-field optical detection of nanoparticles using on-chip microscopy and self-assembled nanolenses. Nat. Photonics 7, 247–254 (2013).
Ray, A. et al. Computational sensing of herpes simplex virus using a cost-effective on-chip microscope. Sci. Rep. 7, 1–9 (2017).
Bigas, M., Cabruja, E., Forest, J. & Salvi, J. Review of CMOS image sensors. Microelectron. J. 37, 433–451 (2006).
Sobieranski, A. C. et al. Portable lensless wide-field microscopy imaging platform based on digital inline holography and multi-frame pixel super-resolution. Light Sci. Appl. 4, e346–e346 (2015).
Ozcan, A. & McLeod, E. Lensless imaging and sensing. Annu. Rev. Biomed. Eng. 18, 77–102 (2016).
McLeod, E. et al. Tunable vapor-condensed nanolenses. ACS Nano 8, 7340–7349 (2014).
Ferraro, P., Grilli, S., Miccio, L. & Vespini, V. Wettability patterning of lithium niobate substrate by modulating pyroelectric effect to form microarray of sessile droplets. Appl. Phys. Lett. 92, 213107 (2008).
Grilli, S., Vespini, V. & Ferraro, P. Surface-charge lithography for direct PDMS micro-patterning. Langmuir 24, 13262–13265 (2008).
Miccio, L. et al. Tunable liquid microlens arrays in electrode-less configuration and their accurate characterization by interference microscopy. Opt. Express 17, 2487–2499 (2009).
Grilli, S. et al. Liquid micro-lens array activated by selective electrowetting on lithium niobate substrates. Opt. Express 16, 8084–8093 (2008).
Schuhladen, S. et al. Variable optofluidic slit aperture. Light 5, e16005 (2016).
Nayfeh, A. H. & Nagy, P. B. Excess attenuation of leaky Lamb waves due to viscous fluid loading. J. Acoustical Soc. Am. 101, 2649–2658 (1997).
Yapura, C. L. & Kinra, V. K. Guided waves in a fluid-solid bilayer. Wave Motion 21, 35–46 (1995).
Rayleigh, Lord XXXIV. On the transmission of light through an atmosphere containing small particles in suspension, and on the origin of the blue of the sky. Lond. Edinb. Dublin Philos. Mag. J. Sci. 47, 375–384 (1899).
McLeod, E. & Ozcan, A. Unconventional methods of imaging: computational microscopy and compact implementations. Rep. Prog. Phys. 79, 076001 (2016).
Giro, R., Bryant, P. W., Engel, M., Neumann, R. F. & Steiner, M. B. Adsorption energy as a metric for wettability at the nanoscale. Sci. Rep. 7, 46317 (2017).
Monteiro, J. M. et al. Cell shape dynamics during the staphylococcal cell cycle. Nat. Commun. 6, 1–12 (2015).
Markey, B. K. Clinical Veterinary Microbiology (Elsevier, 2013).
Xu, F., Markowitz, L. E., Gottlieb, S. L. & Berman, S. M. Seroprevalence of herpes simplex virus types 1 and 2 in pregnant women in the United States. Am. J. Obstet. Gynecol. 196, 43.e1–43.e6 (2007).
Zhang, X., Brodus, D., Hollimon, V. & Hu, H. A brief review of recent developments in the designs that prevent bio-fouling on silicon and silicon-based materials. Chem. Cent. J. 11, 18 (2017).
Bourquin, Y., Reboud, J., Wilson, R., Zhang, Y. & Cooper, J. M. Integrated immunoassay using tuneable surface acoustic waves and lensfree detection. Lab Chip 11, 2725 (2011).
Cular, S., Branch, D. W., Bhethanabotla, V. R., Meyer, G. D. & Craighead, H. G. Removal of nonspecifically bound proteins on microarrays using surface acoustic waves. IEEE Sens. J. 8, 314–320 (2008).
Nyborg, W. L. M. 11—Acoustic Streaming. in Physical Acoustics (ed. Mason, W. P.) Vol. 2, 265–331 (Academic Press, 1965).
Reboud, J. et al. Shaping acoustic fields as a toolset for microfluidic manipulations in diagnostic technologies. Proc. Natl Acad. Sci. 109, 15162–15167 (2012).
Schindelin, J., Rueden, C. T., Hiner, M. C. & Eliceiri, K. W. The ImageJ ecosystem: an open platform for biomedical image analysis. Mol. Reprod. Dev. 82, 518–529 (2015).
Acknowledgements
The authors acknowledge Drs. R. Wilson, M. Tassieri and A. Glidle (University of Glasgow) for their scientific guidance, Dr. H. Koydemir (University of California, Los Angeles) for equipment ordering, Mr. A. Feizi (University of California, Los Angeles) for his help with the Labview code, M. Khan (University of California, Los Angeles) for his help with some of the MATLAB codes, Mr. J. Bronstein (University of Glasgow) for his help with ellipsometry measurements and the James Watt Nanofabrication Centre (University of Glasgow) for device manufacturing. The authors acknowledge the support of the NSF Engineering Research Center (ERC, PATHS-UP), NSF CBET, the ARO Life Sciences Division, the Howard Hughes Medical Institute (HHMI), an EPSRC grant (EP/K027611/1), an ERC Advanced Investigator Award (340117 Biophononics). M.K. acknowledges James Watt, David Brown Mobility and Mac Robertson scholarships. A.R. acknowledges the Ruth L. Kirschstein Institutional National Research Service Award (5T32DK104687-03) from NIH.
Author information
Authors and Affiliations
Contributions
The idea was conceived by M.K., A.R., J.C. and A.O. The experiments were performed by A.R. and M.K. The fluid analysis was performed by M.K., A.D., J.R. and J.C. A.R. and M.D. performed the electron microscopy. D.T. helped developed the imaging platform. M.K. integrated the IDT to the imaging platform. The paper was written via contributions of all the co-authors. A.O. and J.C. supervised the research.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Peer review information Nature Communications thanks Jian Liu and the other, anonymous, reviewer/s for their contribution to the peer review of this work.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ray, A., Khalid, M.A., Demčenko, A. et al. Holographic detection of nanoparticles using acoustically actuated nanolenses. Nat Commun 11, 171 (2020). https://doi.org/10.1038/s41467-019-13802-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-019-13802-1